A low-pass filter of 300 Hz improved the detection of pacemaker spike on remote and bedside electrocardiogram

Abstract. Background:. The current upper-frequency cutoff of 150 Hz sometimes causes loss of pacemaker spike and misdiagnosis. We hypothesized that low-pass filter (LPF) other than 150 Hz could improve the detection of pacemaker spike. This study aimed to examine the effect of different LPF on pacem...

Full description

Bibliographic Details
Main Authors: Jian Sun, Qiu-Feng Lu, Yan Zhao, Peng-Pai Zhang, Jun Wang, Qun-Shan Wang, Xiao-Hong Liu, Yi-Gang Li, Li-Min Chen
Format: Article
Language:English
Published: Wolters Kluwer 2019-03-01
Series:Chinese Medical Journal
Online Access:http://journals.lww.com/10.1097/CM9.0000000000000110
id doaj-82833f666d144b7bb0be3d60939e7bb1
record_format Article
spelling doaj-82833f666d144b7bb0be3d60939e7bb12020-12-02T07:45:48ZengWolters KluwerChinese Medical Journal0366-69992542-56412019-03-01132553454110.1097/CM9.0000000000000110201903050-00005A low-pass filter of 300 Hz improved the detection of pacemaker spike on remote and bedside electrocardiogramJian Sun0Qiu-Feng Lu1Yan Zhao2Peng-Pai Zhang3Jun Wang4Qun-Shan Wang5Xiao-Hong Liu6Yi-Gang Li7Li-Min Chen8Department of Cardiology, Xinhua Hospital, Shanghai Jiao Tong University School of Medicine, Shanghai 200092, China.Department of Cardiology, Xinhua Hospital, Shanghai Jiao Tong University School of Medicine, Shanghai 200092, China.Department of Cardiology, Xinhua Hospital, Shanghai Jiao Tong University School of Medicine, Shanghai 200092, China.Department of Cardiology, Xinhua Hospital, Shanghai Jiao Tong University School of Medicine, Shanghai 200092, China.Department of Cardiology, Xinhua Hospital, Shanghai Jiao Tong University School of Medicine, Shanghai 200092, China.Department of Cardiology, Xinhua Hospital, Shanghai Jiao Tong University School of Medicine, Shanghai 200092, China.Department of Cardiology, Xinhua Hospital, Shanghai Jiao Tong University School of Medicine, Shanghai 200092, China.Department of Cardiology, Xinhua Hospital, Shanghai Jiao Tong University School of Medicine, Shanghai 200092, China.Department of Cardiology, Xinhua Hospital, Shanghai Jiao Tong University School of Medicine, Shanghai 200092, China.Abstract. Background:. The current upper-frequency cutoff of 150 Hz sometimes causes loss of pacemaker spike and misdiagnosis. We hypothesized that low-pass filter (LPF) other than 150 Hz could improve the detection of pacemaker spike. This study aimed to examine the effect of different LPF on pacemaker spike detection in remote and bedside electrocardiogram (ECG). Methods:. Patients with permanent pacemaker implantation were included during routine follow-up. Standard 12-lead ECGs at 6 different upper-frequency cutoff (40, 100, 150, 200, 300, and 400 Hz) were collected. All ECGs were then transmitted to the remote clinic center. Ventricular and atrial pacing were analyzed by 2 independent medical practitioners. Results:. A total of 88 patients’ ECGs were analyzed (mean age 73.8 ± 10.2 years and 85 with dual-chamber pacemakers). About 75.3% (64/85) of patients were diagnosed as atrial pacing by pacemaker programming. Among 6 different upper-frequency cutoff, the 300 Hz turned out to perform best in detecting atrial-paced spike (area under the curve [AUC] = 0.73, 95% confidence interval [CI]: 0.61–0.84 vs. 0.56, 95% CI: 0.61–0.84 at 150 Hz; P = 0.002) on bedside ECGs. Using programming as the golden standard, the 300 Hz LPF has a sensitivity of 59.4%, specificity of 85.7%, positive predictive value of 92.7% and negative predictive value of 40.9% on bedside ECGs. As for the ventricular pacing, the 300 Hz LPF also had a higher accuracy (AUC = 0.93; 95% CI = 0.84–1.00) than that at 150 Hz (AUC = 0.86; 95% CI: 0.77–0.94; P < 0.001) in detecting ventricular-paced spike on bedside ECGs. The results of remote ECGs were similar with bedside ECGs. Conclusions:. A filter of 300 Hz cutoff may be recommended for ECG spike detection. With the recommended parameter, remote ECG can perform as well as bedside ECG.http://journals.lww.com/10.1097/CM9.0000000000000110
collection DOAJ
language English
format Article
sources DOAJ
author Jian Sun
Qiu-Feng Lu
Yan Zhao
Peng-Pai Zhang
Jun Wang
Qun-Shan Wang
Xiao-Hong Liu
Yi-Gang Li
Li-Min Chen
spellingShingle Jian Sun
Qiu-Feng Lu
Yan Zhao
Peng-Pai Zhang
Jun Wang
Qun-Shan Wang
Xiao-Hong Liu
Yi-Gang Li
Li-Min Chen
A low-pass filter of 300 Hz improved the detection of pacemaker spike on remote and bedside electrocardiogram
Chinese Medical Journal
author_facet Jian Sun
Qiu-Feng Lu
Yan Zhao
Peng-Pai Zhang
Jun Wang
Qun-Shan Wang
Xiao-Hong Liu
Yi-Gang Li
Li-Min Chen
author_sort Jian Sun
title A low-pass filter of 300 Hz improved the detection of pacemaker spike on remote and bedside electrocardiogram
title_short A low-pass filter of 300 Hz improved the detection of pacemaker spike on remote and bedside electrocardiogram
title_full A low-pass filter of 300 Hz improved the detection of pacemaker spike on remote and bedside electrocardiogram
title_fullStr A low-pass filter of 300 Hz improved the detection of pacemaker spike on remote and bedside electrocardiogram
title_full_unstemmed A low-pass filter of 300 Hz improved the detection of pacemaker spike on remote and bedside electrocardiogram
title_sort low-pass filter of 300 hz improved the detection of pacemaker spike on remote and bedside electrocardiogram
publisher Wolters Kluwer
series Chinese Medical Journal
issn 0366-6999
2542-5641
publishDate 2019-03-01
description Abstract. Background:. The current upper-frequency cutoff of 150 Hz sometimes causes loss of pacemaker spike and misdiagnosis. We hypothesized that low-pass filter (LPF) other than 150 Hz could improve the detection of pacemaker spike. This study aimed to examine the effect of different LPF on pacemaker spike detection in remote and bedside electrocardiogram (ECG). Methods:. Patients with permanent pacemaker implantation were included during routine follow-up. Standard 12-lead ECGs at 6 different upper-frequency cutoff (40, 100, 150, 200, 300, and 400 Hz) were collected. All ECGs were then transmitted to the remote clinic center. Ventricular and atrial pacing were analyzed by 2 independent medical practitioners. Results:. A total of 88 patients’ ECGs were analyzed (mean age 73.8 ± 10.2 years and 85 with dual-chamber pacemakers). About 75.3% (64/85) of patients were diagnosed as atrial pacing by pacemaker programming. Among 6 different upper-frequency cutoff, the 300 Hz turned out to perform best in detecting atrial-paced spike (area under the curve [AUC] = 0.73, 95% confidence interval [CI]: 0.61–0.84 vs. 0.56, 95% CI: 0.61–0.84 at 150 Hz; P = 0.002) on bedside ECGs. Using programming as the golden standard, the 300 Hz LPF has a sensitivity of 59.4%, specificity of 85.7%, positive predictive value of 92.7% and negative predictive value of 40.9% on bedside ECGs. As for the ventricular pacing, the 300 Hz LPF also had a higher accuracy (AUC = 0.93; 95% CI = 0.84–1.00) than that at 150 Hz (AUC = 0.86; 95% CI: 0.77–0.94; P < 0.001) in detecting ventricular-paced spike on bedside ECGs. The results of remote ECGs were similar with bedside ECGs. Conclusions:. A filter of 300 Hz cutoff may be recommended for ECG spike detection. With the recommended parameter, remote ECG can perform as well as bedside ECG.
url http://journals.lww.com/10.1097/CM9.0000000000000110
work_keys_str_mv AT jiansun alowpassfilterof300hzimprovedthedetectionofpacemakerspikeonremoteandbedsideelectrocardiogram
AT qiufenglu alowpassfilterof300hzimprovedthedetectionofpacemakerspikeonremoteandbedsideelectrocardiogram
AT yanzhao alowpassfilterof300hzimprovedthedetectionofpacemakerspikeonremoteandbedsideelectrocardiogram
AT pengpaizhang alowpassfilterof300hzimprovedthedetectionofpacemakerspikeonremoteandbedsideelectrocardiogram
AT junwang alowpassfilterof300hzimprovedthedetectionofpacemakerspikeonremoteandbedsideelectrocardiogram
AT qunshanwang alowpassfilterof300hzimprovedthedetectionofpacemakerspikeonremoteandbedsideelectrocardiogram
AT xiaohongliu alowpassfilterof300hzimprovedthedetectionofpacemakerspikeonremoteandbedsideelectrocardiogram
AT yigangli alowpassfilterof300hzimprovedthedetectionofpacemakerspikeonremoteandbedsideelectrocardiogram
AT liminchen alowpassfilterof300hzimprovedthedetectionofpacemakerspikeonremoteandbedsideelectrocardiogram
AT jiansun lowpassfilterof300hzimprovedthedetectionofpacemakerspikeonremoteandbedsideelectrocardiogram
AT qiufenglu lowpassfilterof300hzimprovedthedetectionofpacemakerspikeonremoteandbedsideelectrocardiogram
AT yanzhao lowpassfilterof300hzimprovedthedetectionofpacemakerspikeonremoteandbedsideelectrocardiogram
AT pengpaizhang lowpassfilterof300hzimprovedthedetectionofpacemakerspikeonremoteandbedsideelectrocardiogram
AT junwang lowpassfilterof300hzimprovedthedetectionofpacemakerspikeonremoteandbedsideelectrocardiogram
AT qunshanwang lowpassfilterof300hzimprovedthedetectionofpacemakerspikeonremoteandbedsideelectrocardiogram
AT xiaohongliu lowpassfilterof300hzimprovedthedetectionofpacemakerspikeonremoteandbedsideelectrocardiogram
AT yigangli lowpassfilterof300hzimprovedthedetectionofpacemakerspikeonremoteandbedsideelectrocardiogram
AT liminchen lowpassfilterof300hzimprovedthedetectionofpacemakerspikeonremoteandbedsideelectrocardiogram
_version_ 1724408235319885824