Comparison of accuracy of two uncalibrated pulse contour cardiac output monitors in off-pump coronary artery bypass surgery patients using pulmonary artery catheter-thermodilution as a reference
Abstract Background Cardiac output (CO) is a key measure of adequacy of organ and tissue perfusion, especially in critically ill or complex surgical patients. CO monitoring technology continues to evolve. Recently developed CO monitors rely on unique algorithms based on pulse contour analysis of an...
Main Authors: | , , |
---|---|
Format: | Article |
Language: | English |
Published: |
BMC
2021-07-01
|
Series: | BMC Anesthesiology |
Subjects: | |
Online Access: | https://doi.org/10.1186/s12871-021-01415-5 |
id |
doaj-8d35e3cf6404469f9dc78dac5c0511c9 |
---|---|
record_format |
Article |
spelling |
doaj-8d35e3cf6404469f9dc78dac5c0511c92021-07-11T11:12:54ZengBMCBMC Anesthesiology1471-22532021-07-012111710.1186/s12871-021-01415-5Comparison of accuracy of two uncalibrated pulse contour cardiac output monitors in off-pump coronary artery bypass surgery patients using pulmonary artery catheter-thermodilution as a referenceRamakrishna Mukkamala0Benjamin A. Kohl1Aman Mahajan2Department of Anesthesiology and Perioperative Medicine and Department of Bioengineering, University of PittsburghDepartment of Anesthesiology, Thomas Jefferson UniversityDepartment of Anesthesiology and Perioperative Medicine and Department of Bioengineering, University of PittsburghAbstract Background Cardiac output (CO) is a key measure of adequacy of organ and tissue perfusion, especially in critically ill or complex surgical patients. CO monitoring technology continues to evolve. Recently developed CO monitors rely on unique algorithms based on pulse contour analysis of an arterial blood pressure (ABP) waveform. The objective of this investigation was to compare the accuracy of two monitors using different methods of pulse contour analysis – the Retia Argos device and the Edwards Vigileo-FloTrac device – with pulmonary artery catheter (PAC)-thermodilution as a reference. Methods Fifty-eight patients undergoing off-pump coronary artery bypass surgery formed the study cohort. A total of 572 triplets of CO measurements from each device – Argos, Vigileo-FloTrac (third generation), and thermodilution – were available before and after interventions (e.g., vasopressors, fluids, and inotropes). Bland–Altman analysis accounting for repeated measurements per subject and concordance analysis were applied to assess the accuracy of the CO values and intervention-induced CO changes of each pulse contour device against thermodilution. Cluster bootstrapping was employed to statistically compare the root-mean-squared-errors (RMSE = √(μ2 + σ2), where μ and σ are the Bland–Altman bias and precision errors) and concordance rates of the two devices. Results The RMSE (mean (95% confidence intervals)) for CO values was 1.16 (1.00–1.32) L/min for the Argos device and 1.54 (1.33–1.77) L/min for the Vigileo-FloTrac device; the concordance rate for intervention-induced CO changes was 87 (82–92)% for the Argos device and 72 (65–78)% for the Vigileo-FloTrac device; and the RMSE for the CO changes was 17 (15–19)% for the Argos device and 21 (19–23)% for the Vigileo-FloTrac device (p < 0.0167 for all comparisons). Conclusions In comparison with CO measured by the PAC, the Argos device proved to be more accurate than the Vigileo-FloTrac device in CO trending and absolute CO measurement in patients undergoing off-pump coronary artery bypass surgery.https://doi.org/10.1186/s12871-021-01415-5ArgosCardiac output monitoringFloTracPulse contour analysisThermodilution |
collection |
DOAJ |
language |
English |
format |
Article |
sources |
DOAJ |
author |
Ramakrishna Mukkamala Benjamin A. Kohl Aman Mahajan |
spellingShingle |
Ramakrishna Mukkamala Benjamin A. Kohl Aman Mahajan Comparison of accuracy of two uncalibrated pulse contour cardiac output monitors in off-pump coronary artery bypass surgery patients using pulmonary artery catheter-thermodilution as a reference BMC Anesthesiology Argos Cardiac output monitoring FloTrac Pulse contour analysis Thermodilution |
author_facet |
Ramakrishna Mukkamala Benjamin A. Kohl Aman Mahajan |
author_sort |
Ramakrishna Mukkamala |
title |
Comparison of accuracy of two uncalibrated pulse contour cardiac output monitors in off-pump coronary artery bypass surgery patients using pulmonary artery catheter-thermodilution as a reference |
title_short |
Comparison of accuracy of two uncalibrated pulse contour cardiac output monitors in off-pump coronary artery bypass surgery patients using pulmonary artery catheter-thermodilution as a reference |
title_full |
Comparison of accuracy of two uncalibrated pulse contour cardiac output monitors in off-pump coronary artery bypass surgery patients using pulmonary artery catheter-thermodilution as a reference |
title_fullStr |
Comparison of accuracy of two uncalibrated pulse contour cardiac output monitors in off-pump coronary artery bypass surgery patients using pulmonary artery catheter-thermodilution as a reference |
title_full_unstemmed |
Comparison of accuracy of two uncalibrated pulse contour cardiac output monitors in off-pump coronary artery bypass surgery patients using pulmonary artery catheter-thermodilution as a reference |
title_sort |
comparison of accuracy of two uncalibrated pulse contour cardiac output monitors in off-pump coronary artery bypass surgery patients using pulmonary artery catheter-thermodilution as a reference |
publisher |
BMC |
series |
BMC Anesthesiology |
issn |
1471-2253 |
publishDate |
2021-07-01 |
description |
Abstract Background Cardiac output (CO) is a key measure of adequacy of organ and tissue perfusion, especially in critically ill or complex surgical patients. CO monitoring technology continues to evolve. Recently developed CO monitors rely on unique algorithms based on pulse contour analysis of an arterial blood pressure (ABP) waveform. The objective of this investigation was to compare the accuracy of two monitors using different methods of pulse contour analysis – the Retia Argos device and the Edwards Vigileo-FloTrac device – with pulmonary artery catheter (PAC)-thermodilution as a reference. Methods Fifty-eight patients undergoing off-pump coronary artery bypass surgery formed the study cohort. A total of 572 triplets of CO measurements from each device – Argos, Vigileo-FloTrac (third generation), and thermodilution – were available before and after interventions (e.g., vasopressors, fluids, and inotropes). Bland–Altman analysis accounting for repeated measurements per subject and concordance analysis were applied to assess the accuracy of the CO values and intervention-induced CO changes of each pulse contour device against thermodilution. Cluster bootstrapping was employed to statistically compare the root-mean-squared-errors (RMSE = √(μ2 + σ2), where μ and σ are the Bland–Altman bias and precision errors) and concordance rates of the two devices. Results The RMSE (mean (95% confidence intervals)) for CO values was 1.16 (1.00–1.32) L/min for the Argos device and 1.54 (1.33–1.77) L/min for the Vigileo-FloTrac device; the concordance rate for intervention-induced CO changes was 87 (82–92)% for the Argos device and 72 (65–78)% for the Vigileo-FloTrac device; and the RMSE for the CO changes was 17 (15–19)% for the Argos device and 21 (19–23)% for the Vigileo-FloTrac device (p < 0.0167 for all comparisons). Conclusions In comparison with CO measured by the PAC, the Argos device proved to be more accurate than the Vigileo-FloTrac device in CO trending and absolute CO measurement in patients undergoing off-pump coronary artery bypass surgery. |
topic |
Argos Cardiac output monitoring FloTrac Pulse contour analysis Thermodilution |
url |
https://doi.org/10.1186/s12871-021-01415-5 |
work_keys_str_mv |
AT ramakrishnamukkamala comparisonofaccuracyoftwouncalibratedpulsecontourcardiacoutputmonitorsinoffpumpcoronaryarterybypasssurgerypatientsusingpulmonaryarterycatheterthermodilutionasareference AT benjaminakohl comparisonofaccuracyoftwouncalibratedpulsecontourcardiacoutputmonitorsinoffpumpcoronaryarterybypasssurgerypatientsusingpulmonaryarterycatheterthermodilutionasareference AT amanmahajan comparisonofaccuracyoftwouncalibratedpulsecontourcardiacoutputmonitorsinoffpumpcoronaryarterybypasssurgerypatientsusingpulmonaryarterycatheterthermodilutionasareference |
_version_ |
1721309245640540160 |