Cytotoxicity Evaluation of The Bioresorbable and Titanium Plates/Screws Used in Maxillofacial Surgery on Gingival Fibroblasts and Human Mesenchymal Bone Marrow Stem Cells
Objective: Bioresorbable and titanium plates/screws are considered as a standard treatment for fixation of the bone segments of craniofacial area and paying attention to their biocompatibility is an important issue along with other aspects of application. The purpose of the study was to evaluate t...
Main Authors: | , , , , , |
---|---|
Format: | Article |
Language: | English |
Published: |
Royan Institute (ACECR), Tehran
2020-10-01
|
Series: | Cell Journal |
Subjects: | |
Online Access: | https://celljournal.org/journal/article/abstract/6409 |
Summary: | Objective: Bioresorbable and titanium plates/screws are considered as a standard treatment for fixation of the bone
segments of craniofacial area and paying attention to their biocompatibility is an important issue along with other
aspects of application. The purpose of the study was to evaluate the cell viability of two types of plate and screw used
in maxillofacial surgeries in contact with gingival fibroblasts and bone marrow stem cells.
Materials and Methods: In this experimental study after extraction and cultivation of cells from healthy human gingival
tissue and alveolar bone of jaw, cytotoxicity of device was evaluated. In direct contact method, samples had near
vicinity contact with the both cell lines and in indirect contact method, by-products released, like ions, from samples
after 8 weeks were used to assess cytotoxicity. Then cytotoxicity was evaluated on the 2nd, 4th and 6th day with MTS
tests and microscopy. The data were analyzed by one-way ANOVA and independent t tests.
Results: There was a statistically significant difference between the German plate and screw and all the samples
studied on day 6 (P<0.05). Furthermore, a statistically significant difference was observed between both metal samples
and both bio-absorbable samples on day 6 and both cell lines (P<0.05). Comparisons between the two groups with
each other for both cell lines on the 6th day were statistically significant (P<0.05).
Conclusion: Our results suggest that that cytotoxicity of biomaterial, from different brands, were not similar and some
of the biomaterial showed lower degree of toxicity compared to others and specialist using these products showed
be aware of this differences. Our investigation indicates more biocompatibility of bioresorbable plates and screws
compared to titanium. In addition our results suggest that biomaterials were not completely neutral. |
---|---|
ISSN: | 2228-5806 2228-5814 |