Evaluation of self-prepared test kit for outdoor dissolved oxygen determination

A self-prepared test kit for outdoor dissolved oxygen determination was prepared and evaluated against the azide modification Winkler method. Three hundred and five water samples with 3 replicates were taken from ditches at Prince of Songkla University, Pattani campus. The result showed that the 2 m...

Full description

Bibliographic Details
Main Authors: Luangthuvapranit, C., Srichai, N.
Format: Article
Language:English
Published: Prince of Songkla University 2007-01-01
Series:Songklanakarin Journal of Science and Technology (SJST)
Subjects:
Online Access:http://www.sjst.psu.ac.th/journal/29_1_pdf/09Chokchai_101-108.pdf
Description
Summary:A self-prepared test kit for outdoor dissolved oxygen determination was prepared and evaluated against the azide modification Winkler method. Three hundred and five water samples with 3 replicates were taken from ditches at Prince of Songkla University, Pattani campus. The result showed that the 2 methods had good correlation (r = 0.999, p < 0.00) with linear regression Y = 0.9992X + 0.0111 (p< 0.00). When paired t-test was performed, there was no statistically significant difference at p>0.05. Thus, the compact self-prepared test kit, based on azide modification Winkler method, was suitable for quick outdoor determination of dissolved oxygen and cost only one-eighth of the cost of the azide modification method.