Imposed Work of Breathing for Flow Meters with In-Line versus Flow-Through Technique during Simulated Neonatal Breathing.

The ability to determine airflow during nasal CPAP (NCPAP) treatment without adding dead space or resistance would be useful when investigating the physiologic effects of different NCPAP systems on breathing. The aim of this study was to investigate the effect on pressure stability of different flow...

Full description

Bibliographic Details
Main Authors: Snorri Donaldsson, Markus Falk, Baldvin Jonsson, Thomas Drevhammar
Format: Article
Language:English
Published: Public Library of Science (PLoS) 2015-01-01
Series:PLoS ONE
Online Access:http://europepmc.org/articles/PMC4507850?pdf=render
id doaj-84a7a26416ff44639cacb94886ad6ed5
record_format Article
spelling doaj-84a7a26416ff44639cacb94886ad6ed52020-11-24T21:30:03ZengPublic Library of Science (PLoS)PLoS ONE1932-62032015-01-01107e013343210.1371/journal.pone.0133432Imposed Work of Breathing for Flow Meters with In-Line versus Flow-Through Technique during Simulated Neonatal Breathing.Snorri DonaldssonMarkus FalkBaldvin JonssonThomas DrevhammarThe ability to determine airflow during nasal CPAP (NCPAP) treatment without adding dead space or resistance would be useful when investigating the physiologic effects of different NCPAP systems on breathing. The aim of this study was to investigate the effect on pressure stability of different flow measuring devices at the in-line and flow-through position, using simulated neonatal breathing.Six different flow measure devices were evaluated by recording pressure changes and imposed work of breathing for breaths with 16 and 32 ml tidal volumes. The tests were performed initially with the devices in an in line position and with 5 and 10 L/min using flow through technique, without CPAP. The flow meters were then subsequently tested with an Infant Flow CPAP system at 3, 5 and 8 cm H2O pressure using flow through technique. The quality of the recorded signals was compared graphically.The resistance of the measuring devices generated pressure swings and imposed work of breathing. With bias flow, the resistance also generated CPAP pressure. Three of the devices had low resistance and generated no changes in pressure stability or CPAP pressure. The two devices intended for neonatal use had the highest measured resistance.The importance of pressure stability and increased work of breathing during non-invasive respiratory support are insufficiently studied. Clinical trials using flow-through technique have not focused on pressure stability. Our results indicate that a flow-through technique might be a way forward in obtaining a sufficiently high signal quality without the added effects of rebreathing and increased work of breathing. The results should stimulate further research and the development of equipment for dynamic flow measurements in neonates.http://europepmc.org/articles/PMC4507850?pdf=render
collection DOAJ
language English
format Article
sources DOAJ
author Snorri Donaldsson
Markus Falk
Baldvin Jonsson
Thomas Drevhammar
spellingShingle Snorri Donaldsson
Markus Falk
Baldvin Jonsson
Thomas Drevhammar
Imposed Work of Breathing for Flow Meters with In-Line versus Flow-Through Technique during Simulated Neonatal Breathing.
PLoS ONE
author_facet Snorri Donaldsson
Markus Falk
Baldvin Jonsson
Thomas Drevhammar
author_sort Snorri Donaldsson
title Imposed Work of Breathing for Flow Meters with In-Line versus Flow-Through Technique during Simulated Neonatal Breathing.
title_short Imposed Work of Breathing for Flow Meters with In-Line versus Flow-Through Technique during Simulated Neonatal Breathing.
title_full Imposed Work of Breathing for Flow Meters with In-Line versus Flow-Through Technique during Simulated Neonatal Breathing.
title_fullStr Imposed Work of Breathing for Flow Meters with In-Line versus Flow-Through Technique during Simulated Neonatal Breathing.
title_full_unstemmed Imposed Work of Breathing for Flow Meters with In-Line versus Flow-Through Technique during Simulated Neonatal Breathing.
title_sort imposed work of breathing for flow meters with in-line versus flow-through technique during simulated neonatal breathing.
publisher Public Library of Science (PLoS)
series PLoS ONE
issn 1932-6203
publishDate 2015-01-01
description The ability to determine airflow during nasal CPAP (NCPAP) treatment without adding dead space or resistance would be useful when investigating the physiologic effects of different NCPAP systems on breathing. The aim of this study was to investigate the effect on pressure stability of different flow measuring devices at the in-line and flow-through position, using simulated neonatal breathing.Six different flow measure devices were evaluated by recording pressure changes and imposed work of breathing for breaths with 16 and 32 ml tidal volumes. The tests were performed initially with the devices in an in line position and with 5 and 10 L/min using flow through technique, without CPAP. The flow meters were then subsequently tested with an Infant Flow CPAP system at 3, 5 and 8 cm H2O pressure using flow through technique. The quality of the recorded signals was compared graphically.The resistance of the measuring devices generated pressure swings and imposed work of breathing. With bias flow, the resistance also generated CPAP pressure. Three of the devices had low resistance and generated no changes in pressure stability or CPAP pressure. The two devices intended for neonatal use had the highest measured resistance.The importance of pressure stability and increased work of breathing during non-invasive respiratory support are insufficiently studied. Clinical trials using flow-through technique have not focused on pressure stability. Our results indicate that a flow-through technique might be a way forward in obtaining a sufficiently high signal quality without the added effects of rebreathing and increased work of breathing. The results should stimulate further research and the development of equipment for dynamic flow measurements in neonates.
url http://europepmc.org/articles/PMC4507850?pdf=render
work_keys_str_mv AT snorridonaldsson imposedworkofbreathingforflowmeterswithinlineversusflowthroughtechniqueduringsimulatedneonatalbreathing
AT markusfalk imposedworkofbreathingforflowmeterswithinlineversusflowthroughtechniqueduringsimulatedneonatalbreathing
AT baldvinjonsson imposedworkofbreathingforflowmeterswithinlineversusflowthroughtechniqueduringsimulatedneonatalbreathing
AT thomasdrevhammar imposedworkofbreathingforflowmeterswithinlineversusflowthroughtechniqueduringsimulatedneonatalbreathing
_version_ 1725964376492474368