Comparative Analysis of Erythrocyte Sedimentation Rate Measured by Automated and Manual Methods in Anaemic Patients

Purpose Erythrocyte sedimentation rate (ESR) is a widely used indicator of inflammation and a routinely done hematology investigation to monitor patients of autoimmune and infectious diseases. We aimed to compare the ESR results obtained by Roller 20LC automated instrument and standard reference Wes...

Full description

Bibliographic Details
Main Authors: Vikram Narang, Sumit Grover, Amandeep Kaur Kang, Avantika Garg, Neena Sood
Format: Article
Language:English
Published: Thieme Medical and Scientific Publishers Pvt. Ltd. 2020-12-01
Series:Journal of Laboratory Physicians
Subjects:
Online Access:http://www.thieme-connect.de/DOI/DOI?10.1055/s-0040-1721155
id doaj-b356cb0fb80a4ff8b6e8643678e3341f
record_format Article
spelling doaj-b356cb0fb80a4ff8b6e8643678e3341f2020-12-30T23:40:30ZengThieme Medical and Scientific Publishers Pvt. Ltd.Journal of Laboratory Physicians0974-27270974-78262020-12-01120423924310.1055/s-0040-1721155Comparative Analysis of Erythrocyte Sedimentation Rate Measured by Automated and Manual Methods in Anaemic PatientsVikram Narang0Sumit Grover1Amandeep Kaur Kang2Avantika Garg3Neena Sood4Department of Pathology, Dayanand Medical College & Hospital, Ludhiana, Punjab, IndiaDepartment of Pathology, Dayanand Medical College & Hospital, Ludhiana, Punjab, IndiaDepartment of Pathology, Dayanand Medical College & Hospital, Ludhiana, Punjab, IndiaDepartment of Pathology, Dayanand Medical College & Hospital, Ludhiana, Punjab, IndiaDepartment of Pathology, Dayanand Medical College & Hospital, Ludhiana, Punjab, IndiaPurpose Erythrocyte sedimentation rate (ESR) is a widely used indicator of inflammation and a routinely done hematology investigation to monitor patients of autoimmune and infectious diseases. We aimed to compare the ESR results obtained by Roller 20LC automated instrument and standard reference Westergren method and analyzed the effect of anemia (hematocrit) on ESR measurements through the automated method. Methods We analyzed 1377 random anemic OPD patients (hematocrit [HCT] < 35%) for ESR levels measured by Roller 20LC using EDTA blood and Westergren method using citrated blood for a one and half year period from January 1, 2018 to June 30, 2019. Fabry’s formula was used to correct the Westergren ESR. Results The total number of samples after evaluation were divided into low (n = 232), intermediate (n = 417), high (n = 406), and very high range of ESR (≥100 mm/hr; n = 422). Mean difference between values of corrected and automated ESR for the low, intermediate, high and very high ESR range was 2.33 ± 5.03, 10.95 ± 8.04, 28.22 ± 19.11 and 43.3 ± 19.22 mm/hr, respectively. The 95% limit of agreement calculated by the Bland–Altmann analysis between the two methods for low-ESR range was −7.53 to 12.2 (highest correlation coefficient –0.65), while for very high ESR, range was −5.1 to 81.5 (least coefficient of 0.18) (p < 0.001). Conclusion In laboratories with high-sample load and where manual measurement may be tedious, the automated method of ESR measurement can safely replace the Westergren method for low-ESR values in patients with low hematocrit. While for high-ESR values, validation by the standard Westergren method may be needed.http://www.thieme-connect.de/DOI/DOI?10.1055/s-0040-1721155corrected esrhematocritfabry’s formulaanemia
collection DOAJ
language English
format Article
sources DOAJ
author Vikram Narang
Sumit Grover
Amandeep Kaur Kang
Avantika Garg
Neena Sood
spellingShingle Vikram Narang
Sumit Grover
Amandeep Kaur Kang
Avantika Garg
Neena Sood
Comparative Analysis of Erythrocyte Sedimentation Rate Measured by Automated and Manual Methods in Anaemic Patients
Journal of Laboratory Physicians
corrected esr
hematocrit
fabry’s formula
anemia
author_facet Vikram Narang
Sumit Grover
Amandeep Kaur Kang
Avantika Garg
Neena Sood
author_sort Vikram Narang
title Comparative Analysis of Erythrocyte Sedimentation Rate Measured by Automated and Manual Methods in Anaemic Patients
title_short Comparative Analysis of Erythrocyte Sedimentation Rate Measured by Automated and Manual Methods in Anaemic Patients
title_full Comparative Analysis of Erythrocyte Sedimentation Rate Measured by Automated and Manual Methods in Anaemic Patients
title_fullStr Comparative Analysis of Erythrocyte Sedimentation Rate Measured by Automated and Manual Methods in Anaemic Patients
title_full_unstemmed Comparative Analysis of Erythrocyte Sedimentation Rate Measured by Automated and Manual Methods in Anaemic Patients
title_sort comparative analysis of erythrocyte sedimentation rate measured by automated and manual methods in anaemic patients
publisher Thieme Medical and Scientific Publishers Pvt. Ltd.
series Journal of Laboratory Physicians
issn 0974-2727
0974-7826
publishDate 2020-12-01
description Purpose Erythrocyte sedimentation rate (ESR) is a widely used indicator of inflammation and a routinely done hematology investigation to monitor patients of autoimmune and infectious diseases. We aimed to compare the ESR results obtained by Roller 20LC automated instrument and standard reference Westergren method and analyzed the effect of anemia (hematocrit) on ESR measurements through the automated method. Methods We analyzed 1377 random anemic OPD patients (hematocrit [HCT] < 35%) for ESR levels measured by Roller 20LC using EDTA blood and Westergren method using citrated blood for a one and half year period from January 1, 2018 to June 30, 2019. Fabry’s formula was used to correct the Westergren ESR. Results The total number of samples after evaluation were divided into low (n = 232), intermediate (n = 417), high (n = 406), and very high range of ESR (≥100 mm/hr; n = 422). Mean difference between values of corrected and automated ESR for the low, intermediate, high and very high ESR range was 2.33 ± 5.03, 10.95 ± 8.04, 28.22 ± 19.11 and 43.3 ± 19.22 mm/hr, respectively. The 95% limit of agreement calculated by the Bland–Altmann analysis between the two methods for low-ESR range was −7.53 to 12.2 (highest correlation coefficient –0.65), while for very high ESR, range was −5.1 to 81.5 (least coefficient of 0.18) (p < 0.001). Conclusion In laboratories with high-sample load and where manual measurement may be tedious, the automated method of ESR measurement can safely replace the Westergren method for low-ESR values in patients with low hematocrit. While for high-ESR values, validation by the standard Westergren method may be needed.
topic corrected esr
hematocrit
fabry’s formula
anemia
url http://www.thieme-connect.de/DOI/DOI?10.1055/s-0040-1721155
work_keys_str_mv AT vikramnarang comparativeanalysisoferythrocytesedimentationratemeasuredbyautomatedandmanualmethodsinanaemicpatients
AT sumitgrover comparativeanalysisoferythrocytesedimentationratemeasuredbyautomatedandmanualmethodsinanaemicpatients
AT amandeepkaurkang comparativeanalysisoferythrocytesedimentationratemeasuredbyautomatedandmanualmethodsinanaemicpatients
AT avantikagarg comparativeanalysisoferythrocytesedimentationratemeasuredbyautomatedandmanualmethodsinanaemicpatients
AT neenasood comparativeanalysisoferythrocytesedimentationratemeasuredbyautomatedandmanualmethodsinanaemicpatients
_version_ 1724365677364510720