Revalidation of CoaguChek XS Plus System for INR Monitoring in Taiwanese Patients: Effects of Clinical and Genetic Factors

Background:In this study, we evaluated the performance of a point-of-care device, the CoaguChek XS Plus system, in the determination of prothrombin time and international normalized ratio (INR) based on ISO17593: 2007 criteria in Taiwanese patients. The underlying clinical and genetic factors were a...

Full description

Bibliographic Details
Main Authors: Chao-Hua Fu, Wei-Ting Chen, Pi-Yueh Chang, Ming-Ta Michael Lee, Ming-Shien Wen
Format: Article
Language:English
Published: Elsevier 2014-12-01
Series:Biomedical Journal
Subjects:
Online Access:http://www.biomedj.org/article.asp?issn=2319-4170;year=2014;volume=37;issue=6;spage=380;epage=385;aulast=Fu
id doaj-8ffba93d4bde4c8088079681751da14f
record_format Article
spelling doaj-8ffba93d4bde4c8088079681751da14f2021-03-02T10:28:00ZengElsevierBiomedical Journal2319-41702320-28902014-12-0137638038510.4103/2319-4170.132904Revalidation of CoaguChek XS Plus System for INR Monitoring in Taiwanese Patients: Effects of Clinical and Genetic FactorsChao-Hua Fu0Wei-Ting Chen1Pi-Yueh Chang2Ming-Ta Michael Lee3Ming-Shien Wen4Department of Laboratory Medicine, Chang Gung Memorial Hospital at Linkou, Chang Gung University College of Medicine, Taoyuan, TaiwanDepartment of Medicine, Second Section of Cardiology, Chang Gung Memorial Hospital at Linkou, Chang Gung University College of Medicine, Taoyuan, TaiwanDepartment of Laboratory Medicine, Chang Gung Memorial Hospital at Linkou, Chang Gung University College of Medicine; Department of Medical Biotechnology and Laboratory Science, Chang Gung University, Taoyuan, TaiwanInstitute of Biomedical Sciences, Academic Sinica, Taipei, Taiwan; Laboratory for International Alliance, Center for Genomic Medicine, RIKEN, Yokohama, JapanDepartment of Laboratory Medicine, Chang Gung Memorial Hospital at Linkou, Chang Gung University College of Medicine; School of Medicine, Chang Gung University, Taoyuan, TaiwanBackground:In this study, we evaluated the performance of a point-of-care device, the CoaguChek XS Plus system, in the determination of prothrombin time and international normalized ratio (INR) based on ISO17593: 2007 criteria in Taiwanese patients. The underlying clinical and genetic factors were also investigated. Methods: Fifty patients receiving warfarin therapy were enrolled in this study. The accuracy of the CoaguChek XS Plus system was evaluated with linear regression analysis and bias plot by comparing with the data measured using Sysmex CA-1500. The clinical and genetic factors that may have caused a bias of ΃ 0.5 INR were evaluated with Fisher's exact test. Results: From the 50 patients, 93 INR values were collected by each method. Linear regression analysis indicated a high correlation with r = 0.96, a slope of 1.05, and an intercept of - 0.14. Eight patients showed an INR bias ΃0.5 between the two methods. Only aspartate aminotransferase (AST) >34 U/L (3/8, 37.5% vs. 3/42, 7.1%; p = 0.044) and alanine aminotransferase (ALT) >36 U/L (3/8, 37.5% vs. 3/42, 7.1%; p = 0.044) were significantly different from each other. No differences were observed for hypoalbuminemia, elevated creatinine, anemia, and the polymorphisms of VKORC1 and CYP2C9. Conclusions: The CoaguChek XS Plus system presented results that were comparable with those obtained using laboratory CA-1500 method. Both methods fell within INR in the range of 2-4.5 defined by ISO17593:2007 and the clinically recognized therapeutic INR range of 2-3.5. Elevated AST and ALT levels might have interfered with the INR results.http://www.biomedj.org/article.asp?issn=2319-4170;year=2014;volume=37;issue=6;spage=380;epage=385;aulast=FuCoaguChek XS systeminterferenceISO17593:2007point-of-carewarfarin
collection DOAJ
language English
format Article
sources DOAJ
author Chao-Hua Fu
Wei-Ting Chen
Pi-Yueh Chang
Ming-Ta Michael Lee
Ming-Shien Wen
spellingShingle Chao-Hua Fu
Wei-Ting Chen
Pi-Yueh Chang
Ming-Ta Michael Lee
Ming-Shien Wen
Revalidation of CoaguChek XS Plus System for INR Monitoring in Taiwanese Patients: Effects of Clinical and Genetic Factors
Biomedical Journal
CoaguChek XS system
interference
ISO17593:2007
point-of-care
warfarin
author_facet Chao-Hua Fu
Wei-Ting Chen
Pi-Yueh Chang
Ming-Ta Michael Lee
Ming-Shien Wen
author_sort Chao-Hua Fu
title Revalidation of CoaguChek XS Plus System for INR Monitoring in Taiwanese Patients: Effects of Clinical and Genetic Factors
title_short Revalidation of CoaguChek XS Plus System for INR Monitoring in Taiwanese Patients: Effects of Clinical and Genetic Factors
title_full Revalidation of CoaguChek XS Plus System for INR Monitoring in Taiwanese Patients: Effects of Clinical and Genetic Factors
title_fullStr Revalidation of CoaguChek XS Plus System for INR Monitoring in Taiwanese Patients: Effects of Clinical and Genetic Factors
title_full_unstemmed Revalidation of CoaguChek XS Plus System for INR Monitoring in Taiwanese Patients: Effects of Clinical and Genetic Factors
title_sort revalidation of coaguchek xs plus system for inr monitoring in taiwanese patients: effects of clinical and genetic factors
publisher Elsevier
series Biomedical Journal
issn 2319-4170
2320-2890
publishDate 2014-12-01
description Background:In this study, we evaluated the performance of a point-of-care device, the CoaguChek XS Plus system, in the determination of prothrombin time and international normalized ratio (INR) based on ISO17593: 2007 criteria in Taiwanese patients. The underlying clinical and genetic factors were also investigated. Methods: Fifty patients receiving warfarin therapy were enrolled in this study. The accuracy of the CoaguChek XS Plus system was evaluated with linear regression analysis and bias plot by comparing with the data measured using Sysmex CA-1500. The clinical and genetic factors that may have caused a bias of ΃ 0.5 INR were evaluated with Fisher's exact test. Results: From the 50 patients, 93 INR values were collected by each method. Linear regression analysis indicated a high correlation with r = 0.96, a slope of 1.05, and an intercept of - 0.14. Eight patients showed an INR bias ΃0.5 between the two methods. Only aspartate aminotransferase (AST) >34 U/L (3/8, 37.5% vs. 3/42, 7.1%; p = 0.044) and alanine aminotransferase (ALT) >36 U/L (3/8, 37.5% vs. 3/42, 7.1%; p = 0.044) were significantly different from each other. No differences were observed for hypoalbuminemia, elevated creatinine, anemia, and the polymorphisms of VKORC1 and CYP2C9. Conclusions: The CoaguChek XS Plus system presented results that were comparable with those obtained using laboratory CA-1500 method. Both methods fell within INR in the range of 2-4.5 defined by ISO17593:2007 and the clinically recognized therapeutic INR range of 2-3.5. Elevated AST and ALT levels might have interfered with the INR results.
topic CoaguChek XS system
interference
ISO17593:2007
point-of-care
warfarin
url http://www.biomedj.org/article.asp?issn=2319-4170;year=2014;volume=37;issue=6;spage=380;epage=385;aulast=Fu
work_keys_str_mv AT chaohuafu revalidationofcoaguchekxsplussystemforinrmonitoringintaiwanesepatientseffectsofclinicalandgeneticfactors
AT weitingchen revalidationofcoaguchekxsplussystemforinrmonitoringintaiwanesepatientseffectsofclinicalandgeneticfactors
AT piyuehchang revalidationofcoaguchekxsplussystemforinrmonitoringintaiwanesepatientseffectsofclinicalandgeneticfactors
AT mingtamichaellee revalidationofcoaguchekxsplussystemforinrmonitoringintaiwanesepatientseffectsofclinicalandgeneticfactors
AT mingshienwen revalidationofcoaguchekxsplussystemforinrmonitoringintaiwanesepatientseffectsofclinicalandgeneticfactors
_version_ 1724236858209075200