Corneal Topographic and Aberrometric Measurements Obtained with a Multidiagnostic Device in Healthy Eyes: Intrasession Repeatability
Purpose. To evaluate the intrasession repeatability of corneal curvature, eccentricity, and aberrometric measurements obtained with a multidiagnostic device in healthy eyes. Methods. This study enrolled 107 eyes of 107 patients ranging in age from 23 to 65 years. All of them underwent a complete ant...
Main Authors: | , , , , , |
---|---|
Format: | Article |
Language: | English |
Published: |
Hindawi Limited
2017-01-01
|
Series: | Journal of Ophthalmology |
Online Access: | http://dx.doi.org/10.1155/2017/2149145 |
id |
doaj-bed9fe8a589946379e3640fda55bcac2 |
---|---|
record_format |
Article |
spelling |
doaj-bed9fe8a589946379e3640fda55bcac22020-11-24T20:52:39ZengHindawi LimitedJournal of Ophthalmology2090-004X2090-00582017-01-01201710.1155/2017/21491452149145Corneal Topographic and Aberrometric Measurements Obtained with a Multidiagnostic Device in Healthy Eyes: Intrasession RepeatabilityDavid P. Piñero0Alberto López-Navarro1Inmaculada Cabezos2Dolores de Fez3María T. Caballero4Vicent J. Camps5Group of Optics and Visual Perception, Department of Optics, Pharmacology and Anatomy, University of Alicante, Alicante, SpainOptometry Clinic, Prevention Service of the University of Alicante, Alicante, SpainGroup of Optics and Visual Perception, Department of Optics, Pharmacology and Anatomy, University of Alicante, Alicante, SpainGroup of Optics and Visual Perception, Department of Optics, Pharmacology and Anatomy, University of Alicante, Alicante, SpainGroup of Optics and Visual Perception, Department of Optics, Pharmacology and Anatomy, University of Alicante, Alicante, SpainGroup of Optics and Visual Perception, Department of Optics, Pharmacology and Anatomy, University of Alicante, Alicante, SpainPurpose. To evaluate the intrasession repeatability of corneal curvature, eccentricity, and aberrometric measurements obtained with a multidiagnostic device in healthy eyes. Methods. This study enrolled 107 eyes of 107 patients ranging in age from 23 to 65 years. All of them underwent a complete anterior segment examination with the VX120 system (Visionix-Luneau Technologies, Chartres, France). Three consecutive measurements were obtained. The within-subject standard deviation (Sw), intrasubject precision (1.96×Sw), and intraclass correlation coefficient (ICC) were calculated. Results. All Sw for corneal power measurements were below 0.26 D, with ICC above 0.982. The Sw for corneal astigmatism at different areas (3, 5, and 7 mm) was below 0.21 D, with ICC above 0.913. Concerning the axis of astigmatism, its Sw was below 11.27°, with ICC above 0.975. The Sw and ICC for corneal eccentricity were 0.067 and 0.957, respectively. The Sw and ICC for high-order aberration root mean square (RMS) were 0.048 µm and 0.901, respectively. For 3rd- and 4th-order aberrometric parameters, all Sw were below 0.037 µm and all ICC were higher than 0.84, except for quadrafoil RMS (ICC: 0.689). Conclusions. The multidiagnostic device evaluated is able to provide consistent measurements of corneal power, eccentricity, and third- and fourth-order aberrations in healthy eyes.http://dx.doi.org/10.1155/2017/2149145 |
collection |
DOAJ |
language |
English |
format |
Article |
sources |
DOAJ |
author |
David P. Piñero Alberto López-Navarro Inmaculada Cabezos Dolores de Fez María T. Caballero Vicent J. Camps |
spellingShingle |
David P. Piñero Alberto López-Navarro Inmaculada Cabezos Dolores de Fez María T. Caballero Vicent J. Camps Corneal Topographic and Aberrometric Measurements Obtained with a Multidiagnostic Device in Healthy Eyes: Intrasession Repeatability Journal of Ophthalmology |
author_facet |
David P. Piñero Alberto López-Navarro Inmaculada Cabezos Dolores de Fez María T. Caballero Vicent J. Camps |
author_sort |
David P. Piñero |
title |
Corneal Topographic and Aberrometric Measurements Obtained with a Multidiagnostic Device in Healthy Eyes: Intrasession Repeatability |
title_short |
Corneal Topographic and Aberrometric Measurements Obtained with a Multidiagnostic Device in Healthy Eyes: Intrasession Repeatability |
title_full |
Corneal Topographic and Aberrometric Measurements Obtained with a Multidiagnostic Device in Healthy Eyes: Intrasession Repeatability |
title_fullStr |
Corneal Topographic and Aberrometric Measurements Obtained with a Multidiagnostic Device in Healthy Eyes: Intrasession Repeatability |
title_full_unstemmed |
Corneal Topographic and Aberrometric Measurements Obtained with a Multidiagnostic Device in Healthy Eyes: Intrasession Repeatability |
title_sort |
corneal topographic and aberrometric measurements obtained with a multidiagnostic device in healthy eyes: intrasession repeatability |
publisher |
Hindawi Limited |
series |
Journal of Ophthalmology |
issn |
2090-004X 2090-0058 |
publishDate |
2017-01-01 |
description |
Purpose. To evaluate the intrasession repeatability of corneal curvature, eccentricity, and aberrometric measurements obtained with a multidiagnostic device in healthy eyes. Methods. This study enrolled 107 eyes of 107 patients ranging in age from 23 to 65 years. All of them underwent a complete anterior segment examination with the VX120 system (Visionix-Luneau Technologies, Chartres, France). Three consecutive measurements were obtained. The within-subject standard deviation (Sw), intrasubject precision (1.96×Sw), and intraclass correlation coefficient (ICC) were calculated. Results. All Sw for corneal power measurements were below 0.26 D, with ICC above 0.982. The Sw for corneal astigmatism at different areas (3, 5, and 7 mm) was below 0.21 D, with ICC above 0.913. Concerning the axis of astigmatism, its Sw was below 11.27°, with ICC above 0.975. The Sw and ICC for corneal eccentricity were 0.067 and 0.957, respectively. The Sw and ICC for high-order aberration root mean square (RMS) were 0.048 µm and 0.901, respectively. For 3rd- and 4th-order aberrometric parameters, all Sw were below 0.037 µm and all ICC were higher than 0.84, except for quadrafoil RMS (ICC: 0.689). Conclusions. The multidiagnostic device evaluated is able to provide consistent measurements of corneal power, eccentricity, and third- and fourth-order aberrations in healthy eyes. |
url |
http://dx.doi.org/10.1155/2017/2149145 |
work_keys_str_mv |
AT davidppinero cornealtopographicandaberrometricmeasurementsobtainedwithamultidiagnosticdeviceinhealthyeyesintrasessionrepeatability AT albertolopeznavarro cornealtopographicandaberrometricmeasurementsobtainedwithamultidiagnosticdeviceinhealthyeyesintrasessionrepeatability AT inmaculadacabezos cornealtopographicandaberrometricmeasurementsobtainedwithamultidiagnosticdeviceinhealthyeyesintrasessionrepeatability AT doloresdefez cornealtopographicandaberrometricmeasurementsobtainedwithamultidiagnosticdeviceinhealthyeyesintrasessionrepeatability AT mariatcaballero cornealtopographicandaberrometricmeasurementsobtainedwithamultidiagnosticdeviceinhealthyeyesintrasessionrepeatability AT vicentjcamps cornealtopographicandaberrometricmeasurementsobtainedwithamultidiagnosticdeviceinhealthyeyesintrasessionrepeatability |
_version_ |
1716799096502616064 |