Reproducibility of pulmonary blood flow measurements by phase-contrast MRI using different 1.5 T MR scanners at two institutions
Background Magnetic resonance imaging (MRI) can be beneficial for diagnosis of disease by offering quantitative information. However, reproducibility can be a major problem when there is a numerical threshold in multi-institution, multi-vendor situations. Purpose To measure pulmonary blood flow with...
Main Authors: | , , , , |
---|---|
Format: | Article |
Language: | English |
Published: |
SAGE Publishing
2017-01-01
|
Series: | Acta Radiologica Open |
Online Access: | http://arr.sagepub.com/content/6/1/2058460116684370.full.pdf |
id |
doaj-a208a71f24f3464b902fd554ccf94ee9 |
---|---|
record_format |
Article |
spelling |
doaj-a208a71f24f3464b902fd554ccf94ee92020-11-25T03:38:47ZengSAGE PublishingActa Radiologica Open2058-46012017-01-016110.1177/205846011668437010.1177_2058460116684370Reproducibility of pulmonary blood flow measurements by phase-contrast MRI using different 1.5 T MR scanners at two institutionsRin IrahaNanae TsuchiyaTsuneo YamashiroTae IwasawaSadayuki MurayamaBackground Magnetic resonance imaging (MRI) can be beneficial for diagnosis of disease by offering quantitative information. However, reproducibility can be a major problem when there is a numerical threshold in multi-institution, multi-vendor situations. Purpose To measure pulmonary blood flow with phase-contrast (PC) imaging using two different MR scanners (1.5 T) at different institutions in the same participants and to examine the reproducibility of the measurements. Material and Methods Participants were 10 healthy volunteers (5 men; age range, 27–36 years). The measurements included the mean and maximal blood velocities, the mean blood flow volume, and the acceleration time and volume (AT and AV), derived from the time-flow curve of the PC-MRI. Simultaneously obtained maximal, minimal, and mean areas from regions of interest set in the pulmonary artery were also calculated. In order to calculate the reproducibility of the quantitative variables, intra-class correlation coefficients (ICCs) were employed. When an adequate ICC was obtained, Bland–Altman analysis was conducted to identify any systematic bias. Results The ICCs were almost perfect for the mean blood flow volume and the AV (r = 0.82 and 0.80), and were substantial in the mean and maximal areas, and the AT (r = 0.63, 0.74, and 0.64, respectively). However, there was a fixed bias in the area measurement between the two scanners. Also, the AV had a proportional bias. Conclusion Our results reveal that various indices derived from PC-MRI on different MR scanners are promising as common indices for pulmonary flow assessment. Research and clinical use of PC-MRI for the pulmonary artery is expected to extend to multi-institution situations.http://arr.sagepub.com/content/6/1/2058460116684370.full.pdf |
collection |
DOAJ |
language |
English |
format |
Article |
sources |
DOAJ |
author |
Rin Iraha Nanae Tsuchiya Tsuneo Yamashiro Tae Iwasawa Sadayuki Murayama |
spellingShingle |
Rin Iraha Nanae Tsuchiya Tsuneo Yamashiro Tae Iwasawa Sadayuki Murayama Reproducibility of pulmonary blood flow measurements by phase-contrast MRI using different 1.5 T MR scanners at two institutions Acta Radiologica Open |
author_facet |
Rin Iraha Nanae Tsuchiya Tsuneo Yamashiro Tae Iwasawa Sadayuki Murayama |
author_sort |
Rin Iraha |
title |
Reproducibility of pulmonary blood flow measurements by phase-contrast MRI using different 1.5 T MR scanners at two institutions |
title_short |
Reproducibility of pulmonary blood flow measurements by phase-contrast MRI using different 1.5 T MR scanners at two institutions |
title_full |
Reproducibility of pulmonary blood flow measurements by phase-contrast MRI using different 1.5 T MR scanners at two institutions |
title_fullStr |
Reproducibility of pulmonary blood flow measurements by phase-contrast MRI using different 1.5 T MR scanners at two institutions |
title_full_unstemmed |
Reproducibility of pulmonary blood flow measurements by phase-contrast MRI using different 1.5 T MR scanners at two institutions |
title_sort |
reproducibility of pulmonary blood flow measurements by phase-contrast mri using different 1.5 t mr scanners at two institutions |
publisher |
SAGE Publishing |
series |
Acta Radiologica Open |
issn |
2058-4601 |
publishDate |
2017-01-01 |
description |
Background Magnetic resonance imaging (MRI) can be beneficial for diagnosis of disease by offering quantitative information. However, reproducibility can be a major problem when there is a numerical threshold in multi-institution, multi-vendor situations. Purpose To measure pulmonary blood flow with phase-contrast (PC) imaging using two different MR scanners (1.5 T) at different institutions in the same participants and to examine the reproducibility of the measurements. Material and Methods Participants were 10 healthy volunteers (5 men; age range, 27–36 years). The measurements included the mean and maximal blood velocities, the mean blood flow volume, and the acceleration time and volume (AT and AV), derived from the time-flow curve of the PC-MRI. Simultaneously obtained maximal, minimal, and mean areas from regions of interest set in the pulmonary artery were also calculated. In order to calculate the reproducibility of the quantitative variables, intra-class correlation coefficients (ICCs) were employed. When an adequate ICC was obtained, Bland–Altman analysis was conducted to identify any systematic bias. Results The ICCs were almost perfect for the mean blood flow volume and the AV (r = 0.82 and 0.80), and were substantial in the mean and maximal areas, and the AT (r = 0.63, 0.74, and 0.64, respectively). However, there was a fixed bias in the area measurement between the two scanners. Also, the AV had a proportional bias. Conclusion Our results reveal that various indices derived from PC-MRI on different MR scanners are promising as common indices for pulmonary flow assessment. Research and clinical use of PC-MRI for the pulmonary artery is expected to extend to multi-institution situations. |
url |
http://arr.sagepub.com/content/6/1/2058460116684370.full.pdf |
work_keys_str_mv |
AT riniraha reproducibilityofpulmonarybloodflowmeasurementsbyphasecontrastmriusingdifferent15tmrscannersattwoinstitutions AT nanaetsuchiya reproducibilityofpulmonarybloodflowmeasurementsbyphasecontrastmriusingdifferent15tmrscannersattwoinstitutions AT tsuneoyamashiro reproducibilityofpulmonarybloodflowmeasurementsbyphasecontrastmriusingdifferent15tmrscannersattwoinstitutions AT taeiwasawa reproducibilityofpulmonarybloodflowmeasurementsbyphasecontrastmriusingdifferent15tmrscannersattwoinstitutions AT sadayukimurayama reproducibilityofpulmonarybloodflowmeasurementsbyphasecontrastmriusingdifferent15tmrscannersattwoinstitutions |
_version_ |
1724540531725303808 |