Comparison of corneal power obtained from VERION image-guided surgery system and four other devices
Hung-Yuan Lin,1,* Hsin-Yang Chen,1,2,* Han Bor Fam,3 Ya-Jung Chuang,1 Ronald Yeoh,4 Pi-Jung Lin5 1Universal Eye Center, Zhongli Branch, Zhongli County, TaoYuan City, Taiwan, Republic of China; 2Ophthalmology Department, Ningbo First Hospital, Ningbo, Zhejiang Province, People’s Republic o...
Main Authors: | , , , , , |
---|---|
Format: | Article |
Language: | English |
Published: |
Dove Medical Press
2017-07-01
|
Series: | Clinical Ophthalmology |
Subjects: | |
Online Access: | https://www.dovepress.com/comparison-of-corneal-power-obtained-from-verion-image-guided-surgery--peer-reviewed-article-OPTH |
id |
doaj-a60c9bd227294ccebe9865211fe040d3 |
---|---|
record_format |
Article |
spelling |
doaj-a60c9bd227294ccebe9865211fe040d32020-11-24T23:34:29ZengDove Medical PressClinical Ophthalmology1177-54832017-07-01Volume 111291129933717Comparison of corneal power obtained from VERION image-guided surgery system and four other devicesLin HYChen HYFam HBChuang YJYeoh RLin PJHung-Yuan Lin,1,* Hsin-Yang Chen,1,2,* Han Bor Fam,3 Ya-Jung Chuang,1 Ronald Yeoh,4 Pi-Jung Lin5 1Universal Eye Center, Zhongli Branch, Zhongli County, TaoYuan City, Taiwan, Republic of China; 2Ophthalmology Department, Ningbo First Hospital, Ningbo, Zhejiang Province, People’s Republic of China; 3Ophthalmology Department, Tan Tock Seng Hospital, Jalan Tan Tock Seng, 4Ophthalmology Department, Eye and Retina Surgeons, Camden Medical Centre, Singapore; 5Universal Eye Center, Xinnan Branch, Taipei City, Taiwan, Republic of China *These authors contributed equally to this work Purpose: To assess the corneal keratometric values obtained using the VERION image-guided surgery system and other devices.Methods: This study evaluated the right eyes of 115 cataract patients before intraocular lens (IOL) implantation through consecutive tests using 5 devices: VERION Reference Unit , Placido-based corneal topography (OPD-Scan III), monochromatic light-emitting diodes (LenStar LS900 and AL-Scan), and rotary prism technology (auto kerato-refractometer KR-8800). Analyzed parameters were corneal steep and flat keratometric values (Ks and Kf) and corneal astigmatism and axis. These parameters were evaluated using the one-sample two-tailed t-test and the 95% limits of agreement (95% LOAs) between the devices.Results: The mean corneal cylinder value measurements were -0.97±0.63 D, -0.88± 0.60 D, -0.90±0.69 D, -0.90±0.67 D, and -0.83±0.60 D with VERION, LenStar, AL-Scan (2.4 mm), OPD III, and KR-8800, respectively. Only KR-8800 showed a significant difference from VERION in the corneal cylinder value (P<0.05). The mean differences in the Kf and Ks of VERION compared to those of OPD III were 0.18±0.45 D and 0.17±0.38 D (P<0.05), respectively. The 95% LOAs of Bland–Altman analysis for the corneal astigmatism axis of the VERION with LenStar, AL-Scan (2.4 mm), OPD III, and KR-8800 were -26.25° to 58.71°, -20.61° to 47.44°, -25.03° to 58.98°, and -27.85° to 65.17°, respectively.Conclusion: None of the VERION parameters were significantly different from those of AL-Scan and LenStar. AL-Scan (2.4 mm zone) was especially similar to VERION. Wide LOAs are potential contributors to axis error in patients with toric IOL implants. Keywords: phacoemulsification, toric IOL, corneal astigmatism, surgical induced astigmatism, image-guided surgery systemhttps://www.dovepress.com/comparison-of-corneal-power-obtained-from-verion-image-guided-surgery--peer-reviewed-article-OPTHphacoemulsificationtoric IOLcorneal astigmatismsurgical induced astigmatismimage-guided Surgery System |
collection |
DOAJ |
language |
English |
format |
Article |
sources |
DOAJ |
author |
Lin HY Chen HY Fam HB Chuang YJ Yeoh R Lin PJ |
spellingShingle |
Lin HY Chen HY Fam HB Chuang YJ Yeoh R Lin PJ Comparison of corneal power obtained from VERION image-guided surgery system and four other devices Clinical Ophthalmology phacoemulsification toric IOL corneal astigmatism surgical induced astigmatism image-guided Surgery System |
author_facet |
Lin HY Chen HY Fam HB Chuang YJ Yeoh R Lin PJ |
author_sort |
Lin HY |
title |
Comparison of corneal power obtained from VERION image-guided surgery system and four other devices |
title_short |
Comparison of corneal power obtained from VERION image-guided surgery system and four other devices |
title_full |
Comparison of corneal power obtained from VERION image-guided surgery system and four other devices |
title_fullStr |
Comparison of corneal power obtained from VERION image-guided surgery system and four other devices |
title_full_unstemmed |
Comparison of corneal power obtained from VERION image-guided surgery system and four other devices |
title_sort |
comparison of corneal power obtained from verion image-guided surgery system and four other devices |
publisher |
Dove Medical Press |
series |
Clinical Ophthalmology |
issn |
1177-5483 |
publishDate |
2017-07-01 |
description |
Hung-Yuan Lin,1,* Hsin-Yang Chen,1,2,* Han Bor Fam,3 Ya-Jung Chuang,1 Ronald Yeoh,4 Pi-Jung Lin5 1Universal Eye Center, Zhongli Branch, Zhongli County, TaoYuan City, Taiwan, Republic of China; 2Ophthalmology Department, Ningbo First Hospital, Ningbo, Zhejiang Province, People’s Republic of China; 3Ophthalmology Department, Tan Tock Seng Hospital, Jalan Tan Tock Seng, 4Ophthalmology Department, Eye and Retina Surgeons, Camden Medical Centre, Singapore; 5Universal Eye Center, Xinnan Branch, Taipei City, Taiwan, Republic of China *These authors contributed equally to this work Purpose: To assess the corneal keratometric values obtained using the VERION image-guided surgery system and other devices.Methods: This study evaluated the right eyes of 115 cataract patients before intraocular lens (IOL) implantation through consecutive tests using 5 devices: VERION Reference Unit , Placido-based corneal topography (OPD-Scan III), monochromatic light-emitting diodes (LenStar LS900 and AL-Scan), and rotary prism technology (auto kerato-refractometer KR-8800). Analyzed parameters were corneal steep and flat keratometric values (Ks and Kf) and corneal astigmatism and axis. These parameters were evaluated using the one-sample two-tailed t-test and the 95% limits of agreement (95% LOAs) between the devices.Results: The mean corneal cylinder value measurements were -0.97±0.63 D, -0.88± 0.60 D, -0.90±0.69 D, -0.90±0.67 D, and -0.83±0.60 D with VERION, LenStar, AL-Scan (2.4 mm), OPD III, and KR-8800, respectively. Only KR-8800 showed a significant difference from VERION in the corneal cylinder value (P<0.05). The mean differences in the Kf and Ks of VERION compared to those of OPD III were 0.18±0.45 D and 0.17±0.38 D (P<0.05), respectively. The 95% LOAs of Bland–Altman analysis for the corneal astigmatism axis of the VERION with LenStar, AL-Scan (2.4 mm), OPD III, and KR-8800 were -26.25° to 58.71°, -20.61° to 47.44°, -25.03° to 58.98°, and -27.85° to 65.17°, respectively.Conclusion: None of the VERION parameters were significantly different from those of AL-Scan and LenStar. AL-Scan (2.4 mm zone) was especially similar to VERION. Wide LOAs are potential contributors to axis error in patients with toric IOL implants. Keywords: phacoemulsification, toric IOL, corneal astigmatism, surgical induced astigmatism, image-guided surgery system |
topic |
phacoemulsification toric IOL corneal astigmatism surgical induced astigmatism image-guided Surgery System |
url |
https://www.dovepress.com/comparison-of-corneal-power-obtained-from-verion-image-guided-surgery--peer-reviewed-article-OPTH |
work_keys_str_mv |
AT linhy comparisonofcornealpowerobtainedfromverionimageguidedsurgerysystemandfourotherdevices AT chenhy comparisonofcornealpowerobtainedfromverionimageguidedsurgerysystemandfourotherdevices AT famhb comparisonofcornealpowerobtainedfromverionimageguidedsurgerysystemandfourotherdevices AT chuangyj comparisonofcornealpowerobtainedfromverionimageguidedsurgerysystemandfourotherdevices AT yeohr comparisonofcornealpowerobtainedfromverionimageguidedsurgerysystemandfourotherdevices AT linpj comparisonofcornealpowerobtainedfromverionimageguidedsurgerysystemandfourotherdevices |
_version_ |
1725529216087228416 |