Revalidation of CoaguChek XS Plus System for INR Monitoring in Taiwanese Patients: Effects of Clinical and Genetic Factors
Background:In this study, we evaluated the performance of a point-of-care device, the CoaguChek XS Plus system, in the determination of prothrombin time and international normalized ratio (INR) based on ISO17593: 2007 criteria in Taiwanese patients. The underlying clinical and genetic factors were a...
Main Authors: | , , , , |
---|---|
Format: | Article |
Language: | English |
Published: |
Elsevier
2014-12-01
|
Series: | Biomedical Journal |
Subjects: | |
Online Access: | http://www.biomedj.org/article.asp?issn=2319-4170;year=2014;volume=37;issue=6;spage=380;epage=385;aulast=Fu |
id |
doaj-8ffba93d4bde4c8088079681751da14f |
---|---|
record_format |
Article |
spelling |
doaj-8ffba93d4bde4c8088079681751da14f2021-03-02T10:28:00ZengElsevierBiomedical Journal2319-41702320-28902014-12-0137638038510.4103/2319-4170.132904Revalidation of CoaguChek XS Plus System for INR Monitoring in Taiwanese Patients: Effects of Clinical and Genetic FactorsChao-Hua Fu0Wei-Ting Chen1Pi-Yueh Chang2Ming-Ta Michael Lee3Ming-Shien Wen4Department of Laboratory Medicine, Chang Gung Memorial Hospital at Linkou, Chang Gung University College of Medicine, Taoyuan, TaiwanDepartment of Medicine, Second Section of Cardiology, Chang Gung Memorial Hospital at Linkou, Chang Gung University College of Medicine, Taoyuan, TaiwanDepartment of Laboratory Medicine, Chang Gung Memorial Hospital at Linkou, Chang Gung University College of Medicine; Department of Medical Biotechnology and Laboratory Science, Chang Gung University, Taoyuan, TaiwanInstitute of Biomedical Sciences, Academic Sinica, Taipei, Taiwan; Laboratory for International Alliance, Center for Genomic Medicine, RIKEN, Yokohama, JapanDepartment of Laboratory Medicine, Chang Gung Memorial Hospital at Linkou, Chang Gung University College of Medicine; School of Medicine, Chang Gung University, Taoyuan, TaiwanBackground:In this study, we evaluated the performance of a point-of-care device, the CoaguChek XS Plus system, in the determination of prothrombin time and international normalized ratio (INR) based on ISO17593: 2007 criteria in Taiwanese patients. The underlying clinical and genetic factors were also investigated. Methods: Fifty patients receiving warfarin therapy were enrolled in this study. The accuracy of the CoaguChek XS Plus system was evaluated with linear regression analysis and bias plot by comparing with the data measured using Sysmex CA-1500. The clinical and genetic factors that may have caused a bias of 0.5 INR were evaluated with Fisher's exact test. Results: From the 50 patients, 93 INR values were collected by each method. Linear regression analysis indicated a high correlation with r = 0.96, a slope of 1.05, and an intercept of - 0.14. Eight patients showed an INR bias 0.5 between the two methods. Only aspartate aminotransferase (AST) >34 U/L (3/8, 37.5% vs. 3/42, 7.1%; p = 0.044) and alanine aminotransferase (ALT) >36 U/L (3/8, 37.5% vs. 3/42, 7.1%; p = 0.044) were significantly different from each other. No differences were observed for hypoalbuminemia, elevated creatinine, anemia, and the polymorphisms of VKORC1 and CYP2C9. Conclusions: The CoaguChek XS Plus system presented results that were comparable with those obtained using laboratory CA-1500 method. Both methods fell within INR in the range of 2-4.5 defined by ISO17593:2007 and the clinically recognized therapeutic INR range of 2-3.5. Elevated AST and ALT levels might have interfered with the INR results.http://www.biomedj.org/article.asp?issn=2319-4170;year=2014;volume=37;issue=6;spage=380;epage=385;aulast=FuCoaguChek XS systeminterferenceISO17593:2007point-of-carewarfarin |
collection |
DOAJ |
language |
English |
format |
Article |
sources |
DOAJ |
author |
Chao-Hua Fu Wei-Ting Chen Pi-Yueh Chang Ming-Ta Michael Lee Ming-Shien Wen |
spellingShingle |
Chao-Hua Fu Wei-Ting Chen Pi-Yueh Chang Ming-Ta Michael Lee Ming-Shien Wen Revalidation of CoaguChek XS Plus System for INR Monitoring in Taiwanese Patients: Effects of Clinical and Genetic Factors Biomedical Journal CoaguChek XS system interference ISO17593:2007 point-of-care warfarin |
author_facet |
Chao-Hua Fu Wei-Ting Chen Pi-Yueh Chang Ming-Ta Michael Lee Ming-Shien Wen |
author_sort |
Chao-Hua Fu |
title |
Revalidation of CoaguChek XS Plus System for INR Monitoring in Taiwanese Patients: Effects of Clinical and Genetic Factors |
title_short |
Revalidation of CoaguChek XS Plus System for INR Monitoring in Taiwanese Patients: Effects of Clinical and Genetic Factors |
title_full |
Revalidation of CoaguChek XS Plus System for INR Monitoring in Taiwanese Patients: Effects of Clinical and Genetic Factors |
title_fullStr |
Revalidation of CoaguChek XS Plus System for INR Monitoring in Taiwanese Patients: Effects of Clinical and Genetic Factors |
title_full_unstemmed |
Revalidation of CoaguChek XS Plus System for INR Monitoring in Taiwanese Patients: Effects of Clinical and Genetic Factors |
title_sort |
revalidation of coaguchek xs plus system for inr monitoring in taiwanese patients: effects of clinical and genetic factors |
publisher |
Elsevier |
series |
Biomedical Journal |
issn |
2319-4170 2320-2890 |
publishDate |
2014-12-01 |
description |
Background:In this study, we evaluated the performance of a point-of-care device, the CoaguChek XS Plus system, in the determination of prothrombin time and international normalized ratio (INR) based on ISO17593: 2007 criteria in Taiwanese patients. The underlying clinical and genetic factors were also investigated.
Methods: Fifty patients receiving warfarin therapy were enrolled in this study. The accuracy of the CoaguChek XS Plus system was evaluated with linear regression analysis and bias plot by comparing with the data measured using Sysmex CA-1500. The clinical and genetic factors that may have caused a bias of 0.5 INR were evaluated with Fisher's exact test.
Results: From the 50 patients, 93 INR values were collected by each method. Linear regression analysis indicated a high correlation with r = 0.96, a slope of 1.05, and an intercept of - 0.14. Eight patients showed an INR bias 0.5 between the two methods. Only aspartate aminotransferase (AST) >34 U/L (3/8, 37.5% vs. 3/42, 7.1%; p = 0.044) and alanine aminotransferase (ALT) >36 U/L (3/8, 37.5% vs. 3/42, 7.1%; p = 0.044) were significantly different from each other. No differences were observed for hypoalbuminemia, elevated creatinine, anemia, and the polymorphisms of VKORC1 and CYP2C9.
Conclusions: The CoaguChek XS Plus system presented results that were comparable with those obtained using laboratory CA-1500 method. Both methods fell within INR in the range of 2-4.5 defined by ISO17593:2007 and the clinically recognized therapeutic INR range of 2-3.5. Elevated AST and ALT levels might have interfered with the INR results. |
topic |
CoaguChek XS system interference ISO17593:2007 point-of-care warfarin |
url |
http://www.biomedj.org/article.asp?issn=2319-4170;year=2014;volume=37;issue=6;spage=380;epage=385;aulast=Fu |
work_keys_str_mv |
AT chaohuafu revalidationofcoaguchekxsplussystemforinrmonitoringintaiwanesepatientseffectsofclinicalandgeneticfactors AT weitingchen revalidationofcoaguchekxsplussystemforinrmonitoringintaiwanesepatientseffectsofclinicalandgeneticfactors AT piyuehchang revalidationofcoaguchekxsplussystemforinrmonitoringintaiwanesepatientseffectsofclinicalandgeneticfactors AT mingtamichaellee revalidationofcoaguchekxsplussystemforinrmonitoringintaiwanesepatientseffectsofclinicalandgeneticfactors AT mingshienwen revalidationofcoaguchekxsplussystemforinrmonitoringintaiwanesepatientseffectsofclinicalandgeneticfactors |
_version_ |
1724236858209075200 |