A cardiovascular magnetic resonance (CMR) safe metal braided catheter design for interventional CMR at 1.5 T: freedom from radiofrequency induced heating and preserved mechanical performance

Abstract Background Catheter designs incorporating metallic braiding have high torque control and kink resistance compared with unbraided alternatives. However, metallic segments longer than a quarter wavelength (~ 12 cm for 1.5 T scanner) are prone to radiofrequency (RF) induced heating during card...

Full description

Bibliographic Details
Main Authors: Korel D. Yildirim, Burcu Basar, A. E. Campbell-Washburn, Daniel A. Herzka, Ozgur Kocaturk, Robert J. Lederman
Format: Article
Language:English
Published: BMC 2019-03-01
Series:Journal of Cardiovascular Magnetic Resonance
Subjects:
Online Access:http://link.springer.com/article/10.1186/s12968-019-0526-7