Differences in Central Corneal Thickness between Spectral Domain-Optical Coherence Tomography and Ultrasound Pachymetry in Patients with Dry Eye Disease

Purpose. To compare central corneal thickness (CCT) values via Spectral Domain-Optical Coherence Tomography (SD-OCT) and ultrasonic pachymetry in patients with severe dry eye disease (DED) to determine the level of agreement between these 2 methods. Methods. The paired samples t-test was used to com...

Full description

Bibliographic Details
Main Authors: Ali Riza Cenk Celebi, G. Ertugrul Mirza
Format: Article
Language:English
Published: Hindawi Limited 2016-01-01
Series:Journal of Ophthalmology
Online Access:http://dx.doi.org/10.1155/2016/2623719
id doaj-77c76fcb3f6041cfb2a26a756b1b42ad
record_format Article
spelling doaj-77c76fcb3f6041cfb2a26a756b1b42ad2020-11-24T23:15:02ZengHindawi LimitedJournal of Ophthalmology2090-004X2090-00582016-01-01201610.1155/2016/26237192623719Differences in Central Corneal Thickness between Spectral Domain-Optical Coherence Tomography and Ultrasound Pachymetry in Patients with Dry Eye DiseaseAli Riza Cenk Celebi0G. Ertugrul Mirza1Department of Ophthalmology, Acibadem University School of Medicine, 34303 Istanbul, TurkeyDepartment of Ophthalmology, Erciyes University School of Medicine, 38030 Kayseri, TurkeyPurpose. To compare central corneal thickness (CCT) values via Spectral Domain-Optical Coherence Tomography (SD-OCT) and ultrasonic pachymetry in patients with severe dry eye disease (DED) to determine the level of agreement between these 2 methods. Methods. The paired samples t-test was used to compare CCT values in severe DED patients. Matching analysis between methods was performed using intraclass correlation coefficient (ICC). Intrasession reliability of the measurement methods was calculated via the concordance correlation coefficient (CCC), variation equivalent, and Pearson’s correlation coefficient. The Bland-Altman procedure was used to graphically represent the differences between CCT values. Results. The study included 56 eyes of 24 female and 4 male patients. Mean age of the patients was 50.9±11.3 years. Mean CCT via Cirrus SD-OCT was 523.82±30.98 μm versus 530.050±31.85 μm via ultrasonic pachymetry (paired samples t-test, P<0.001). The Bland-Altman plot showed good agreement between the examiners. The ICC for repeatability was 0.974. The CCC between the 2 methods’ CCT values was 0.973. The variation equivalent was 0.976 and Pearson’s correlation coefficient was 99.3%, which also indicated high correlation between the 2 methods’ measurements. Conclusions. The present findings show that in patients with severe DED Cirrus SD-OCT provides reliable intraobserver CCT values.http://dx.doi.org/10.1155/2016/2623719
collection DOAJ
language English
format Article
sources DOAJ
author Ali Riza Cenk Celebi
G. Ertugrul Mirza
spellingShingle Ali Riza Cenk Celebi
G. Ertugrul Mirza
Differences in Central Corneal Thickness between Spectral Domain-Optical Coherence Tomography and Ultrasound Pachymetry in Patients with Dry Eye Disease
Journal of Ophthalmology
author_facet Ali Riza Cenk Celebi
G. Ertugrul Mirza
author_sort Ali Riza Cenk Celebi
title Differences in Central Corneal Thickness between Spectral Domain-Optical Coherence Tomography and Ultrasound Pachymetry in Patients with Dry Eye Disease
title_short Differences in Central Corneal Thickness between Spectral Domain-Optical Coherence Tomography and Ultrasound Pachymetry in Patients with Dry Eye Disease
title_full Differences in Central Corneal Thickness between Spectral Domain-Optical Coherence Tomography and Ultrasound Pachymetry in Patients with Dry Eye Disease
title_fullStr Differences in Central Corneal Thickness between Spectral Domain-Optical Coherence Tomography and Ultrasound Pachymetry in Patients with Dry Eye Disease
title_full_unstemmed Differences in Central Corneal Thickness between Spectral Domain-Optical Coherence Tomography and Ultrasound Pachymetry in Patients with Dry Eye Disease
title_sort differences in central corneal thickness between spectral domain-optical coherence tomography and ultrasound pachymetry in patients with dry eye disease
publisher Hindawi Limited
series Journal of Ophthalmology
issn 2090-004X
2090-0058
publishDate 2016-01-01
description Purpose. To compare central corneal thickness (CCT) values via Spectral Domain-Optical Coherence Tomography (SD-OCT) and ultrasonic pachymetry in patients with severe dry eye disease (DED) to determine the level of agreement between these 2 methods. Methods. The paired samples t-test was used to compare CCT values in severe DED patients. Matching analysis between methods was performed using intraclass correlation coefficient (ICC). Intrasession reliability of the measurement methods was calculated via the concordance correlation coefficient (CCC), variation equivalent, and Pearson’s correlation coefficient. The Bland-Altman procedure was used to graphically represent the differences between CCT values. Results. The study included 56 eyes of 24 female and 4 male patients. Mean age of the patients was 50.9±11.3 years. Mean CCT via Cirrus SD-OCT was 523.82±30.98 μm versus 530.050±31.85 μm via ultrasonic pachymetry (paired samples t-test, P<0.001). The Bland-Altman plot showed good agreement between the examiners. The ICC for repeatability was 0.974. The CCC between the 2 methods’ CCT values was 0.973. The variation equivalent was 0.976 and Pearson’s correlation coefficient was 99.3%, which also indicated high correlation between the 2 methods’ measurements. Conclusions. The present findings show that in patients with severe DED Cirrus SD-OCT provides reliable intraobserver CCT values.
url http://dx.doi.org/10.1155/2016/2623719
work_keys_str_mv AT alirizacenkcelebi differencesincentralcornealthicknessbetweenspectraldomainopticalcoherencetomographyandultrasoundpachymetryinpatientswithdryeyedisease
AT gertugrulmirza differencesincentralcornealthicknessbetweenspectraldomainopticalcoherencetomographyandultrasoundpachymetryinpatientswithdryeyedisease
_version_ 1725592374699098112