Same sized three-way indwelling urinary catheters from various manufacturers present different irrigation and drainage properties

Background: The three-way indwelling urinary catheter (IUC) is used for continuous bladder irrigation and is considered the cornerstone for clinical treatment of patients with macroscopic hematuria. Although there seems to be a logical relationship between catheter size and efficacy of irrigation an...

Full description

Bibliographic Details
Main Authors: Arie Carneiro, Marcelo Langer Wroclawski, Guilherme Andrade Peixoto, Jonathan Doyun Cha, Natasha Kouvaleski Saviano Moran, Felipe Ko Chen, Henrique Nelson Satkunas, José Roberto Augusto Campos, Alides Maria Mendes Rasabone Garcia, Manoj Monga, Gustavo Caserta Lemos
Format: Article
Language:English
Published: SAGE Publishing 2020-01-01
Series:Therapeutic Advances in Urology
Online Access:https://doi.org/10.1177/1756287219889496
id doaj-c1e207fc3d6540b7b3a94c6c2f4046ef
record_format Article
spelling doaj-c1e207fc3d6540b7b3a94c6c2f4046ef2020-11-25T03:42:15ZengSAGE PublishingTherapeutic Advances in Urology1756-28802020-01-011210.1177/1756287219889496Same sized three-way indwelling urinary catheters from various manufacturers present different irrigation and drainage propertiesArie CarneiroMarcelo Langer WroclawskiGuilherme Andrade PeixotoJonathan Doyun ChaNatasha Kouvaleski Saviano MoranFelipe Ko ChenHenrique Nelson SatkunasJosé Roberto Augusto CamposAlides Maria Mendes Rasabone GarciaManoj MongaGustavo Caserta LemosBackground: The three-way indwelling urinary catheter (IUC) is used for continuous bladder irrigation and is considered the cornerstone for clinical treatment of patients with macroscopic hematuria. Although there seems to be a logical relationship between catheter size and efficacy of irrigation and drainage, we often observe relevant variations in these parameters between different brands of catheters available on the market. The aim of this study was to compare the mechanical properties of different models of latex and silicone three-way catheters in an in vitro setting that resembles clinical use. Methods: Three different three-way catheters were evaluated: Gold Silicone-Coated Rusch ® (Model A), 100% Silicone Rusch ® (Model B) and X-Flow Coloplast ® (Model C). Irrigation channel, drainage channel, and overall cross-sectional areas were all digitally measured. Irrigation and drainage channel flow rates were measured and correlated with their corresponding catheter cross-sectional area values. Results: Different catheter models of the same caliber have different internal irrigation port diameters, internal drainage port diameters and internal cuff port diameters. The Model C IUC internal irrigation port diameter is significantly larger than models A and B. When flows were evaluated, we found that in the same model, the increase in caliber of the IUC was related to an increased drainage flow, but not to an increased irrigation flow. Conclusion: Precise measurements of the internal architecture of the three-way catheter, rather than relying on the caliber itself, could assist surgeons in choosing the best product for each specific patient, while minimizing complications.https://doi.org/10.1177/1756287219889496
collection DOAJ
language English
format Article
sources DOAJ
author Arie Carneiro
Marcelo Langer Wroclawski
Guilherme Andrade Peixoto
Jonathan Doyun Cha
Natasha Kouvaleski Saviano Moran
Felipe Ko Chen
Henrique Nelson Satkunas
José Roberto Augusto Campos
Alides Maria Mendes Rasabone Garcia
Manoj Monga
Gustavo Caserta Lemos
spellingShingle Arie Carneiro
Marcelo Langer Wroclawski
Guilherme Andrade Peixoto
Jonathan Doyun Cha
Natasha Kouvaleski Saviano Moran
Felipe Ko Chen
Henrique Nelson Satkunas
José Roberto Augusto Campos
Alides Maria Mendes Rasabone Garcia
Manoj Monga
Gustavo Caserta Lemos
Same sized three-way indwelling urinary catheters from various manufacturers present different irrigation and drainage properties
Therapeutic Advances in Urology
author_facet Arie Carneiro
Marcelo Langer Wroclawski
Guilherme Andrade Peixoto
Jonathan Doyun Cha
Natasha Kouvaleski Saviano Moran
Felipe Ko Chen
Henrique Nelson Satkunas
José Roberto Augusto Campos
Alides Maria Mendes Rasabone Garcia
Manoj Monga
Gustavo Caserta Lemos
author_sort Arie Carneiro
title Same sized three-way indwelling urinary catheters from various manufacturers present different irrigation and drainage properties
title_short Same sized three-way indwelling urinary catheters from various manufacturers present different irrigation and drainage properties
title_full Same sized three-way indwelling urinary catheters from various manufacturers present different irrigation and drainage properties
title_fullStr Same sized three-way indwelling urinary catheters from various manufacturers present different irrigation and drainage properties
title_full_unstemmed Same sized three-way indwelling urinary catheters from various manufacturers present different irrigation and drainage properties
title_sort same sized three-way indwelling urinary catheters from various manufacturers present different irrigation and drainage properties
publisher SAGE Publishing
series Therapeutic Advances in Urology
issn 1756-2880
publishDate 2020-01-01
description Background: The three-way indwelling urinary catheter (IUC) is used for continuous bladder irrigation and is considered the cornerstone for clinical treatment of patients with macroscopic hematuria. Although there seems to be a logical relationship between catheter size and efficacy of irrigation and drainage, we often observe relevant variations in these parameters between different brands of catheters available on the market. The aim of this study was to compare the mechanical properties of different models of latex and silicone three-way catheters in an in vitro setting that resembles clinical use. Methods: Three different three-way catheters were evaluated: Gold Silicone-Coated Rusch ® (Model A), 100% Silicone Rusch ® (Model B) and X-Flow Coloplast ® (Model C). Irrigation channel, drainage channel, and overall cross-sectional areas were all digitally measured. Irrigation and drainage channel flow rates were measured and correlated with their corresponding catheter cross-sectional area values. Results: Different catheter models of the same caliber have different internal irrigation port diameters, internal drainage port diameters and internal cuff port diameters. The Model C IUC internal irrigation port diameter is significantly larger than models A and B. When flows were evaluated, we found that in the same model, the increase in caliber of the IUC was related to an increased drainage flow, but not to an increased irrigation flow. Conclusion: Precise measurements of the internal architecture of the three-way catheter, rather than relying on the caliber itself, could assist surgeons in choosing the best product for each specific patient, while minimizing complications.
url https://doi.org/10.1177/1756287219889496
work_keys_str_mv AT ariecarneiro samesizedthreewayindwellingurinarycathetersfromvariousmanufacturerspresentdifferentirrigationanddrainageproperties
AT marcelolangerwroclawski samesizedthreewayindwellingurinarycathetersfromvariousmanufacturerspresentdifferentirrigationanddrainageproperties
AT guilhermeandradepeixoto samesizedthreewayindwellingurinarycathetersfromvariousmanufacturerspresentdifferentirrigationanddrainageproperties
AT jonathandoyuncha samesizedthreewayindwellingurinarycathetersfromvariousmanufacturerspresentdifferentirrigationanddrainageproperties
AT natashakouvaleskisavianomoran samesizedthreewayindwellingurinarycathetersfromvariousmanufacturerspresentdifferentirrigationanddrainageproperties
AT felipekochen samesizedthreewayindwellingurinarycathetersfromvariousmanufacturerspresentdifferentirrigationanddrainageproperties
AT henriquenelsonsatkunas samesizedthreewayindwellingurinarycathetersfromvariousmanufacturerspresentdifferentirrigationanddrainageproperties
AT joserobertoaugustocampos samesizedthreewayindwellingurinarycathetersfromvariousmanufacturerspresentdifferentirrigationanddrainageproperties
AT alidesmariamendesrasabonegarcia samesizedthreewayindwellingurinarycathetersfromvariousmanufacturerspresentdifferentirrigationanddrainageproperties
AT manojmonga samesizedthreewayindwellingurinarycathetersfromvariousmanufacturerspresentdifferentirrigationanddrainageproperties
AT gustavocasertalemos samesizedthreewayindwellingurinarycathetersfromvariousmanufacturerspresentdifferentirrigationanddrainageproperties
_version_ 1724526253600407552