Accurate Coregistration between Ultra-High-Resolution Micro-SPECT and Circular Cone-Beam Micro-CT Scanners
Introduction. Spatially registering SPECT with CT makes it possible to anatomically localize SPECT tracers. In this study, an accurate method for the coregistration of ultra-high-resolution SPECT volumes and multiple cone-beam CT volumes is developed and validated, which does not require markers dur...
Main Authors: | , , , , |
---|---|
Format: | Article |
Language: | English |
Published: |
Hindawi Limited
2010-01-01
|
Series: | International Journal of Biomedical Imaging |
Online Access: | http://dx.doi.org/10.1155/2010/654506 |
id |
doaj-af9bd044f574487e930fd29f2fed0488 |
---|---|
record_format |
Article |
spelling |
doaj-af9bd044f574487e930fd29f2fed04882020-11-24T22:35:43ZengHindawi LimitedInternational Journal of Biomedical Imaging1687-41881687-41962010-01-01201010.1155/2010/654506654506Accurate Coregistration between Ultra-High-Resolution Micro-SPECT and Circular Cone-Beam Micro-CT ScannersChangguo Ji0Frans van der Have1Hugo Gratama van Andel2Ruud Ramakers3Freek Beekman4Image Sciences Institute and Rudolf Magnus Institute, University Medical Center Utrecht, Universiteitsweg 100, 3584 CG Utrecht, The NetherlandsImage Sciences Institute and Rudolf Magnus Institute, University Medical Center Utrecht, Universiteitsweg 100, 3584 CG Utrecht, The NetherlandsDepartment of Radiation, Radionuclides, and Reactors, Section Radiation Detection and Medical imaging, Delft University of Technology, Mekelweg 15, 2629 JB Delft, The NetherlandsImage Sciences Institute and Rudolf Magnus Institute, University Medical Center Utrecht, Universiteitsweg 100, 3584 CG Utrecht, The NetherlandsImage Sciences Institute and Rudolf Magnus Institute, University Medical Center Utrecht, Universiteitsweg 100, 3584 CG Utrecht, The NetherlandsIntroduction. Spatially registering SPECT with CT makes it possible to anatomically localize SPECT tracers. In this study, an accurate method for the coregistration of ultra-high-resolution SPECT volumes and multiple cone-beam CT volumes is developed and validated, which does not require markers during animal scanning. Methods. Transferable animal beds were developed with an accurate mounting interface. Simple calibration phantoms make it possible to obtain both the spatial transformation matrix for stitching multiple CT scans of different parts of the animal and to register SPECT and CT. The spatial transformation for image coregistration is calculated once using Horn's matching algorithm. Animal images can then be coregistered without using markers. Results. For mouse-sized objects, average coregistration errors between SPECT and CT in X, Y, and Z directions are within 0.04 mm, 0.10 mm, and 0.19 mm, respectively. For rat-sized objects, these numbers are 0.22 mm, 0.14 mm, and 0.28 mm. Average 3D coregistration errors were within 0.24 mm and 0.42 mm for mouse and rat imaging, respectively. Conclusion. Extending the field-of-view of cone-beam CT by stitching is improved by prior registration of the CT volumes. The accuracy of registration between SPECT and CT is typically better than the image resolution of current ultra-high-resolution SPECT.http://dx.doi.org/10.1155/2010/654506 |
collection |
DOAJ |
language |
English |
format |
Article |
sources |
DOAJ |
author |
Changguo Ji Frans van der Have Hugo Gratama van Andel Ruud Ramakers Freek Beekman |
spellingShingle |
Changguo Ji Frans van der Have Hugo Gratama van Andel Ruud Ramakers Freek Beekman Accurate Coregistration between Ultra-High-Resolution Micro-SPECT and Circular Cone-Beam Micro-CT Scanners International Journal of Biomedical Imaging |
author_facet |
Changguo Ji Frans van der Have Hugo Gratama van Andel Ruud Ramakers Freek Beekman |
author_sort |
Changguo Ji |
title |
Accurate Coregistration between Ultra-High-Resolution Micro-SPECT and Circular Cone-Beam Micro-CT Scanners |
title_short |
Accurate Coregistration between Ultra-High-Resolution Micro-SPECT and Circular Cone-Beam Micro-CT Scanners |
title_full |
Accurate Coregistration between Ultra-High-Resolution Micro-SPECT and Circular Cone-Beam Micro-CT Scanners |
title_fullStr |
Accurate Coregistration between Ultra-High-Resolution Micro-SPECT and Circular Cone-Beam Micro-CT Scanners |
title_full_unstemmed |
Accurate Coregistration between Ultra-High-Resolution Micro-SPECT and Circular Cone-Beam Micro-CT Scanners |
title_sort |
accurate coregistration between ultra-high-resolution micro-spect and circular cone-beam micro-ct scanners |
publisher |
Hindawi Limited |
series |
International Journal of Biomedical Imaging |
issn |
1687-4188 1687-4196 |
publishDate |
2010-01-01 |
description |
Introduction. Spatially registering SPECT with CT makes it possible to anatomically localize SPECT tracers. In this study, an accurate method for the coregistration of ultra-high-resolution SPECT volumes and multiple cone-beam CT volumes is developed and validated, which does not require markers during animal scanning.
Methods. Transferable animal beds were developed with an accurate mounting interface. Simple calibration phantoms make it possible to obtain both the spatial transformation matrix for stitching multiple CT scans of different parts of the animal and to register SPECT and CT. The spatial transformation for image coregistration is calculated once using Horn's matching algorithm. Animal images can then be coregistered without using markers.
Results. For mouse-sized objects, average coregistration errors between SPECT and CT in X, Y, and Z directions are within 0.04 mm, 0.10 mm, and 0.19 mm, respectively. For rat-sized objects, these numbers are 0.22 mm, 0.14 mm, and 0.28 mm. Average 3D coregistration errors were within 0.24 mm and 0.42 mm for mouse and rat imaging, respectively.
Conclusion. Extending the field-of-view of cone-beam CT by stitching is improved by prior registration of the CT volumes. The accuracy of registration between SPECT and CT is typically better than the image resolution of current ultra-high-resolution SPECT. |
url |
http://dx.doi.org/10.1155/2010/654506 |
work_keys_str_mv |
AT changguoji accuratecoregistrationbetweenultrahighresolutionmicrospectandcircularconebeammicroctscanners AT fransvanderhave accuratecoregistrationbetweenultrahighresolutionmicrospectandcircularconebeammicroctscanners AT hugogratamavanandel accuratecoregistrationbetweenultrahighresolutionmicrospectandcircularconebeammicroctscanners AT ruudramakers accuratecoregistrationbetweenultrahighresolutionmicrospectandcircularconebeammicroctscanners AT freekbeekman accuratecoregistrationbetweenultrahighresolutionmicrospectandcircularconebeammicroctscanners |
_version_ |
1725723050014408704 |