Methodology using a portable X-ray fluorescence device for on-site and rapid evaluation of heavy-atom contamination in wounds: a model study for application to plutonium contamination.

Workers decommissioning the Fukushima-Daiichi nuclear power plant damaged from the Great East Japan Earthquake and resulting tsunami are at risk of injury with possible contamination from radioactive heavy atoms including actinides, such as plutonium. We propose a new methodology for on-site and rap...

Full description

Bibliographic Details
Main Authors: Hiroshi Yoshii, Kouta Yanagihara, Hitoshi Imaseki, Tsuyoshi Hamano, Hirokuni Yamanishi, Masayo Inagaki, Yasuhiro Sakai, Nobuyuki Sugiura, Osamu Kurihara, Kazuo Sakai
Format: Article
Language:English
Published: Public Library of Science (PLoS) 2014-01-01
Series:PLoS ONE
Online Access:http://europepmc.org/articles/PMC4092075?pdf=render
id doaj-8e83e08448e84449b359c5bf7536bb5d
record_format Article
spelling doaj-8e83e08448e84449b359c5bf7536bb5d2020-11-25T01:27:43ZengPublic Library of Science (PLoS)PLoS ONE1932-62032014-01-0197e10196610.1371/journal.pone.0101966Methodology using a portable X-ray fluorescence device for on-site and rapid evaluation of heavy-atom contamination in wounds: a model study for application to plutonium contamination.Hiroshi YoshiiKouta YanagiharaHitoshi ImasekiTsuyoshi HamanoHirokuni YamanishiMasayo InagakiYasuhiro SakaiNobuyuki SugiuraOsamu KuriharaKazuo SakaiWorkers decommissioning the Fukushima-Daiichi nuclear power plant damaged from the Great East Japan Earthquake and resulting tsunami are at risk of injury with possible contamination from radioactive heavy atoms including actinides, such as plutonium. We propose a new methodology for on-site and rapid evaluation of heavy-atom contamination in wounds using a portable X-ray fluorescence (XRF) device. In the present study, stable lead was used as the model contaminant substitute for radioactive heavy atoms. First, the wound model was developed by placing a liquid blood phantom on an epoxy resin wound phantom contaminated with lead. Next, the correlation between the concentration of contaminant and the XRF peak intensity was formulated considering the thickness of blood exiting the wound. Methods to determine the minimum detection limit (MDL) of contaminants at any maximal equivalent dose to the wound by XRF measurement were also established. For example, in this system, at a maximal equivalent dose of 16.5 mSv to the wound and blood thickness of 0.5 mm, the MDL value for lead was 1.2 ppm (3.1 nmol). The radioactivity of 239Pu corresponding to 3.1 nmol is 1.7 kBq, which is lower than the radioactivity of 239Pu contaminating puncture wounds in previous severe accidents. In conclusion, the established methodology could be beneficial for future development of a method to evaluate plutonium contamination in wounds. Highlights: Methodology for evaluation of heavy-atom contamination in a wound was established. A portable X-ray fluorescence device enables on-site, rapid and direct evaluation. This method is expected to be used for evaluation of plutonium contamination in wounds.http://europepmc.org/articles/PMC4092075?pdf=render
collection DOAJ
language English
format Article
sources DOAJ
author Hiroshi Yoshii
Kouta Yanagihara
Hitoshi Imaseki
Tsuyoshi Hamano
Hirokuni Yamanishi
Masayo Inagaki
Yasuhiro Sakai
Nobuyuki Sugiura
Osamu Kurihara
Kazuo Sakai
spellingShingle Hiroshi Yoshii
Kouta Yanagihara
Hitoshi Imaseki
Tsuyoshi Hamano
Hirokuni Yamanishi
Masayo Inagaki
Yasuhiro Sakai
Nobuyuki Sugiura
Osamu Kurihara
Kazuo Sakai
Methodology using a portable X-ray fluorescence device for on-site and rapid evaluation of heavy-atom contamination in wounds: a model study for application to plutonium contamination.
PLoS ONE
author_facet Hiroshi Yoshii
Kouta Yanagihara
Hitoshi Imaseki
Tsuyoshi Hamano
Hirokuni Yamanishi
Masayo Inagaki
Yasuhiro Sakai
Nobuyuki Sugiura
Osamu Kurihara
Kazuo Sakai
author_sort Hiroshi Yoshii
title Methodology using a portable X-ray fluorescence device for on-site and rapid evaluation of heavy-atom contamination in wounds: a model study for application to plutonium contamination.
title_short Methodology using a portable X-ray fluorescence device for on-site and rapid evaluation of heavy-atom contamination in wounds: a model study for application to plutonium contamination.
title_full Methodology using a portable X-ray fluorescence device for on-site and rapid evaluation of heavy-atom contamination in wounds: a model study for application to plutonium contamination.
title_fullStr Methodology using a portable X-ray fluorescence device for on-site and rapid evaluation of heavy-atom contamination in wounds: a model study for application to plutonium contamination.
title_full_unstemmed Methodology using a portable X-ray fluorescence device for on-site and rapid evaluation of heavy-atom contamination in wounds: a model study for application to plutonium contamination.
title_sort methodology using a portable x-ray fluorescence device for on-site and rapid evaluation of heavy-atom contamination in wounds: a model study for application to plutonium contamination.
publisher Public Library of Science (PLoS)
series PLoS ONE
issn 1932-6203
publishDate 2014-01-01
description Workers decommissioning the Fukushima-Daiichi nuclear power plant damaged from the Great East Japan Earthquake and resulting tsunami are at risk of injury with possible contamination from radioactive heavy atoms including actinides, such as plutonium. We propose a new methodology for on-site and rapid evaluation of heavy-atom contamination in wounds using a portable X-ray fluorescence (XRF) device. In the present study, stable lead was used as the model contaminant substitute for radioactive heavy atoms. First, the wound model was developed by placing a liquid blood phantom on an epoxy resin wound phantom contaminated with lead. Next, the correlation between the concentration of contaminant and the XRF peak intensity was formulated considering the thickness of blood exiting the wound. Methods to determine the minimum detection limit (MDL) of contaminants at any maximal equivalent dose to the wound by XRF measurement were also established. For example, in this system, at a maximal equivalent dose of 16.5 mSv to the wound and blood thickness of 0.5 mm, the MDL value for lead was 1.2 ppm (3.1 nmol). The radioactivity of 239Pu corresponding to 3.1 nmol is 1.7 kBq, which is lower than the radioactivity of 239Pu contaminating puncture wounds in previous severe accidents. In conclusion, the established methodology could be beneficial for future development of a method to evaluate plutonium contamination in wounds. Highlights: Methodology for evaluation of heavy-atom contamination in a wound was established. A portable X-ray fluorescence device enables on-site, rapid and direct evaluation. This method is expected to be used for evaluation of plutonium contamination in wounds.
url http://europepmc.org/articles/PMC4092075?pdf=render
work_keys_str_mv AT hiroshiyoshii methodologyusingaportablexrayfluorescencedeviceforonsiteandrapidevaluationofheavyatomcontaminationinwoundsamodelstudyforapplicationtoplutoniumcontamination
AT koutayanagihara methodologyusingaportablexrayfluorescencedeviceforonsiteandrapidevaluationofheavyatomcontaminationinwoundsamodelstudyforapplicationtoplutoniumcontamination
AT hitoshiimaseki methodologyusingaportablexrayfluorescencedeviceforonsiteandrapidevaluationofheavyatomcontaminationinwoundsamodelstudyforapplicationtoplutoniumcontamination
AT tsuyoshihamano methodologyusingaportablexrayfluorescencedeviceforonsiteandrapidevaluationofheavyatomcontaminationinwoundsamodelstudyforapplicationtoplutoniumcontamination
AT hirokuniyamanishi methodologyusingaportablexrayfluorescencedeviceforonsiteandrapidevaluationofheavyatomcontaminationinwoundsamodelstudyforapplicationtoplutoniumcontamination
AT masayoinagaki methodologyusingaportablexrayfluorescencedeviceforonsiteandrapidevaluationofheavyatomcontaminationinwoundsamodelstudyforapplicationtoplutoniumcontamination
AT yasuhirosakai methodologyusingaportablexrayfluorescencedeviceforonsiteandrapidevaluationofheavyatomcontaminationinwoundsamodelstudyforapplicationtoplutoniumcontamination
AT nobuyukisugiura methodologyusingaportablexrayfluorescencedeviceforonsiteandrapidevaluationofheavyatomcontaminationinwoundsamodelstudyforapplicationtoplutoniumcontamination
AT osamukurihara methodologyusingaportablexrayfluorescencedeviceforonsiteandrapidevaluationofheavyatomcontaminationinwoundsamodelstudyforapplicationtoplutoniumcontamination
AT kazuosakai methodologyusingaportablexrayfluorescencedeviceforonsiteandrapidevaluationofheavyatomcontaminationinwoundsamodelstudyforapplicationtoplutoniumcontamination
_version_ 1725103693751123968