Repeatability and Interchangeability of Topometric, Anterior Chamber and Corneal Wavefront Data Between Two Scheimpflug Camera Devices

Hesham Mohamed Gharieb,1,2 Hisham Samy Shalaby,3 Ihab Saad Othman4,5 1Faculty of Medicine, Ain Shams University, Cairo, Egypt; 2Research and Development, Eye World Hospital, Giza, Egypt; 3Department of Ophthalmology, Faculty of Medicine, Ain Shams University, Cairo, Egypt; 4Eye World Hospital, Giza,...

Full description

Bibliographic Details
Main Authors: Gharieb HM, Shalaby HS, Othman IS
Format: Article
Language:English
Published: Dove Medical Press 2020-11-01
Series:Clinical Ophthalmology
Subjects:
Online Access:https://www.dovepress.com/repeatability-and-interchangeability-of-topometric-anterior-chamber-an-peer-reviewed-article-OPTH
id doaj-918c8309f1c648d28b26b5d08b0b839e
record_format Article
spelling doaj-918c8309f1c648d28b26b5d08b0b839e2020-11-25T04:07:15ZengDove Medical PressClinical Ophthalmology1177-54832020-11-01Volume 143801381058992Repeatability and Interchangeability of Topometric, Anterior Chamber and Corneal Wavefront Data Between Two Scheimpflug Camera DevicesGharieb HMShalaby HSOthman ISHesham Mohamed Gharieb,1,2 Hisham Samy Shalaby,3 Ihab Saad Othman4,5 1Faculty of Medicine, Ain Shams University, Cairo, Egypt; 2Research and Development, Eye World Hospital, Giza, Egypt; 3Department of Ophthalmology, Faculty of Medicine, Ain Shams University, Cairo, Egypt; 4Eye World Hospital, Giza, Egypt; 5Faculty of Medicine, Cairo University, Giza, EgyptCorrespondence: Hisham Samy ShalabyDepartment of Ophthalmology, Faculty of Medicine, Ain Shams University, Abbassia Square, Cairo 11591, EgyptTel +20 1003641789Email hisham_ophth@yahoo.comPurpose: This study aims to assess the intra-operator repeatability and correlation of the Pentacam HR (device 1) and Sirius (device 2) in measuring anterior segment parameters and to evaluate the agreement of their readings and therefore their interchangeability in a clinical setting.Methods: This is a prospective non-randomized study was conducted on the right eyes of 102 subjects coming to Eye World Hospital, Giza, Egypt. With each machine, four scans were taken by a single examiner. Each device was used to measure keratometric indices, corneal thickness, anterior chamber depth, anterior chamber angle, corneal diameter and corneal optical aberrations.Results: Both devices show high repeatability for corneal thickness, corneal diameter, anterior chamber depth and keratometric indices (except for maximum keratometry, where device 1 shows high repeatability and device 2 shows low repeatability). On the other hand, both devices show poor repeatability for anterior chamber angle, Q-values, root mean square, spherical, coma and trefoil aberrations. The readings of the two devices are strongly correlated as regards only keratometric indices, corneal thickness and anterior chamber depth. In addition, the readings of the devices are in good agreement as regards only keratometric indices (except maximum keratometry), corneal thickness, anterior chamber depth, anterior chamber angle, root mean square, spherical and trefoil aberrations.Conclusion: Both devices showed variable intra-observer repeatability, with the device 1 showing slightly higher repeatability. Despite the similarity between some of the readings of the two devices, caution is advised before considering them interchangeable. We therefore do not recommend using them in alternation in refractive surgery.Keywords: Scheimpflug, wavefront, corneal topographyhttps://www.dovepress.com/repeatability-and-interchangeability-of-topometric-anterior-chamber-an-peer-reviewed-article-OPTHscheimpflugwavefrontcorneal topography
collection DOAJ
language English
format Article
sources DOAJ
author Gharieb HM
Shalaby HS
Othman IS
spellingShingle Gharieb HM
Shalaby HS
Othman IS
Repeatability and Interchangeability of Topometric, Anterior Chamber and Corneal Wavefront Data Between Two Scheimpflug Camera Devices
Clinical Ophthalmology
scheimpflug
wavefront
corneal topography
author_facet Gharieb HM
Shalaby HS
Othman IS
author_sort Gharieb HM
title Repeatability and Interchangeability of Topometric, Anterior Chamber and Corneal Wavefront Data Between Two Scheimpflug Camera Devices
title_short Repeatability and Interchangeability of Topometric, Anterior Chamber and Corneal Wavefront Data Between Two Scheimpflug Camera Devices
title_full Repeatability and Interchangeability of Topometric, Anterior Chamber and Corneal Wavefront Data Between Two Scheimpflug Camera Devices
title_fullStr Repeatability and Interchangeability of Topometric, Anterior Chamber and Corneal Wavefront Data Between Two Scheimpflug Camera Devices
title_full_unstemmed Repeatability and Interchangeability of Topometric, Anterior Chamber and Corneal Wavefront Data Between Two Scheimpflug Camera Devices
title_sort repeatability and interchangeability of topometric, anterior chamber and corneal wavefront data between two scheimpflug camera devices
publisher Dove Medical Press
series Clinical Ophthalmology
issn 1177-5483
publishDate 2020-11-01
description Hesham Mohamed Gharieb,1,2 Hisham Samy Shalaby,3 Ihab Saad Othman4,5 1Faculty of Medicine, Ain Shams University, Cairo, Egypt; 2Research and Development, Eye World Hospital, Giza, Egypt; 3Department of Ophthalmology, Faculty of Medicine, Ain Shams University, Cairo, Egypt; 4Eye World Hospital, Giza, Egypt; 5Faculty of Medicine, Cairo University, Giza, EgyptCorrespondence: Hisham Samy ShalabyDepartment of Ophthalmology, Faculty of Medicine, Ain Shams University, Abbassia Square, Cairo 11591, EgyptTel +20 1003641789Email hisham_ophth@yahoo.comPurpose: This study aims to assess the intra-operator repeatability and correlation of the Pentacam HR (device 1) and Sirius (device 2) in measuring anterior segment parameters and to evaluate the agreement of their readings and therefore their interchangeability in a clinical setting.Methods: This is a prospective non-randomized study was conducted on the right eyes of 102 subjects coming to Eye World Hospital, Giza, Egypt. With each machine, four scans were taken by a single examiner. Each device was used to measure keratometric indices, corneal thickness, anterior chamber depth, anterior chamber angle, corneal diameter and corneal optical aberrations.Results: Both devices show high repeatability for corneal thickness, corneal diameter, anterior chamber depth and keratometric indices (except for maximum keratometry, where device 1 shows high repeatability and device 2 shows low repeatability). On the other hand, both devices show poor repeatability for anterior chamber angle, Q-values, root mean square, spherical, coma and trefoil aberrations. The readings of the two devices are strongly correlated as regards only keratometric indices, corneal thickness and anterior chamber depth. In addition, the readings of the devices are in good agreement as regards only keratometric indices (except maximum keratometry), corneal thickness, anterior chamber depth, anterior chamber angle, root mean square, spherical and trefoil aberrations.Conclusion: Both devices showed variable intra-observer repeatability, with the device 1 showing slightly higher repeatability. Despite the similarity between some of the readings of the two devices, caution is advised before considering them interchangeable. We therefore do not recommend using them in alternation in refractive surgery.Keywords: Scheimpflug, wavefront, corneal topography
topic scheimpflug
wavefront
corneal topography
url https://www.dovepress.com/repeatability-and-interchangeability-of-topometric-anterior-chamber-an-peer-reviewed-article-OPTH
work_keys_str_mv AT ghariebhm repeatabilityandinterchangeabilityoftopometricanteriorchamberandcornealwavefrontdatabetweentwoscheimpflugcameradevices
AT shalabyhs repeatabilityandinterchangeabilityoftopometricanteriorchamberandcornealwavefrontdatabetweentwoscheimpflugcameradevices
AT othmanis repeatabilityandinterchangeabilityoftopometricanteriorchamberandcornealwavefrontdatabetweentwoscheimpflugcameradevices
_version_ 1724429430075424768