Film Reject Analysis and Radiation Doses Received by Patients in Selected Hospitals in Southwestern Nigeria

Introduction: A reject rate is the percentage of diagnostic images repeated due to errors during radiological examinations. The present study aimed to evaluate the patient radiation doses and analyze the film reject rate as part of quality assurance program in three diagnostic centers in Nigeria. Ma...

Full description

Bibliographic Details
Main Authors: WASIU ERINOSO, Rachel Obed, Christopher Olowookere
Format: Article
Language:English
Published: Mashhad University of Medical Sciences 2017-12-01
Series:Iranian Journal of Medical Physics
Subjects:
Online Access:http://ijmp.mums.ac.ir/article_8881_a682b5ba20c17921b41a71d1124a6bca.pdf
id doaj-9021e88d512242d49f8fc3e0411a4680
record_format Article
spelling doaj-9021e88d512242d49f8fc3e0411a46802020-11-25T01:09:24ZengMashhad University of Medical SciencesIranian Journal of Medical Physics2345-36722345-36722017-12-0114418318910.22038/ijmp.2017.19521.11798881Film Reject Analysis and Radiation Doses Received by Patients in Selected Hospitals in Southwestern NigeriaWASIU ERINOSO0Rachel Obed1Christopher Olowookere2UNIVERSITY OF IBADANUniversity of IbadanAjayi Crowther UniversityIntroduction: A reject rate is the percentage of diagnostic images repeated due to errors during radiological examinations. The present study aimed to evaluate the patient radiation doses and analyze the film reject rate as part of quality assurance program in three diagnostic centers in Nigeria. Materials and Methods: This study was conducted in three hospitals, namely Federal Medical Center (FMC), General Hospital (GH), and Sacred Heart Hospital (SHH), located in Abeokuta, Ogun State, Southwestern Nigeria. For the purpose of the study, the accepted and rejected radiographs during different X-ray examinations were recorded. A total of 376 rejected and accepted radiographs were evaluated in the three hospitals, and the economic losses due to rejected films were determined. The quality control (QC) tests, which involve kilo voltage (kV), milliampere seconds (mAs), etc, were carried out on the facilities of two out of three hospitals using Victoreen 6000m QC kits. The results of the QC tests and exposure parameters were used to estimate the patient doses for different examinations carried out during the study. Results: Based on the results of the study, the SHH  had the highest estimated annual loss of $225, followed by the FMC and GH with annual monetary losses of $208 and $166, respectively. In addition, the anteroposterior projection of the lumbosacral spine had the highest mean dose (15±1.64 mGy) in this study, which was observed in the SHH. Additionally, at FMC, all the estimated doses were low. Conclusion: Regarding the monetary loss and increase in patient dose burden involved in repeated examinations, it is essential to train personnel on the factors leading to repeated exposures.http://ijmp.mums.ac.ir/article_8881_a682b5ba20c17921b41a71d1124a6bca.pdfDiagnostic ImagingImage QualityQuality ControlRadiation Dose
collection DOAJ
language English
format Article
sources DOAJ
author WASIU ERINOSO
Rachel Obed
Christopher Olowookere
spellingShingle WASIU ERINOSO
Rachel Obed
Christopher Olowookere
Film Reject Analysis and Radiation Doses Received by Patients in Selected Hospitals in Southwestern Nigeria
Iranian Journal of Medical Physics
Diagnostic Imaging
Image Quality
Quality Control
Radiation Dose
author_facet WASIU ERINOSO
Rachel Obed
Christopher Olowookere
author_sort WASIU ERINOSO
title Film Reject Analysis and Radiation Doses Received by Patients in Selected Hospitals in Southwestern Nigeria
title_short Film Reject Analysis and Radiation Doses Received by Patients in Selected Hospitals in Southwestern Nigeria
title_full Film Reject Analysis and Radiation Doses Received by Patients in Selected Hospitals in Southwestern Nigeria
title_fullStr Film Reject Analysis and Radiation Doses Received by Patients in Selected Hospitals in Southwestern Nigeria
title_full_unstemmed Film Reject Analysis and Radiation Doses Received by Patients in Selected Hospitals in Southwestern Nigeria
title_sort film reject analysis and radiation doses received by patients in selected hospitals in southwestern nigeria
publisher Mashhad University of Medical Sciences
series Iranian Journal of Medical Physics
issn 2345-3672
2345-3672
publishDate 2017-12-01
description Introduction: A reject rate is the percentage of diagnostic images repeated due to errors during radiological examinations. The present study aimed to evaluate the patient radiation doses and analyze the film reject rate as part of quality assurance program in three diagnostic centers in Nigeria. Materials and Methods: This study was conducted in three hospitals, namely Federal Medical Center (FMC), General Hospital (GH), and Sacred Heart Hospital (SHH), located in Abeokuta, Ogun State, Southwestern Nigeria. For the purpose of the study, the accepted and rejected radiographs during different X-ray examinations were recorded. A total of 376 rejected and accepted radiographs were evaluated in the three hospitals, and the economic losses due to rejected films were determined. The quality control (QC) tests, which involve kilo voltage (kV), milliampere seconds (mAs), etc, were carried out on the facilities of two out of three hospitals using Victoreen 6000m QC kits. The results of the QC tests and exposure parameters were used to estimate the patient doses for different examinations carried out during the study. Results: Based on the results of the study, the SHH  had the highest estimated annual loss of $225, followed by the FMC and GH with annual monetary losses of $208 and $166, respectively. In addition, the anteroposterior projection of the lumbosacral spine had the highest mean dose (15±1.64 mGy) in this study, which was observed in the SHH. Additionally, at FMC, all the estimated doses were low. Conclusion: Regarding the monetary loss and increase in patient dose burden involved in repeated examinations, it is essential to train personnel on the factors leading to repeated exposures.
topic Diagnostic Imaging
Image Quality
Quality Control
Radiation Dose
url http://ijmp.mums.ac.ir/article_8881_a682b5ba20c17921b41a71d1124a6bca.pdf
work_keys_str_mv AT wasiuerinoso filmrejectanalysisandradiationdosesreceivedbypatientsinselectedhospitalsinsouthwesternnigeria
AT rachelobed filmrejectanalysisandradiationdosesreceivedbypatientsinselectedhospitalsinsouthwesternnigeria
AT christopherolowookere filmrejectanalysisandradiationdosesreceivedbypatientsinselectedhospitalsinsouthwesternnigeria
_version_ 1725179076028661760