Attenuation Correction for Small Animal PET Images: A Comparison of Two Methods

In order to extract quantitative parameters from PET images, several physical effects such as photon attenuation, scatter, and partial volume must be taken into account. The main objectives of this work were the evaluation of photon attenuation in small animals and the implementation of two attenuat...

Full description

Bibliographic Details
Main Authors: Daniela D'Ambrosio, Federico Zagni, Antonello E. Spinelli, Mario Marengo
Format: Article
Language:English
Published: Hindawi Limited 2013-01-01
Series:Computational and Mathematical Methods in Medicine
Online Access:http://dx.doi.org/10.1155/2013/103476
Description
Summary:In order to extract quantitative parameters from PET images, several physical effects such as photon attenuation, scatter, and partial volume must be taken into account. The main objectives of this work were the evaluation of photon attenuation in small animals and the implementation of two attenuation correction methods based on X-rays CT and segmentation of emission images. The accuracy of the first method with respect to the beam hardening effect was investigated by using Monte Carlo simulations. Mouse- and rat-sized phantoms were acquired in order to evaluate attenuation correction in terms of counts increment and recovery of uniform activity concentration. Both methods were applied to mice and rat images acquired with several radiotracers such asF18-FDG, 11C-acetate, 68Ga-chloride, and F18-NaF. The accuracy of the proposed methods was evaluated in heart and tumour tissues using F18-FDG images and in liver, kidney, and spinal column tissues using C11-acetate, Ga68-chloride, and F18-NaF images, respectively. In vivo results from animal studies show that, except for bone scans, differences between the proposed methods were about 10% in rats and 3% in mice. In conclusion, both methods provide equivalent results; however, the segmentation-based approach has several advantages being less time consuming and simple to implement.
ISSN:1748-670X
1748-6718