Summary: | This paper describes the evaluation of simple and fast solubilization methods for the determination of Ca, Mg, and K in glycerin samples from biodiesel production by atomic spectrometry. The solubilization in water was compared with two other methods: solubilization in formic acid and solubilization in ethanol. Using solubilization in water, determination of the three analytes was possible; the values of limits of detection for Ca, K, Mg were 0.31, 0.06, and 0.16 mg kg−1, respectively. Because no adequate reference material was available, the accuracy was evaluated by assessing the recoveries tests with both solubilization methods; the evaluation ranged from 90% to 115%, with values of relative standard deviation >8%, indicating good accuracy of the measure. Four crude glycerin samples obtained from biodiesel plants of Rio Grande do Sul were analyzed after treatment with the different methods of solubilization, and the obtained results of Ca, Mg, and K concentration were in agreement with the values obtained from both solubilization methods. Therefore, solubilization in water is concluded to be a simpler, faster, and viable method for sample preparation of glycerin.
|