Calculation of Inter- and Intra-Fraction Motion Errors at External Radiotherapy Using a Markerless Strategy Based on Image Registration Combined with Correlation Model

Introduction: A new method based on image registration technique and an intelligent correlation model to calculate. The present study aimed to propose inter- and intra-fraction motion errors in order to address the limitations of conventional Patient positioning methods. Material and Methods: The co...

Full description

Bibliographic Details
Main Authors: Payam Samadi Miandoab, Ahmad Esmaili Torshabi, Sohelia parandeh
Format: Article
Language:English
Published: Mashhad University of Medical Sciences 2019-05-01
Series:Iranian Journal of Medical Physics
Subjects:
Online Access:http://ijmp.mums.ac.ir/article_11054_3f137c26b93885b85ad54146b82b6790.pdf
id doaj-a01576f4618d4bdfb03983814683d818
record_format Article
spelling doaj-a01576f4618d4bdfb03983814683d8182020-11-25T00:36:59ZengMashhad University of Medical SciencesIranian Journal of Medical Physics2345-36722345-36722019-05-0116322423110.22038/ijmp.2018.30477.134811054Calculation of Inter- and Intra-Fraction Motion Errors at External Radiotherapy Using a Markerless Strategy Based on Image Registration Combined with Correlation ModelPayam Samadi Miandoab0Ahmad Esmaili Torshabi1Sohelia parandeh2Department of Electrical and Computer Engineering, Medical Radiation Group, Graduate University of Advanced Technology, Haft Bagh Highway, Knowledge Paradise, Kerman, Iran.Medical Radiation Division, Dept. of Electrical and Computer Eng. Graduate University of Advanced Technology, Kerman, IranDepartment of Electrical and Computer Engineering, Medical Radiation Group, Graduate University of Advanced Technology, Haft Bagh Highway, Knowledge Paradise, Kerman, Iran.Introduction: A new method based on image registration technique and an intelligent correlation model to calculate. The present study aimed to propose inter- and intra-fraction motion errors in order to address the limitations of conventional Patient positioning methods. Material and Methods: The configuration of the markerless method was accomplished by using four-dimensional computed tomography (4DCT) datasets. Firstly, the MeVisLab software package was used to extract a three-dimensional (3D) surface model of the patient and determine the tumor location. Then, the patient-specific 3D surface model which also included the breathing phases was imported into the MATLAB software package in order to define several control points on the thorax region as virtual external markers. Finally, based on the correlation of breathing signals/patient position with breathing signals/tumor coordinate, an adaptive neuro fuzzy inference system was proposed to both verify and align the inter- and intra-fraction motion errors in radiotherapy, if needed. In order to validate the proposed method, the 4DCT data acquired from four real patients was considered. Results: Final results revealed that our hybrid configuration method was capable of aligning patient setup with lower uncertainties, compared to other available methods. In addition, the 3D root-mean-square error has been reduced from 5.26 to 1.5 mm for all patients. Conclusion: In this study, a markerless method based on the image registration technique in combination with a correlation model was proposed to address the limitations of the available methods, including dependence on operator’s attention, use of passive markers, and rigid-only constraint for patient setup.http://ijmp.mums.ac.ir/article_11054_3f137c26b93885b85ad54146b82b6790.pdfImage ProcessingImage Guided Radiation Therapy (IGRT)patient positioning
collection DOAJ
language English
format Article
sources DOAJ
author Payam Samadi Miandoab
Ahmad Esmaili Torshabi
Sohelia parandeh
spellingShingle Payam Samadi Miandoab
Ahmad Esmaili Torshabi
Sohelia parandeh
Calculation of Inter- and Intra-Fraction Motion Errors at External Radiotherapy Using a Markerless Strategy Based on Image Registration Combined with Correlation Model
Iranian Journal of Medical Physics
Image Processing
Image Guided Radiation Therapy (IGRT)
patient positioning
author_facet Payam Samadi Miandoab
Ahmad Esmaili Torshabi
Sohelia parandeh
author_sort Payam Samadi Miandoab
title Calculation of Inter- and Intra-Fraction Motion Errors at External Radiotherapy Using a Markerless Strategy Based on Image Registration Combined with Correlation Model
title_short Calculation of Inter- and Intra-Fraction Motion Errors at External Radiotherapy Using a Markerless Strategy Based on Image Registration Combined with Correlation Model
title_full Calculation of Inter- and Intra-Fraction Motion Errors at External Radiotherapy Using a Markerless Strategy Based on Image Registration Combined with Correlation Model
title_fullStr Calculation of Inter- and Intra-Fraction Motion Errors at External Radiotherapy Using a Markerless Strategy Based on Image Registration Combined with Correlation Model
title_full_unstemmed Calculation of Inter- and Intra-Fraction Motion Errors at External Radiotherapy Using a Markerless Strategy Based on Image Registration Combined with Correlation Model
title_sort calculation of inter- and intra-fraction motion errors at external radiotherapy using a markerless strategy based on image registration combined with correlation model
publisher Mashhad University of Medical Sciences
series Iranian Journal of Medical Physics
issn 2345-3672
2345-3672
publishDate 2019-05-01
description Introduction: A new method based on image registration technique and an intelligent correlation model to calculate. The present study aimed to propose inter- and intra-fraction motion errors in order to address the limitations of conventional Patient positioning methods. Material and Methods: The configuration of the markerless method was accomplished by using four-dimensional computed tomography (4DCT) datasets. Firstly, the MeVisLab software package was used to extract a three-dimensional (3D) surface model of the patient and determine the tumor location. Then, the patient-specific 3D surface model which also included the breathing phases was imported into the MATLAB software package in order to define several control points on the thorax region as virtual external markers. Finally, based on the correlation of breathing signals/patient position with breathing signals/tumor coordinate, an adaptive neuro fuzzy inference system was proposed to both verify and align the inter- and intra-fraction motion errors in radiotherapy, if needed. In order to validate the proposed method, the 4DCT data acquired from four real patients was considered. Results: Final results revealed that our hybrid configuration method was capable of aligning patient setup with lower uncertainties, compared to other available methods. In addition, the 3D root-mean-square error has been reduced from 5.26 to 1.5 mm for all patients. Conclusion: In this study, a markerless method based on the image registration technique in combination with a correlation model was proposed to address the limitations of the available methods, including dependence on operator’s attention, use of passive markers, and rigid-only constraint for patient setup.
topic Image Processing
Image Guided Radiation Therapy (IGRT)
patient positioning
url http://ijmp.mums.ac.ir/article_11054_3f137c26b93885b85ad54146b82b6790.pdf
work_keys_str_mv AT payamsamadimiandoab calculationofinterandintrafractionmotionerrorsatexternalradiotherapyusingamarkerlessstrategybasedonimageregistrationcombinedwithcorrelationmodel
AT ahmadesmailitorshabi calculationofinterandintrafractionmotionerrorsatexternalradiotherapyusingamarkerlessstrategybasedonimageregistrationcombinedwithcorrelationmodel
AT soheliaparandeh calculationofinterandintrafractionmotionerrorsatexternalradiotherapyusingamarkerlessstrategybasedonimageregistrationcombinedwithcorrelationmodel
_version_ 1725303138597994496