Ultra-Long-Term Reliable Encapsulation Using an Atomic Layer Deposited HfO<sub>2</sub>/Al<sub>2</sub>O<sub>3</sub>/HfO<sub>2</sub> Triple-Interlayer for Biomedical Implants

Long-term packaging of miniaturized, flexible implantable medical devices is essential for the next generation of medical devices. Polymer materials that are biocompatible and flexible have attracted extensive interest for the packaging of implantable medical devices, however realizing these devices...

Full description

Bibliographic Details
Main Authors: Changzheng Li, Maarten Cauwe, Yang Yang, David Schaubroeck, Lothar Mader, Maaike Op de Beeck
Format: Article
Language:English
Published: MDPI AG 2019-09-01
Series:Coatings
Subjects:
EIS
Online Access:https://www.mdpi.com/2079-6412/9/9/579
id doaj-9f6e4d74884c4fa3b9e03b5b953f61d7
record_format Article
spelling doaj-9f6e4d74884c4fa3b9e03b5b953f61d72020-11-25T01:22:45ZengMDPI AGCoatings2079-64122019-09-019957910.3390/coatings9090579coatings9090579Ultra-Long-Term Reliable Encapsulation Using an Atomic Layer Deposited HfO<sub>2</sub>/Al<sub>2</sub>O<sub>3</sub>/HfO<sub>2</sub> Triple-Interlayer for Biomedical ImplantsChangzheng Li0Maarten Cauwe1Yang Yang2David Schaubroeck3Lothar Mader4Maaike Op de Beeck5Centre for Microsystems Technology (CMST), imec and Ghent University, Technologiepark 126, 9052 Gent, BelgiumCentre for Microsystems Technology (CMST), imec and Ghent University, Technologiepark 126, 9052 Gent, BelgiumCentre for Microsystems Technology (CMST), imec and Ghent University, Technologiepark 126, 9052 Gent, BelgiumCentre for Microsystems Technology (CMST), imec and Ghent University, Technologiepark 126, 9052 Gent, BelgiumCentre for Microsystems Technology (CMST), imec and Ghent University, Technologiepark 126, 9052 Gent, BelgiumCentre for Microsystems Technology (CMST), imec and Ghent University, Technologiepark 126, 9052 Gent, BelgiumLong-term packaging of miniaturized, flexible implantable medical devices is essential for the next generation of medical devices. Polymer materials that are biocompatible and flexible have attracted extensive interest for the packaging of implantable medical devices, however realizing these devices with long-term hermeticity up to several years remains a great challenge. Here, polyimide (PI) based hermetic encapsulation was greatly improved by atomic layer deposition (ALD) of a nanoscale-thin, biocompatible sandwich stack of HfO<sub>2</sub>/Al<sub>2</sub>O<sub>3</sub>/HfO<sub>2</sub> (ALD-3) between two polyimide layers. A thin copper film covered with a PI/ALD-3/PI barrier maintained excellent electrochemical performance over 1028 days (2.8 years) during acceleration tests at 60 &#176;C in phosphate buffered saline solution (PBS). This stability is equivalent to approximately 14 years at 37 &#176;C. The coatings were monitored in situ through electrochemical impedance spectroscopy (EIS), were inspected by microscope, and were further analyzed using equivalent circuit modeling. The failure mode of ALD Al<sub>2</sub>O<sub>3</sub>, ALD-3, and PI soaking in PBS is discussed. Encapsulation using ultrathin ALD-3 combined with PI for the packaging of implantable medical devices is robust at the acceleration temperature condition for more than 2.8 years, showing that it has great potential as reliable packaging for long-term implantable devices.https://www.mdpi.com/2079-6412/9/9/579Al<sub>2</sub>O<sub>3</sub>HfO<sub>2</sub>polyimideEISmoisture barrierslifetimeimplantable medical devices
collection DOAJ
language English
format Article
sources DOAJ
author Changzheng Li
Maarten Cauwe
Yang Yang
David Schaubroeck
Lothar Mader
Maaike Op de Beeck
spellingShingle Changzheng Li
Maarten Cauwe
Yang Yang
David Schaubroeck
Lothar Mader
Maaike Op de Beeck
Ultra-Long-Term Reliable Encapsulation Using an Atomic Layer Deposited HfO<sub>2</sub>/Al<sub>2</sub>O<sub>3</sub>/HfO<sub>2</sub> Triple-Interlayer for Biomedical Implants
Coatings
Al<sub>2</sub>O<sub>3</sub>
HfO<sub>2</sub>
polyimide
EIS
moisture barriers
lifetime
implantable medical devices
author_facet Changzheng Li
Maarten Cauwe
Yang Yang
David Schaubroeck
Lothar Mader
Maaike Op de Beeck
author_sort Changzheng Li
title Ultra-Long-Term Reliable Encapsulation Using an Atomic Layer Deposited HfO<sub>2</sub>/Al<sub>2</sub>O<sub>3</sub>/HfO<sub>2</sub> Triple-Interlayer for Biomedical Implants
title_short Ultra-Long-Term Reliable Encapsulation Using an Atomic Layer Deposited HfO<sub>2</sub>/Al<sub>2</sub>O<sub>3</sub>/HfO<sub>2</sub> Triple-Interlayer for Biomedical Implants
title_full Ultra-Long-Term Reliable Encapsulation Using an Atomic Layer Deposited HfO<sub>2</sub>/Al<sub>2</sub>O<sub>3</sub>/HfO<sub>2</sub> Triple-Interlayer for Biomedical Implants
title_fullStr Ultra-Long-Term Reliable Encapsulation Using an Atomic Layer Deposited HfO<sub>2</sub>/Al<sub>2</sub>O<sub>3</sub>/HfO<sub>2</sub> Triple-Interlayer for Biomedical Implants
title_full_unstemmed Ultra-Long-Term Reliable Encapsulation Using an Atomic Layer Deposited HfO<sub>2</sub>/Al<sub>2</sub>O<sub>3</sub>/HfO<sub>2</sub> Triple-Interlayer for Biomedical Implants
title_sort ultra-long-term reliable encapsulation using an atomic layer deposited hfo<sub>2</sub>/al<sub>2</sub>o<sub>3</sub>/hfo<sub>2</sub> triple-interlayer for biomedical implants
publisher MDPI AG
series Coatings
issn 2079-6412
publishDate 2019-09-01
description Long-term packaging of miniaturized, flexible implantable medical devices is essential for the next generation of medical devices. Polymer materials that are biocompatible and flexible have attracted extensive interest for the packaging of implantable medical devices, however realizing these devices with long-term hermeticity up to several years remains a great challenge. Here, polyimide (PI) based hermetic encapsulation was greatly improved by atomic layer deposition (ALD) of a nanoscale-thin, biocompatible sandwich stack of HfO<sub>2</sub>/Al<sub>2</sub>O<sub>3</sub>/HfO<sub>2</sub> (ALD-3) between two polyimide layers. A thin copper film covered with a PI/ALD-3/PI barrier maintained excellent electrochemical performance over 1028 days (2.8 years) during acceleration tests at 60 &#176;C in phosphate buffered saline solution (PBS). This stability is equivalent to approximately 14 years at 37 &#176;C. The coatings were monitored in situ through electrochemical impedance spectroscopy (EIS), were inspected by microscope, and were further analyzed using equivalent circuit modeling. The failure mode of ALD Al<sub>2</sub>O<sub>3</sub>, ALD-3, and PI soaking in PBS is discussed. Encapsulation using ultrathin ALD-3 combined with PI for the packaging of implantable medical devices is robust at the acceleration temperature condition for more than 2.8 years, showing that it has great potential as reliable packaging for long-term implantable devices.
topic Al<sub>2</sub>O<sub>3</sub>
HfO<sub>2</sub>
polyimide
EIS
moisture barriers
lifetime
implantable medical devices
url https://www.mdpi.com/2079-6412/9/9/579
work_keys_str_mv AT changzhengli ultralongtermreliableencapsulationusinganatomiclayerdepositedhfosub2subalsub2subosub3subhfosub2subtripleinterlayerforbiomedicalimplants
AT maartencauwe ultralongtermreliableencapsulationusinganatomiclayerdepositedhfosub2subalsub2subosub3subhfosub2subtripleinterlayerforbiomedicalimplants
AT yangyang ultralongtermreliableencapsulationusinganatomiclayerdepositedhfosub2subalsub2subosub3subhfosub2subtripleinterlayerforbiomedicalimplants
AT davidschaubroeck ultralongtermreliableencapsulationusinganatomiclayerdepositedhfosub2subalsub2subosub3subhfosub2subtripleinterlayerforbiomedicalimplants
AT lotharmader ultralongtermreliableencapsulationusinganatomiclayerdepositedhfosub2subalsub2subosub3subhfosub2subtripleinterlayerforbiomedicalimplants
AT maaikeopdebeeck ultralongtermreliableencapsulationusinganatomiclayerdepositedhfosub2subalsub2subosub3subhfosub2subtripleinterlayerforbiomedicalimplants
_version_ 1725125696696614912