Does Every Calculation Formula Fit for All Types of Intraocular Lenses? Optimization of Constants for Tecnis ZA9003 and ZCB00 Is Necessary
<i>Background and Objectives</i>: To evaluate the performance of intraocular lenses (IOLs) using power calculation formulas on different types of IOL. <i>Materials and Methods</i>: 120 eyes and four IOL types (BioLine Yellow Accurate Aspheric IOL (i-Medical), TECNIS ZCB00, TE...
Main Authors: | , , , |
---|---|
Format: | Article |
Language: | English |
Published: |
MDPI AG
2021-03-01
|
Series: | Medicina |
Subjects: | |
Online Access: | https://www.mdpi.com/1648-9144/57/4/319 |
Summary: | <i>Background and Objectives</i>: To evaluate the performance of intraocular lenses (IOLs) using power calculation formulas on different types of IOL. <i>Materials and Methods</i>: 120 eyes and four IOL types (BioLine Yellow Accurate Aspheric IOL (i-Medical), TECNIS ZCB00, TECNIS ZA9003 (Johnson & Johnson) (3-piece IOL) and Softec HD (Lenstec)) were analyzed. The performance of Haigis, Barret Universal II and SKR-II formulas were compared between IOL types. The mean prediction error (ME) and mean absolute prediction error (MAE) were analyzed. <i>Results</i>: The overall percentage of eyes predicted within ±0.25 diopters (D) was 40.8% for Barret; 39.2% Haigis and 31.7% for SRK-II. Barret and Haigis had a significantly lower MAE than SRK-II (<i>p</i> < 0.05). The results differed among IOL types. The largest portion of eyes predicted within ±0.25 D was with the Barret formula in ZCB00 (33.3%) and ZA9003 (43.3%). Haigis was the most accurate in Softec HD (50%) and SRK-II in Biolline Yellow IOL (50%). ZCB00 showed a clinically significant hypermetropic ME compared to other IOLs. <i>Conclusions</i>: In general, Barret formulas had the best performance as a universal formula. However, the formula should be chosen according to the type of IOL in order to obtain the best results. Constant optimizations are necessary for the Tecnis IOL ZCB00 and ZA9003, as all of the analyzed formulas achieved a clinically significant poor performance in this type of IOL. ZCB00 also showed a hypermetropic shift in ME in all the formulas. |
---|---|
ISSN: | 1010-660X 1648-9144 |