Use of the reticulocyte channel warmed to 41 °C of the XN-9000 analyzer in samples with the presence of cold agglutinins

Objectives: The purpose of this study was to compare data obtained from the reticulocyte channel (RET channel) heated to 41 °C with those obtained from impedance channel (I-Channel) at room temperature in the samples with the mean corpuscular hemoglobin concentration (MCHC) < 370 g/L and in sampl...

Full description

Bibliographic Details
Main Authors: Vincenzo Roccaforte, Flavia Sciarini, Vanessa Proserpio, Ruggero Buonocore, Emanuela Marina Zavaroni, Silvia Burati, Marco Bussetti, Giammaria Liuzzi, Rosalba Monica Russo, Wanda Patrizia Porreca, Maria Luisa De Angelis, Carlo Federico Perno, Claudio Bonato, Stefano Pastori
Format: Article
Language:English
Published: Elsevier 2021-04-01
Series:Hematology, Transfusion and Cell Therapy
Subjects:
Online Access:http://www.sciencedirect.com/science/article/pii/S2531137920300249
id doaj-c28f1aeb84df4cdc89d210d48ddc0ff8
record_format Article
spelling doaj-c28f1aeb84df4cdc89d210d48ddc0ff82021-06-15T04:14:59ZengElsevierHematology, Transfusion and Cell Therapy2531-13792021-04-01432147155Use of the reticulocyte channel warmed to 41 °C of the XN-9000 analyzer in samples with the presence of cold agglutininsVincenzo Roccaforte0Flavia Sciarini1Vanessa Proserpio2Ruggero Buonocore3Emanuela Marina Zavaroni4Silvia Burati5Marco Bussetti6Giammaria Liuzzi7Rosalba Monica Russo8Wanda Patrizia Porreca9Maria Luisa De Angelis10Carlo Federico Perno11Claudio Bonato12Stefano Pastori13ASST Grande Ospedale Metropolitano Niguarda, Milan, Italy; Hospital Alessandro Manzoni, Lecco, Italy; University of Study Milan, Milan, Italy; Corresponding author at: Department of Laboratory Medicine, Area of Chemical Clinical Analysis, ASST Grande Ospedale Metropolitano Niguarda, Piazza Ospedale Maggiore 3, Milan, Italy.Hospital Alessandro Manzoni, Lecco, ItalyHospital Alessandro Manzoni, Lecco, ItalyASST Grande Ospedale Metropolitano Niguarda, Milan, ItalyASST Grande Ospedale Metropolitano Niguarda, Milan, ItalyASST Grande Ospedale Metropolitano Niguarda, Milan, ItalyASST Grande Ospedale Metropolitano Niguarda, Milan, ItalyASST Grande Ospedale Metropolitano Niguarda, Milan, ItalyASST Grande Ospedale Metropolitano Niguarda, Milan, ItalyASST Grande Ospedale Metropolitano Niguarda, Milan, ItalyASST Grande Ospedale Metropolitano Niguarda, Milan, ItalyASST Grande Ospedale Metropolitano Niguarda, Milan, ItalyHospital Alessandro Manzoni, Lecco, ItalyASST Grande Ospedale Metropolitano Niguarda, Milan, ItalyObjectives: The purpose of this study was to compare data obtained from the reticulocyte channel (RET channel) heated to 41 °C with those obtained from impedance channel (I-Channel) at room temperature in the samples with the mean corpuscular hemoglobin concentration (MCHC) < 370 g/L and in samples with the MCHC > 370 g/L, in the presence of cold agglutinins. Methods: In this study, 60 blood samples (group 1) with the MCHC < 370 g/L (without cold agglutinins) and 78 blood samples (group 2) with the MCHC > 370 g/L (with cold agglutinins) were used to compare the two analytical channels of the XN-9000 analyzer in different preanalytical conditions. The parameters evaluated in both groups were the following: red blood cell (RBC), hemoglobin (HGB), hematocrit (HCT), mean cell volume (MCV), RBC-most frequent volume (R-MFV), mean hemoglobin concentration (MCH) and mean cellular hemoglobin concentration (MCHC). Results: The results of this study showed an excellent correlation with both channels of the XN-9000 analyzer in samples with and without cold agglutinins, except for the MCHC. The bias between the values obtained in the I-channel and those obtained in the RET channel of both groups was insignificant and remained within the limits of acceptability, as reported by Ricos et al. for all considered parameters, except for MCHC. Conclusions: The presence of cold agglutinins in blood samples can be detected by a spurious lowering of the RBC count and by a spurious increase in the MCHC. The RET channel represents a great opportunity to correct the RBC count in a rapid manner without preheating. However, neither methodology can completely solve the residual presence of cold agglutinins in all samples, despite the MCHC values being < 370 g/L.http://www.sciencedirect.com/science/article/pii/S2531137920300249Cold agglutininsSysmexXN-9000RET channelAutoimmune hemolytic anemia
collection DOAJ
language English
format Article
sources DOAJ
author Vincenzo Roccaforte
Flavia Sciarini
Vanessa Proserpio
Ruggero Buonocore
Emanuela Marina Zavaroni
Silvia Burati
Marco Bussetti
Giammaria Liuzzi
Rosalba Monica Russo
Wanda Patrizia Porreca
Maria Luisa De Angelis
Carlo Federico Perno
Claudio Bonato
Stefano Pastori
spellingShingle Vincenzo Roccaforte
Flavia Sciarini
Vanessa Proserpio
Ruggero Buonocore
Emanuela Marina Zavaroni
Silvia Burati
Marco Bussetti
Giammaria Liuzzi
Rosalba Monica Russo
Wanda Patrizia Porreca
Maria Luisa De Angelis
Carlo Federico Perno
Claudio Bonato
Stefano Pastori
Use of the reticulocyte channel warmed to 41 °C of the XN-9000 analyzer in samples with the presence of cold agglutinins
Hematology, Transfusion and Cell Therapy
Cold agglutinins
Sysmex
XN-9000
RET channel
Autoimmune hemolytic anemia
author_facet Vincenzo Roccaforte
Flavia Sciarini
Vanessa Proserpio
Ruggero Buonocore
Emanuela Marina Zavaroni
Silvia Burati
Marco Bussetti
Giammaria Liuzzi
Rosalba Monica Russo
Wanda Patrizia Porreca
Maria Luisa De Angelis
Carlo Federico Perno
Claudio Bonato
Stefano Pastori
author_sort Vincenzo Roccaforte
title Use of the reticulocyte channel warmed to 41 °C of the XN-9000 analyzer in samples with the presence of cold agglutinins
title_short Use of the reticulocyte channel warmed to 41 °C of the XN-9000 analyzer in samples with the presence of cold agglutinins
title_full Use of the reticulocyte channel warmed to 41 °C of the XN-9000 analyzer in samples with the presence of cold agglutinins
title_fullStr Use of the reticulocyte channel warmed to 41 °C of the XN-9000 analyzer in samples with the presence of cold agglutinins
title_full_unstemmed Use of the reticulocyte channel warmed to 41 °C of the XN-9000 analyzer in samples with the presence of cold agglutinins
title_sort use of the reticulocyte channel warmed to 41 °c of the xn-9000 analyzer in samples with the presence of cold agglutinins
publisher Elsevier
series Hematology, Transfusion and Cell Therapy
issn 2531-1379
publishDate 2021-04-01
description Objectives: The purpose of this study was to compare data obtained from the reticulocyte channel (RET channel) heated to 41 °C with those obtained from impedance channel (I-Channel) at room temperature in the samples with the mean corpuscular hemoglobin concentration (MCHC) < 370 g/L and in samples with the MCHC > 370 g/L, in the presence of cold agglutinins. Methods: In this study, 60 blood samples (group 1) with the MCHC < 370 g/L (without cold agglutinins) and 78 blood samples (group 2) with the MCHC > 370 g/L (with cold agglutinins) were used to compare the two analytical channels of the XN-9000 analyzer in different preanalytical conditions. The parameters evaluated in both groups were the following: red blood cell (RBC), hemoglobin (HGB), hematocrit (HCT), mean cell volume (MCV), RBC-most frequent volume (R-MFV), mean hemoglobin concentration (MCH) and mean cellular hemoglobin concentration (MCHC). Results: The results of this study showed an excellent correlation with both channels of the XN-9000 analyzer in samples with and without cold agglutinins, except for the MCHC. The bias between the values obtained in the I-channel and those obtained in the RET channel of both groups was insignificant and remained within the limits of acceptability, as reported by Ricos et al. for all considered parameters, except for MCHC. Conclusions: The presence of cold agglutinins in blood samples can be detected by a spurious lowering of the RBC count and by a spurious increase in the MCHC. The RET channel represents a great opportunity to correct the RBC count in a rapid manner without preheating. However, neither methodology can completely solve the residual presence of cold agglutinins in all samples, despite the MCHC values being < 370 g/L.
topic Cold agglutinins
Sysmex
XN-9000
RET channel
Autoimmune hemolytic anemia
url http://www.sciencedirect.com/science/article/pii/S2531137920300249
work_keys_str_mv AT vincenzoroccaforte useofthereticulocytechannelwarmedto41cofthexn9000analyzerinsampleswiththepresenceofcoldagglutinins
AT flaviasciarini useofthereticulocytechannelwarmedto41cofthexn9000analyzerinsampleswiththepresenceofcoldagglutinins
AT vanessaproserpio useofthereticulocytechannelwarmedto41cofthexn9000analyzerinsampleswiththepresenceofcoldagglutinins
AT ruggerobuonocore useofthereticulocytechannelwarmedto41cofthexn9000analyzerinsampleswiththepresenceofcoldagglutinins
AT emanuelamarinazavaroni useofthereticulocytechannelwarmedto41cofthexn9000analyzerinsampleswiththepresenceofcoldagglutinins
AT silviaburati useofthereticulocytechannelwarmedto41cofthexn9000analyzerinsampleswiththepresenceofcoldagglutinins
AT marcobussetti useofthereticulocytechannelwarmedto41cofthexn9000analyzerinsampleswiththepresenceofcoldagglutinins
AT giammarialiuzzi useofthereticulocytechannelwarmedto41cofthexn9000analyzerinsampleswiththepresenceofcoldagglutinins
AT rosalbamonicarusso useofthereticulocytechannelwarmedto41cofthexn9000analyzerinsampleswiththepresenceofcoldagglutinins
AT wandapatriziaporreca useofthereticulocytechannelwarmedto41cofthexn9000analyzerinsampleswiththepresenceofcoldagglutinins
AT marialuisadeangelis useofthereticulocytechannelwarmedto41cofthexn9000analyzerinsampleswiththepresenceofcoldagglutinins
AT carlofedericoperno useofthereticulocytechannelwarmedto41cofthexn9000analyzerinsampleswiththepresenceofcoldagglutinins
AT claudiobonato useofthereticulocytechannelwarmedto41cofthexn9000analyzerinsampleswiththepresenceofcoldagglutinins
AT stefanopastori useofthereticulocytechannelwarmedto41cofthexn9000analyzerinsampleswiththepresenceofcoldagglutinins
_version_ 1721377608526987264