Design of mini phantom and measurement of cobalt-60 beam data parameters

Low cost mini phantoms were fabricated indigenously with different water equivalent material such as polymethyl methacrylate and Bee′s wax of different shapes (with dome top surface and flat top surface). The beam parameters of the Co-60 machine, such as head scatter correction factor (S&...

Full description

Bibliographic Details
Main Authors: Senthilkumar S, Ramakrishnan V
Format: Article
Language:English
Published: Wolters Kluwer Medknow Publications 2008-01-01
Series:Journal of Medical Physics
Subjects:
Online Access:http://www.jmp.org.in/article.asp?issn=0971-6203;year=2008;volume=33;issue=3;spage=100;epage=107;aulast=Senthilkumar
id doaj-7e0d3c9ed250446fb6316e7bb23967f6
record_format Article
spelling doaj-7e0d3c9ed250446fb6316e7bb23967f62020-11-24T21:19:14ZengWolters Kluwer Medknow PublicationsJournal of Medical Physics0971-62031998-39132008-01-01333100107Design of mini phantom and measurement of cobalt-60 beam data parametersSenthilkumar SRamakrishnan VLow cost mini phantoms were fabricated indigenously with different water equivalent material such as polymethyl methacrylate and Bee&#x2032;s wax of different shapes (with dome top surface and flat top surface). The beam parameters of the Co-60 machine, such as head scatter correction factor (S<sub> h</sub> ), phantom scatter correction factor (S<sub> P</sub> ), total scatter correction factor (S<sub> C,P</sub> ), collimator exchange effect were measured. Output ratio measurements were taken for both mini phantom and water phantom for different square and rectangular field sizes. Normalized output ratios were compared with ESTRO published values and (Storchi and Van Gasteren) S and G data. The percentage of variation between the measured and the literature values is about 0.7&#x0025;. Collimator exchange effect were measured for water and mini phantom for different field size, were compared with ESTRO value. This was found to be 0.5&#x0025; and 1.0&#x0025; respectively. Phantom scatter correction factors were calculated for square and rectangular filed sizes; this was compared with ESTRO values, found to be 0.7&#x0025; for square and 1.0&#x0025; for rectangular filed size. It was also noted that there were no appreciable variation observed in ion chamber readings of different materials of mini phantoms for dome and flat surfaces. Mini phantom measurements were done for all types of phantoms and the measured values were compared with the existing data and they were in good agreement with the published values.http://www.jmp.org.in/article.asp?issn=0971-6203;year=2008;volume=33;issue=3;spage=100;epage=107;aulast=SenthilkumarCollimator scatter correction factors (S<sub> C</sub> )collimator exchange effectmini phantomphantom scatter correction factors (S<sub> P</sub> )total scatter correction factors (S<sub> CP</sub> )
collection DOAJ
language English
format Article
sources DOAJ
author Senthilkumar S
Ramakrishnan V
spellingShingle Senthilkumar S
Ramakrishnan V
Design of mini phantom and measurement of cobalt-60 beam data parameters
Journal of Medical Physics
Collimator scatter correction factors (S<sub> C</sub> )
collimator exchange effect
mini phantom
phantom scatter correction factors (S<sub> P</sub> )
total scatter correction factors (S<sub> C
P</sub> )
author_facet Senthilkumar S
Ramakrishnan V
author_sort Senthilkumar S
title Design of mini phantom and measurement of cobalt-60 beam data parameters
title_short Design of mini phantom and measurement of cobalt-60 beam data parameters
title_full Design of mini phantom and measurement of cobalt-60 beam data parameters
title_fullStr Design of mini phantom and measurement of cobalt-60 beam data parameters
title_full_unstemmed Design of mini phantom and measurement of cobalt-60 beam data parameters
title_sort design of mini phantom and measurement of cobalt-60 beam data parameters
publisher Wolters Kluwer Medknow Publications
series Journal of Medical Physics
issn 0971-6203
1998-3913
publishDate 2008-01-01
description Low cost mini phantoms were fabricated indigenously with different water equivalent material such as polymethyl methacrylate and Bee&#x2032;s wax of different shapes (with dome top surface and flat top surface). The beam parameters of the Co-60 machine, such as head scatter correction factor (S<sub> h</sub> ), phantom scatter correction factor (S<sub> P</sub> ), total scatter correction factor (S<sub> C,P</sub> ), collimator exchange effect were measured. Output ratio measurements were taken for both mini phantom and water phantom for different square and rectangular field sizes. Normalized output ratios were compared with ESTRO published values and (Storchi and Van Gasteren) S and G data. The percentage of variation between the measured and the literature values is about 0.7&#x0025;. Collimator exchange effect were measured for water and mini phantom for different field size, were compared with ESTRO value. This was found to be 0.5&#x0025; and 1.0&#x0025; respectively. Phantom scatter correction factors were calculated for square and rectangular filed sizes; this was compared with ESTRO values, found to be 0.7&#x0025; for square and 1.0&#x0025; for rectangular filed size. It was also noted that there were no appreciable variation observed in ion chamber readings of different materials of mini phantoms for dome and flat surfaces. Mini phantom measurements were done for all types of phantoms and the measured values were compared with the existing data and they were in good agreement with the published values.
topic Collimator scatter correction factors (S<sub> C</sub> )
collimator exchange effect
mini phantom
phantom scatter correction factors (S<sub> P</sub> )
total scatter correction factors (S<sub> C
P</sub> )
url http://www.jmp.org.in/article.asp?issn=0971-6203;year=2008;volume=33;issue=3;spage=100;epage=107;aulast=Senthilkumar
work_keys_str_mv AT senthilkumars designofminiphantomandmeasurementofcobalt60beamdataparameters
AT ramakrishnanv designofminiphantomandmeasurementofcobalt60beamdataparameters
_version_ 1726006352143187968