Cytotoxicity Evaluation of The Bioresorbable and Titanium Plates/Screws Used in Maxillofacial Surgery on Gingival Fibroblasts and Human Mesenchymal Bone Marrow Stem Cells

Objective: Bioresorbable and titanium plates/screws are considered as a standard treatment for fixation of the bone segments of craniofacial area and paying attention to their biocompatibility is an important issue along with other aspects of application. The purpose of the study was to evaluate t...

Full description

Bibliographic Details
Main Authors: Masoud Vatani, Mohammad Hossein Beigi, Fatemeh Ejeian, Ahmad Mottaghi, Afshin Yadegari-Naeini, Mohammad Hossein Nasr-Esfahani
Format: Article
Language:English
Published: Royan Institute (ACECR), Tehran 2020-10-01
Series:Cell Journal
Subjects:
Online Access:https://celljournal.org/journal/article/abstract/6409
id doaj-b3b1dcaec23b4f2983a0c75253e3c614
record_format Article
spelling doaj-b3b1dcaec23b4f2983a0c75253e3c6142020-11-25T02:28:40ZengRoyan Institute (ACECR), TehranCell Journal2228-58062228-58142020-10-0122331031810.22074/cellj.2020.6409Cytotoxicity Evaluation of The Bioresorbable and Titanium Plates/Screws Used in Maxillofacial Surgery on Gingival Fibroblasts and Human Mesenchymal Bone Marrow Stem CellsMasoud Vatani0Mohammad Hossein Beigi1Fatemeh Ejeian2Ahmad Mottaghi3Afshin Yadegari-Naeini4Mohammad Hossein Nasr-Esfahani5Department of Oral and Maxillofacial Surgery, School of Dentistry, Islamic Azad University, Khorasgan Branch, Isfahan, IranDepartment of Cellular Biotechnology, Cell Science Research Center, Royan Institute for Biotechnology, ACECR, Isfahan, IranDepartment of Cellular Biotechnology, Cell Science Research Center, Royan Institute for Biotechnology, ACECR, Isfahan, IranDepartment of Oral and Maxillofacial Surgery, School of Dentistry, Islamic Azad University, Khorasgan Branch, Isfahan, IranDepartment of Oral and Maxillofacial Surgery, School of Dentistry, Islamic Azad University, Khorasgan Branch, Isfahan, IranDepartment of Cellular Biotechnology, Cell Science Research Center, Royan Institute for Biotechnology, ACECR, Isfahan, IranObjective: Bioresorbable and titanium plates/screws are considered as a standard treatment for fixation of the bone segments of craniofacial area and paying attention to their biocompatibility is an important issue along with other aspects of application. The purpose of the study was to evaluate the cell viability of two types of plate and screw used in maxillofacial surgeries in contact with gingival fibroblasts and bone marrow stem cells. Materials and Methods: In this experimental study after extraction and cultivation of cells from healthy human gingival tissue and alveolar bone of jaw, cytotoxicity of device was evaluated. In direct contact method, samples had near vicinity contact with the both cell lines and in indirect contact method, by-products released, like ions, from samples after 8 weeks were used to assess cytotoxicity. Then cytotoxicity was evaluated on the 2nd, 4th and 6th day with MTS tests and microscopy. The data were analyzed by one-way ANOVA and independent t tests. Results: There was a statistically significant difference between the German plate and screw and all the samples studied on day 6 (P<0.05). Furthermore, a statistically significant difference was observed between both metal samples and both bio-absorbable samples on day 6 and both cell lines (P<0.05). Comparisons between the two groups with each other for both cell lines on the 6th day were statistically significant (P<0.05). Conclusion: Our results suggest that that cytotoxicity of biomaterial, from different brands, were not similar and some of the biomaterial showed lower degree of toxicity compared to others and specialist using these products showed be aware of this differences. Our investigation indicates more biocompatibility of bioresorbable plates and screws compared to titanium. In addition our results suggest that biomaterials were not completely neutral.https://celljournal.org/journal/article/abstract/6409bone marrow stem cellcytotoxicity testdental implant materialsfibroblast cells
collection DOAJ
language English
format Article
sources DOAJ
author Masoud Vatani
Mohammad Hossein Beigi
Fatemeh Ejeian
Ahmad Mottaghi
Afshin Yadegari-Naeini
Mohammad Hossein Nasr-Esfahani
spellingShingle Masoud Vatani
Mohammad Hossein Beigi
Fatemeh Ejeian
Ahmad Mottaghi
Afshin Yadegari-Naeini
Mohammad Hossein Nasr-Esfahani
Cytotoxicity Evaluation of The Bioresorbable and Titanium Plates/Screws Used in Maxillofacial Surgery on Gingival Fibroblasts and Human Mesenchymal Bone Marrow Stem Cells
Cell Journal
bone marrow stem cell
cytotoxicity test
dental implant materials
fibroblast cells
author_facet Masoud Vatani
Mohammad Hossein Beigi
Fatemeh Ejeian
Ahmad Mottaghi
Afshin Yadegari-Naeini
Mohammad Hossein Nasr-Esfahani
author_sort Masoud Vatani
title Cytotoxicity Evaluation of The Bioresorbable and Titanium Plates/Screws Used in Maxillofacial Surgery on Gingival Fibroblasts and Human Mesenchymal Bone Marrow Stem Cells
title_short Cytotoxicity Evaluation of The Bioresorbable and Titanium Plates/Screws Used in Maxillofacial Surgery on Gingival Fibroblasts and Human Mesenchymal Bone Marrow Stem Cells
title_full Cytotoxicity Evaluation of The Bioresorbable and Titanium Plates/Screws Used in Maxillofacial Surgery on Gingival Fibroblasts and Human Mesenchymal Bone Marrow Stem Cells
title_fullStr Cytotoxicity Evaluation of The Bioresorbable and Titanium Plates/Screws Used in Maxillofacial Surgery on Gingival Fibroblasts and Human Mesenchymal Bone Marrow Stem Cells
title_full_unstemmed Cytotoxicity Evaluation of The Bioresorbable and Titanium Plates/Screws Used in Maxillofacial Surgery on Gingival Fibroblasts and Human Mesenchymal Bone Marrow Stem Cells
title_sort cytotoxicity evaluation of the bioresorbable and titanium plates/screws used in maxillofacial surgery on gingival fibroblasts and human mesenchymal bone marrow stem cells
publisher Royan Institute (ACECR), Tehran
series Cell Journal
issn 2228-5806
2228-5814
publishDate 2020-10-01
description Objective: Bioresorbable and titanium plates/screws are considered as a standard treatment for fixation of the bone segments of craniofacial area and paying attention to their biocompatibility is an important issue along with other aspects of application. The purpose of the study was to evaluate the cell viability of two types of plate and screw used in maxillofacial surgeries in contact with gingival fibroblasts and bone marrow stem cells. Materials and Methods: In this experimental study after extraction and cultivation of cells from healthy human gingival tissue and alveolar bone of jaw, cytotoxicity of device was evaluated. In direct contact method, samples had near vicinity contact with the both cell lines and in indirect contact method, by-products released, like ions, from samples after 8 weeks were used to assess cytotoxicity. Then cytotoxicity was evaluated on the 2nd, 4th and 6th day with MTS tests and microscopy. The data were analyzed by one-way ANOVA and independent t tests. Results: There was a statistically significant difference between the German plate and screw and all the samples studied on day 6 (P<0.05). Furthermore, a statistically significant difference was observed between both metal samples and both bio-absorbable samples on day 6 and both cell lines (P<0.05). Comparisons between the two groups with each other for both cell lines on the 6th day were statistically significant (P<0.05). Conclusion: Our results suggest that that cytotoxicity of biomaterial, from different brands, were not similar and some of the biomaterial showed lower degree of toxicity compared to others and specialist using these products showed be aware of this differences. Our investigation indicates more biocompatibility of bioresorbable plates and screws compared to titanium. In addition our results suggest that biomaterials were not completely neutral.
topic bone marrow stem cell
cytotoxicity test
dental implant materials
fibroblast cells
url https://celljournal.org/journal/article/abstract/6409
work_keys_str_mv AT masoudvatani cytotoxicityevaluationofthebioresorbableandtitaniumplatesscrewsusedinmaxillofacialsurgeryongingivalfibroblastsandhumanmesenchymalbonemarrowstemcells
AT mohammadhosseinbeigi cytotoxicityevaluationofthebioresorbableandtitaniumplatesscrewsusedinmaxillofacialsurgeryongingivalfibroblastsandhumanmesenchymalbonemarrowstemcells
AT fatemehejeian cytotoxicityevaluationofthebioresorbableandtitaniumplatesscrewsusedinmaxillofacialsurgeryongingivalfibroblastsandhumanmesenchymalbonemarrowstemcells
AT ahmadmottaghi cytotoxicityevaluationofthebioresorbableandtitaniumplatesscrewsusedinmaxillofacialsurgeryongingivalfibroblastsandhumanmesenchymalbonemarrowstemcells
AT afshinyadegarinaeini cytotoxicityevaluationofthebioresorbableandtitaniumplatesscrewsusedinmaxillofacialsurgeryongingivalfibroblastsandhumanmesenchymalbonemarrowstemcells
AT mohammadhosseinnasresfahani cytotoxicityevaluationofthebioresorbableandtitaniumplatesscrewsusedinmaxillofacialsurgeryongingivalfibroblastsandhumanmesenchymalbonemarrowstemcells
_version_ 1724837285588893696