Multidimensional Sternal Fixation to Overcome a “Floating” Sternum

This case report describes the repair of a complete sternal dehiscence of the lower right sternum using sternal wires, manubrial plates, and a Talon closure device for rigid, multidimensional sternal fixation. Sternal dehiscence is a rare but significant cause of morbidity for patients undergoing m...

Full description

Bibliographic Details
Main Authors: William Rothstein, Tyler Spata, Bryan Whitson, Ahmet Kilic
Format: Article
Language:English
Published: Hindawi Limited 2014-01-01
Series:Case Reports in Surgery
Online Access:http://dx.doi.org/10.1155/2014/690160
Description
Summary:This case report describes the repair of a complete sternal dehiscence of the lower right sternum using sternal wires, manubrial plates, and a Talon closure device for rigid, multidimensional sternal fixation. Sternal dehiscence is a rare but significant cause of morbidity for patients undergoing median sternotomy. The risk factors for this complication are well described and although sternal wires have traditionally been used for primary closure, rigid fixation with sternal plates is a viable alternative to avoid dehiscence in this high-risk cohort.
ISSN:2090-6900