Evaluation of Digital Mammography Device Quality Control Measurements and Image Quality

Objective: In this study, we aimed to establish a standard protocol of measurements by taking quality control tests belonging to digital mammography devices widely used in our country. Breast cancer occurs in glandular tissue. For this reason, it is necessary to measure the mean glandular tissue val...

Full description

Bibliographic Details
Main Authors: Rukiye Şanlı, Özcan Gündoğdu
Format: Article
Language:English
Published: Kocaeli University 2018-09-01
Series:Kocaeli Üniversitesi Sağlık Bilimleri Dergisi
Subjects:
CNR
AGD
Online Access:http://dergipark.org.tr/kusbed/issue/39194/455426?publisher=kocaeli
id doaj-7fb99d475b39476286ae50dcd5bde1ce
record_format Article
spelling doaj-7fb99d475b39476286ae50dcd5bde1ce2020-11-24T21:28:52ZengKocaeli UniversityKocaeli Üniversitesi Sağlık Bilimleri Dergisi2149-85712018-09-0143889410.30934/kusbed.455426111Evaluation of Digital Mammography Device Quality Control Measurements and Image QualityRukiye ŞanlıÖzcan GündoğduObjective: In this study, we aimed to establish a standard protocol of measurements by taking quality control tests belonging to digital mammography devices widely used in our country. Breast cancer occurs in glandular tissue. For this reason, it is necessary to measure the mean glandular tissue value and follow the standard values. The accuracy of  the system has been tested  by considering whether the values specified are within the acceptable limit values specified by the EUREF standard. Performing quality control tests periodically will increase the quality of the image and reduce patient doses by minimizing repetitive work. Methods: The measurements were performed on a Fujifilm Amulet Innovality digital mammography system. Contrast detail Pro-Mamphantomand Pro-Mam software to simulate different breast tissues for measurements, Ray SafeXi detector for dosimetry, MTF test for image qualities, Image J software was used to measure the average pixel and standard deviation of the pattern the image. The PMMA phantoms of different thicknesses have the mean glandular tissue change (AGD), contrast-to-noiseratio (CNR), figure of merit (FOM) values and modular transfer function (MTF), detective quantum efficiency (DQE) and noise power spectrum (NPS) graphics are compared. HVL and CNR values of W/Rh filter and W/Al filter are compared. Results: The mean glandular dose was measured  as 0.92 mGy on a 45 mm PMMA plate in normal dose mode. When the W/Rh filter is used, the Contrast Noise Ratio (CNR) is 5.1, while the W/Al filter is 4.5. The FOM value with a 45 mm PMMA plate at 26 kVp was measured as 42.905 with the W/Rh filter. When there is a W/Al filter, it is measured as 50.115. Conclusion: The average glandular dose at 45 mm to the EUREF standard should be less than 2.5 mGy, the measured values are within the limit values. Breast dose, air kerma and AGD values increased while CNR and FOM values decreased with increasing breast thickness. When W/Rh filter is used, HVL, FOM and CNR values are higher because the X rays are more cured. Thus, it has been determined that it is possible to obtain a better image quality at a lower dose in the W/Rh combination.http://dergipark.org.tr/kusbed/issue/39194/455426?publisher=kocaeliPMMACNRAGDImageJPro-Mamfull field digital mammographyCNRAGDdijital mamografi cihazıPMMAImageJPro-Mamdijital mamografi cihazı
collection DOAJ
language English
format Article
sources DOAJ
author Rukiye Şanlı
Özcan Gündoğdu
spellingShingle Rukiye Şanlı
Özcan Gündoğdu
Evaluation of Digital Mammography Device Quality Control Measurements and Image Quality
Kocaeli Üniversitesi Sağlık Bilimleri Dergisi
PMMA
CNR
AGD
ImageJ
Pro-Mam
full field digital mammography
CNR
AGD
dijital mamografi cihazı
PMMA
ImageJ
Pro-Mam
dijital mamografi cihazı
author_facet Rukiye Şanlı
Özcan Gündoğdu
author_sort Rukiye Şanlı
title Evaluation of Digital Mammography Device Quality Control Measurements and Image Quality
title_short Evaluation of Digital Mammography Device Quality Control Measurements and Image Quality
title_full Evaluation of Digital Mammography Device Quality Control Measurements and Image Quality
title_fullStr Evaluation of Digital Mammography Device Quality Control Measurements and Image Quality
title_full_unstemmed Evaluation of Digital Mammography Device Quality Control Measurements and Image Quality
title_sort evaluation of digital mammography device quality control measurements and image quality
publisher Kocaeli University
series Kocaeli Üniversitesi Sağlık Bilimleri Dergisi
issn 2149-8571
publishDate 2018-09-01
description Objective: In this study, we aimed to establish a standard protocol of measurements by taking quality control tests belonging to digital mammography devices widely used in our country. Breast cancer occurs in glandular tissue. For this reason, it is necessary to measure the mean glandular tissue value and follow the standard values. The accuracy of  the system has been tested  by considering whether the values specified are within the acceptable limit values specified by the EUREF standard. Performing quality control tests periodically will increase the quality of the image and reduce patient doses by minimizing repetitive work. Methods: The measurements were performed on a Fujifilm Amulet Innovality digital mammography system. Contrast detail Pro-Mamphantomand Pro-Mam software to simulate different breast tissues for measurements, Ray SafeXi detector for dosimetry, MTF test for image qualities, Image J software was used to measure the average pixel and standard deviation of the pattern the image. The PMMA phantoms of different thicknesses have the mean glandular tissue change (AGD), contrast-to-noiseratio (CNR), figure of merit (FOM) values and modular transfer function (MTF), detective quantum efficiency (DQE) and noise power spectrum (NPS) graphics are compared. HVL and CNR values of W/Rh filter and W/Al filter are compared. Results: The mean glandular dose was measured  as 0.92 mGy on a 45 mm PMMA plate in normal dose mode. When the W/Rh filter is used, the Contrast Noise Ratio (CNR) is 5.1, while the W/Al filter is 4.5. The FOM value with a 45 mm PMMA plate at 26 kVp was measured as 42.905 with the W/Rh filter. When there is a W/Al filter, it is measured as 50.115. Conclusion: The average glandular dose at 45 mm to the EUREF standard should be less than 2.5 mGy, the measured values are within the limit values. Breast dose, air kerma and AGD values increased while CNR and FOM values decreased with increasing breast thickness. When W/Rh filter is used, HVL, FOM and CNR values are higher because the X rays are more cured. Thus, it has been determined that it is possible to obtain a better image quality at a lower dose in the W/Rh combination.
topic PMMA
CNR
AGD
ImageJ
Pro-Mam
full field digital mammography
CNR
AGD
dijital mamografi cihazı
PMMA
ImageJ
Pro-Mam
dijital mamografi cihazı
url http://dergipark.org.tr/kusbed/issue/39194/455426?publisher=kocaeli
work_keys_str_mv AT rukiyesanlı evaluationofdigitalmammographydevicequalitycontrolmeasurementsandimagequality
AT ozcangundogdu evaluationofdigitalmammographydevicequalitycontrolmeasurementsandimagequality
_version_ 1725968847072133120