Verification of setup errors in external beam radiation therapy using electronic portal imaging

The objective of this study was to conduct an audit on QA aspects of treatment delivery by the verification of the treatment fields′ position on different days to document the efficiency of immobilization methods and reproducibility of treatment. A retrospective study was carried out on 6...

Full description

Bibliographic Details
Main Authors: Murthy K, Al-Rahbi Zakiya, Sivakumar S, Davis C, Ravichandran R, El Ghamrawy Kamal
Format: Article
Language:English
Published: Wolters Kluwer Medknow Publications 2008-01-01
Series:Journal of Medical Physics
Subjects:
Online Access:http://www.jmp.org.in/article.asp?issn=0971-6203;year=2008;volume=33;issue=2;spage=49;epage=53;aulast=Murthy
id doaj-8db03183dd3f4520aa2434f573d66300
record_format Article
spelling doaj-8db03183dd3f4520aa2434f573d663002020-11-24T21:57:29ZengWolters Kluwer Medknow PublicationsJournal of Medical Physics0971-62031998-39132008-01-013324953Verification of setup errors in external beam radiation therapy using electronic portal imagingMurthy KAl-Rahbi ZakiyaSivakumar SDavis CRavichandran REl Ghamrawy KamalThe objective of this study was to conduct an audit on QA aspects of treatment delivery by the verification of the treatment fields′ position on different days to document the efficiency of immobilization methods and reproducibility of treatment. A retrospective study was carried out on 60 patients, each 20 treated for head and neck, breast, and pelvic sites; and a total of 506 images obtained by electronic portal imaging device (EPID) were analyzed. The portal images acquired using the EPID systems attached to the Varian linear accelerators were superimposed on the reference images. The anatomy matching software (Varian portal Vision. 6.0) was used, and the displacements in two dimensions and rotation were noted for each treated field to study the patient setup errors. The percentages of mean deviations more than 3 mm in ′lateral (X) and longitudinal (Y)′ directions were 17.5%, 11.25%, and 7.5% for breast, pelvis, and head and neck cases respectively. In all cases, the percentage of mean deviation with more than 5 mm error was 0.83%. The maximum average mean deviation in all the cases was 1.87. The average mean SD along X and Y directions in all the cases was less than 2.65. The results revealed that the ranges of setup errors are site specific and immobilization methods improve reproducibility. The observed variations were well within the limits. The study confirmed the accuracy and quality of treatments delivered to the patients. http://www.jmp.org.in/article.asp?issn=0971-6203;year=2008;volume=33;issue=2;spage=49;epage=53;aulast=MurthyAnatomy matching softwareelectronic portal imaging deviceimmobilization castssetup errortumor control probability
collection DOAJ
language English
format Article
sources DOAJ
author Murthy K
Al-Rahbi Zakiya
Sivakumar S
Davis C
Ravichandran R
El Ghamrawy Kamal
spellingShingle Murthy K
Al-Rahbi Zakiya
Sivakumar S
Davis C
Ravichandran R
El Ghamrawy Kamal
Verification of setup errors in external beam radiation therapy using electronic portal imaging
Journal of Medical Physics
Anatomy matching software
electronic portal imaging device
immobilization casts
setup error
tumor control probability
author_facet Murthy K
Al-Rahbi Zakiya
Sivakumar S
Davis C
Ravichandran R
El Ghamrawy Kamal
author_sort Murthy K
title Verification of setup errors in external beam radiation therapy using electronic portal imaging
title_short Verification of setup errors in external beam radiation therapy using electronic portal imaging
title_full Verification of setup errors in external beam radiation therapy using electronic portal imaging
title_fullStr Verification of setup errors in external beam radiation therapy using electronic portal imaging
title_full_unstemmed Verification of setup errors in external beam radiation therapy using electronic portal imaging
title_sort verification of setup errors in external beam radiation therapy using electronic portal imaging
publisher Wolters Kluwer Medknow Publications
series Journal of Medical Physics
issn 0971-6203
1998-3913
publishDate 2008-01-01
description The objective of this study was to conduct an audit on QA aspects of treatment delivery by the verification of the treatment fields′ position on different days to document the efficiency of immobilization methods and reproducibility of treatment. A retrospective study was carried out on 60 patients, each 20 treated for head and neck, breast, and pelvic sites; and a total of 506 images obtained by electronic portal imaging device (EPID) were analyzed. The portal images acquired using the EPID systems attached to the Varian linear accelerators were superimposed on the reference images. The anatomy matching software (Varian portal Vision. 6.0) was used, and the displacements in two dimensions and rotation were noted for each treated field to study the patient setup errors. The percentages of mean deviations more than 3 mm in ′lateral (X) and longitudinal (Y)′ directions were 17.5%, 11.25%, and 7.5% for breast, pelvis, and head and neck cases respectively. In all cases, the percentage of mean deviation with more than 5 mm error was 0.83%. The maximum average mean deviation in all the cases was 1.87. The average mean SD along X and Y directions in all the cases was less than 2.65. The results revealed that the ranges of setup errors are site specific and immobilization methods improve reproducibility. The observed variations were well within the limits. The study confirmed the accuracy and quality of treatments delivered to the patients.
topic Anatomy matching software
electronic portal imaging device
immobilization casts
setup error
tumor control probability
url http://www.jmp.org.in/article.asp?issn=0971-6203;year=2008;volume=33;issue=2;spage=49;epage=53;aulast=Murthy
work_keys_str_mv AT murthyk verificationofsetuperrorsinexternalbeamradiationtherapyusingelectronicportalimaging
AT alrahbizakiya verificationofsetuperrorsinexternalbeamradiationtherapyusingelectronicportalimaging
AT sivakumars verificationofsetuperrorsinexternalbeamradiationtherapyusingelectronicportalimaging
AT davisc verificationofsetuperrorsinexternalbeamradiationtherapyusingelectronicportalimaging
AT ravichandranr verificationofsetuperrorsinexternalbeamradiationtherapyusingelectronicportalimaging
AT elghamrawykamal verificationofsetuperrorsinexternalbeamradiationtherapyusingelectronicportalimaging
_version_ 1725855253087125504