Effective Assessment of Diabetes Control Using Personal Glucometers (CONTOURLINK, Bayer, Germany; CALLA, Wellion, Austria; LINUS, Agamatrix, USA)

Aim of this trial was to assess (1) the accuracy and precision of electrochemistry-based glucometers CONTOURLINK, CALLA, and LINUS and (2) the diabetes control using Ambulatory Glycaemic Profiles (AGP) as markers of therapeutic effectiveness. Glucometers and COBAS INTEGRA 400 Plus analyzer were used...

Full description

Bibliographic Details
Main Authors: R. Chlup, B. Doubravova, J. Bartek, J. Zapletalova, O. Krystynik, V. Prochazka
Format: Article
Language:English
Published: Hindawi Limited 2013-01-01
Series:Disease Markers
Online Access:http://dx.doi.org/10.1155/2013/702427
id doaj-aceab4c4646746bf974bd13ce24b002d
record_format Article
spelling doaj-aceab4c4646746bf974bd13ce24b002d2020-11-24T23:58:52ZengHindawi LimitedDisease Markers0278-02401875-86302013-01-0135689590510.1155/2013/702427702427Effective Assessment of Diabetes Control Using Personal Glucometers (CONTOURLINK, Bayer, Germany; CALLA, Wellion, Austria; LINUS, Agamatrix, USA)R. Chlup0B. Doubravova1J. Bartek2J. Zapletalova3O. Krystynik4V. Prochazka5Department of Physiology, Faculty of Medicine and Dentistry, Palacky University Olomouc, 779 00 Olomouc, Czech RepublicDepartment of Biochemistry, Institute of Neurology and Geriatrics Moravsky Beroun, 793 05 Moravsky Beroun, Czech RepublicDepartment of Medical Chemistry and Biochemistry, Faculty of Medicine and Dentistry, Palacky University Olomouc, 779 00 Olomouc, Czech RepublicDepartment of Medical Biophysics, Faculty of Medicine and Dentistry, Palacky University Olomouc, 779 00 Olomouc, Czech Republic2nd Department of Medicine, Gastroenterology and Hepatology, Teaching Hospital Olomouc, 775 20 Olomouc, Czech Republic2nd Department of Medicine, Gastroenterology and Hepatology, Teaching Hospital Olomouc, 775 20 Olomouc, Czech RepublicAim of this trial was to assess (1) the accuracy and precision of electrochemistry-based glucometers CONTOURLINK, CALLA, and LINUS and (2) the diabetes control using Ambulatory Glycaemic Profiles (AGP) as markers of therapeutic effectiveness. Glucometers and COBAS INTEGRA 400 Plus analyzer were used by one laboratory professional to estimate P-glucose (PG) in 112 out-patients. There were 112 sets of 12 PG estimations analyzed. In each set, means of 3 capillary PG estimations on 3 respective glucometers and on INTEGRA analyzer were calculated. The statistical program SPSS, v. 15, was applied. The mean INTEGRA PG values ranged from 2.7 to 25.3 mmol/L. There were strong correlations between mean PG on INTEGRA versus CONTOURLINK, versus CALLA, and versus LINUS; PG deviations from INTEGRA were mostly within the range ±15%. Wilcoxon Signed Rank Test revealed differences between CONTOURLINK-INTEGRA and LINUS-INTEGRA; CALLA-INTEGRA showed no difference. SD INTEGRA = 0.061 mmol/L, SD CONTOURLINK/SD CALLA/SD LINUS were 0.256/0.290/0.286 mmol/L. All patients were trained to perform defined 10-point PG profiles to adapt food, exercise, and insulin doses. The PG differences between all tested glucometers and reference values were in borderline of ISO 15197 but worthy of consideration. AGP are helpful markers of diabetes control.http://dx.doi.org/10.1155/2013/702427
collection DOAJ
language English
format Article
sources DOAJ
author R. Chlup
B. Doubravova
J. Bartek
J. Zapletalova
O. Krystynik
V. Prochazka
spellingShingle R. Chlup
B. Doubravova
J. Bartek
J. Zapletalova
O. Krystynik
V. Prochazka
Effective Assessment of Diabetes Control Using Personal Glucometers (CONTOURLINK, Bayer, Germany; CALLA, Wellion, Austria; LINUS, Agamatrix, USA)
Disease Markers
author_facet R. Chlup
B. Doubravova
J. Bartek
J. Zapletalova
O. Krystynik
V. Prochazka
author_sort R. Chlup
title Effective Assessment of Diabetes Control Using Personal Glucometers (CONTOURLINK, Bayer, Germany; CALLA, Wellion, Austria; LINUS, Agamatrix, USA)
title_short Effective Assessment of Diabetes Control Using Personal Glucometers (CONTOURLINK, Bayer, Germany; CALLA, Wellion, Austria; LINUS, Agamatrix, USA)
title_full Effective Assessment of Diabetes Control Using Personal Glucometers (CONTOURLINK, Bayer, Germany; CALLA, Wellion, Austria; LINUS, Agamatrix, USA)
title_fullStr Effective Assessment of Diabetes Control Using Personal Glucometers (CONTOURLINK, Bayer, Germany; CALLA, Wellion, Austria; LINUS, Agamatrix, USA)
title_full_unstemmed Effective Assessment of Diabetes Control Using Personal Glucometers (CONTOURLINK, Bayer, Germany; CALLA, Wellion, Austria; LINUS, Agamatrix, USA)
title_sort effective assessment of diabetes control using personal glucometers (contourlink, bayer, germany; calla, wellion, austria; linus, agamatrix, usa)
publisher Hindawi Limited
series Disease Markers
issn 0278-0240
1875-8630
publishDate 2013-01-01
description Aim of this trial was to assess (1) the accuracy and precision of electrochemistry-based glucometers CONTOURLINK, CALLA, and LINUS and (2) the diabetes control using Ambulatory Glycaemic Profiles (AGP) as markers of therapeutic effectiveness. Glucometers and COBAS INTEGRA 400 Plus analyzer were used by one laboratory professional to estimate P-glucose (PG) in 112 out-patients. There were 112 sets of 12 PG estimations analyzed. In each set, means of 3 capillary PG estimations on 3 respective glucometers and on INTEGRA analyzer were calculated. The statistical program SPSS, v. 15, was applied. The mean INTEGRA PG values ranged from 2.7 to 25.3 mmol/L. There were strong correlations between mean PG on INTEGRA versus CONTOURLINK, versus CALLA, and versus LINUS; PG deviations from INTEGRA were mostly within the range ±15%. Wilcoxon Signed Rank Test revealed differences between CONTOURLINK-INTEGRA and LINUS-INTEGRA; CALLA-INTEGRA showed no difference. SD INTEGRA = 0.061 mmol/L, SD CONTOURLINK/SD CALLA/SD LINUS were 0.256/0.290/0.286 mmol/L. All patients were trained to perform defined 10-point PG profiles to adapt food, exercise, and insulin doses. The PG differences between all tested glucometers and reference values were in borderline of ISO 15197 but worthy of consideration. AGP are helpful markers of diabetes control.
url http://dx.doi.org/10.1155/2013/702427
work_keys_str_mv AT rchlup effectiveassessmentofdiabetescontrolusingpersonalglucometerscontourlinkbayergermanycallawellionaustrialinusagamatrixusa
AT bdoubravova effectiveassessmentofdiabetescontrolusingpersonalglucometerscontourlinkbayergermanycallawellionaustrialinusagamatrixusa
AT jbartek effectiveassessmentofdiabetescontrolusingpersonalglucometerscontourlinkbayergermanycallawellionaustrialinusagamatrixusa
AT jzapletalova effectiveassessmentofdiabetescontrolusingpersonalglucometerscontourlinkbayergermanycallawellionaustrialinusagamatrixusa
AT okrystynik effectiveassessmentofdiabetescontrolusingpersonalglucometerscontourlinkbayergermanycallawellionaustrialinusagamatrixusa
AT vprochazka effectiveassessmentofdiabetescontrolusingpersonalglucometerscontourlinkbayergermanycallawellionaustrialinusagamatrixusa
_version_ 1725449349461180416