Low Dose Rate X-Ray Radiation Fields Implementation for Study of Energy Response of the Dosimeters Based on Scintillation Detectors

Near background low dose rate measurements are important part of the environmental radiation monitoring. It is possible to fulfill energy response verification for the high sensitive dosimeters based on inorganic scintillation detectors in low energy region when creating reference X-ray fields with...

Full description

Bibliographic Details
Main Authors: A. V. Novichenko, R. V. Lukashevich, K. G. Senkovsky
Format: Article
Language:English
Published: Belarusian National Technical University 2019-06-01
Series:Pribory i Metody Izmerenij
Subjects:
Online Access:https://pimi.bntu.by/jour/article/view/436
id doaj-71d15081bbd649f3a1be95662198f3eb
record_format Article
spelling doaj-71d15081bbd649f3a1be95662198f3eb2021-07-29T08:39:11ZengBelarusian National Technical UniversityPribory i Metody Izmerenij2220-95062414-04732019-06-0110212813710.21122/2220-9506-2019-10-2-128-137346Low Dose Rate X-Ray Radiation Fields Implementation for Study of Energy Response of the Dosimeters Based on Scintillation DetectorsA. V. Novichenko0R. V. Lukashevich1K. G. Senkovsky2Scientific production unitary enterprise «АТОМТЕХ»Scientific production unitary enterprise «АТОМТЕХ»Scientific production unitary enterprise «АТОМТЕХ»Near background low dose rate measurements are important part of the environmental radiation monitoring. It is possible to fulfill energy response verification for the high sensitive dosimeters based on inorganic scintillation detectors in low energy region when creating reference X-ray fields with dose rates up to 5 µSv/h. The aim of this work was to create and study reference X-ray fields with low dose rate and narrow spectrum in the energy range from 15 to 250 keV using high-purity metal filters as a part of X-ray irradiator of AT300 X-ray calibration facility.To determine the main characteristics of created X-ray fields highly sensitive comparators of photon radiation based on NaI(Tl) scintillation detectors was used. The comparators were developed in “ATOMTEX”. To verify comparators energy response the reference AT5350/1 dosimeter and ionization chamber TM32003 with sensitive volume 10000 cm3 were used.Characteristics of X-ray fields that were created on the AT300 X-ray calibration facility to verify the energy response of high sensitive dosimeters based on scintillation detectors were investigated. The possibility to calibrate high sensitive dosimetric measuring instruments based on scintillation detectors in the energy range up to 250 keV in X-ray beams was shown.https://pimi.bntu.by/jour/article/view/436dose ratecomparatorscintillation detectorenergy responsex-ray tube
collection DOAJ
language English
format Article
sources DOAJ
author A. V. Novichenko
R. V. Lukashevich
K. G. Senkovsky
spellingShingle A. V. Novichenko
R. V. Lukashevich
K. G. Senkovsky
Low Dose Rate X-Ray Radiation Fields Implementation for Study of Energy Response of the Dosimeters Based on Scintillation Detectors
Pribory i Metody Izmerenij
dose rate
comparator
scintillation detector
energy response
x-ray tube
author_facet A. V. Novichenko
R. V. Lukashevich
K. G. Senkovsky
author_sort A. V. Novichenko
title Low Dose Rate X-Ray Radiation Fields Implementation for Study of Energy Response of the Dosimeters Based on Scintillation Detectors
title_short Low Dose Rate X-Ray Radiation Fields Implementation for Study of Energy Response of the Dosimeters Based on Scintillation Detectors
title_full Low Dose Rate X-Ray Radiation Fields Implementation for Study of Energy Response of the Dosimeters Based on Scintillation Detectors
title_fullStr Low Dose Rate X-Ray Radiation Fields Implementation for Study of Energy Response of the Dosimeters Based on Scintillation Detectors
title_full_unstemmed Low Dose Rate X-Ray Radiation Fields Implementation for Study of Energy Response of the Dosimeters Based on Scintillation Detectors
title_sort low dose rate x-ray radiation fields implementation for study of energy response of the dosimeters based on scintillation detectors
publisher Belarusian National Technical University
series Pribory i Metody Izmerenij
issn 2220-9506
2414-0473
publishDate 2019-06-01
description Near background low dose rate measurements are important part of the environmental radiation monitoring. It is possible to fulfill energy response verification for the high sensitive dosimeters based on inorganic scintillation detectors in low energy region when creating reference X-ray fields with dose rates up to 5 µSv/h. The aim of this work was to create and study reference X-ray fields with low dose rate and narrow spectrum in the energy range from 15 to 250 keV using high-purity metal filters as a part of X-ray irradiator of AT300 X-ray calibration facility.To determine the main characteristics of created X-ray fields highly sensitive comparators of photon radiation based on NaI(Tl) scintillation detectors was used. The comparators were developed in “ATOMTEX”. To verify comparators energy response the reference AT5350/1 dosimeter and ionization chamber TM32003 with sensitive volume 10000 cm3 were used.Characteristics of X-ray fields that were created on the AT300 X-ray calibration facility to verify the energy response of high sensitive dosimeters based on scintillation detectors were investigated. The possibility to calibrate high sensitive dosimetric measuring instruments based on scintillation detectors in the energy range up to 250 keV in X-ray beams was shown.
topic dose rate
comparator
scintillation detector
energy response
x-ray tube
url https://pimi.bntu.by/jour/article/view/436
work_keys_str_mv AT avnovichenko lowdoseratexrayradiationfieldsimplementationforstudyofenergyresponseofthedosimetersbasedonscintillationdetectors
AT rvlukashevich lowdoseratexrayradiationfieldsimplementationforstudyofenergyresponseofthedosimetersbasedonscintillationdetectors
AT kgsenkovsky lowdoseratexrayradiationfieldsimplementationforstudyofenergyresponseofthedosimetersbasedonscintillationdetectors
_version_ 1721252912424812544