Inevitable high-dose irradiation to lead of implantable cardioverter defibrillator in small cell lung cancer: a case report

Abstract Background Radiotherapy has been shown to cause malfunction of implantable cardioverter-defibrillators, and there are few studies of implantable cardioverter-defibrillators and radiotherapy. We report an unusual case of small cell lung cancer in a patient with an implantable cardioverter-de...

Full description

Bibliographic Details
Main Authors: Jeong Won Lee, Ki Ho Seol
Format: Article
Language:English
Published: BMC 2019-06-01
Series:Journal of Medical Case Reports
Subjects:
Online Access:http://link.springer.com/article/10.1186/s13256-019-2111-y
id doaj-bc23e26db1ba4fce97e15f28ce6c359e
record_format Article
spelling doaj-bc23e26db1ba4fce97e15f28ce6c359e2020-11-25T02:24:33ZengBMCJournal of Medical Case Reports1752-19472019-06-011311610.1186/s13256-019-2111-yInevitable high-dose irradiation to lead of implantable cardioverter defibrillator in small cell lung cancer: a case reportJeong Won Lee0Ki Ho Seol1Department of Radiation Oncology, Catholic University of Daegu School of MedicineDepartment of Radiation Oncology, Catholic University of Daegu School of MedicineAbstract Background Radiotherapy has been shown to cause malfunction of implantable cardioverter-defibrillators, and there are few studies of implantable cardioverter-defibrillators and radiotherapy. We report an unusual case of small cell lung cancer in a patient with an implantable cardioverter-defibrillator in whom direct irradiation to the electrode and lead could not be avoided. Case presentation We report a case of radiotherapy in a 72-year-old Korean man with a limited stage of small cell lung cancer who had undergone insertion of an implantable cardioverter-defibrillator because of ventricular fibrillation. The radiation dose was 60 Gy in 30 fractions to the thorax. The mean dose and maximum dose estimated at the body of the implantable cardioverter-defibrillator were 0.89 Gy and 2.23 Gy, respectively. The mean and maximum doses of the lead and electrode were 17.12 Gy and 55.72 Gy in the lead and 1.81 Gy and 7.10 Gy in the electrode, respectively, because part of the lead and electrode was inevitably in the irradiated fields. The function of the patient’s implantable cardioverter-defibrillator was checked daily, and no change in implantable cardioverter-defibrillator function was observed for the duration of radiotherapy. The patient was tolerated the treatment well without severe complications. Computed tomography performed at 4 weeks after radiotherapy showed a good response with regression of the tumor. The patient was alive with complete remission of the tumor and without any implantable cardioverter-defibrillator dysfunction more than 36 months after the end of treatment. Conclusions This case demonstrates that radiotherapy may be a safe and effective treatment modality through careful monitoring of implantable cardioverter-defibrillators in patients with lung cancer who have implantable cardioverter-defibrillators.http://link.springer.com/article/10.1186/s13256-019-2111-yImplantable cardioverter-defibrillatorLung cancerRadiotherapy
collection DOAJ
language English
format Article
sources DOAJ
author Jeong Won Lee
Ki Ho Seol
spellingShingle Jeong Won Lee
Ki Ho Seol
Inevitable high-dose irradiation to lead of implantable cardioverter defibrillator in small cell lung cancer: a case report
Journal of Medical Case Reports
Implantable cardioverter-defibrillator
Lung cancer
Radiotherapy
author_facet Jeong Won Lee
Ki Ho Seol
author_sort Jeong Won Lee
title Inevitable high-dose irradiation to lead of implantable cardioverter defibrillator in small cell lung cancer: a case report
title_short Inevitable high-dose irradiation to lead of implantable cardioverter defibrillator in small cell lung cancer: a case report
title_full Inevitable high-dose irradiation to lead of implantable cardioverter defibrillator in small cell lung cancer: a case report
title_fullStr Inevitable high-dose irradiation to lead of implantable cardioverter defibrillator in small cell lung cancer: a case report
title_full_unstemmed Inevitable high-dose irradiation to lead of implantable cardioverter defibrillator in small cell lung cancer: a case report
title_sort inevitable high-dose irradiation to lead of implantable cardioverter defibrillator in small cell lung cancer: a case report
publisher BMC
series Journal of Medical Case Reports
issn 1752-1947
publishDate 2019-06-01
description Abstract Background Radiotherapy has been shown to cause malfunction of implantable cardioverter-defibrillators, and there are few studies of implantable cardioverter-defibrillators and radiotherapy. We report an unusual case of small cell lung cancer in a patient with an implantable cardioverter-defibrillator in whom direct irradiation to the electrode and lead could not be avoided. Case presentation We report a case of radiotherapy in a 72-year-old Korean man with a limited stage of small cell lung cancer who had undergone insertion of an implantable cardioverter-defibrillator because of ventricular fibrillation. The radiation dose was 60 Gy in 30 fractions to the thorax. The mean dose and maximum dose estimated at the body of the implantable cardioverter-defibrillator were 0.89 Gy and 2.23 Gy, respectively. The mean and maximum doses of the lead and electrode were 17.12 Gy and 55.72 Gy in the lead and 1.81 Gy and 7.10 Gy in the electrode, respectively, because part of the lead and electrode was inevitably in the irradiated fields. The function of the patient’s implantable cardioverter-defibrillator was checked daily, and no change in implantable cardioverter-defibrillator function was observed for the duration of radiotherapy. The patient was tolerated the treatment well without severe complications. Computed tomography performed at 4 weeks after radiotherapy showed a good response with regression of the tumor. The patient was alive with complete remission of the tumor and without any implantable cardioverter-defibrillator dysfunction more than 36 months after the end of treatment. Conclusions This case demonstrates that radiotherapy may be a safe and effective treatment modality through careful monitoring of implantable cardioverter-defibrillators in patients with lung cancer who have implantable cardioverter-defibrillators.
topic Implantable cardioverter-defibrillator
Lung cancer
Radiotherapy
url http://link.springer.com/article/10.1186/s13256-019-2111-y
work_keys_str_mv AT jeongwonlee inevitablehighdoseirradiationtoleadofimplantablecardioverterdefibrillatorinsmallcelllungcanceracasereport
AT kihoseol inevitablehighdoseirradiationtoleadofimplantablecardioverterdefibrillatorinsmallcelllungcanceracasereport