Analysis and evaluation of temperature field and experiment for magnetorheological fluid testing devices

When a large amount of heat is produced during the operation of a MRF (magnetorheological fluid) testing device, the temperature of the device will increase, which will in turn affect the characteristics of the MRF. Exploring the temperature field characteristics of the MRF yield stress testing devi...

Full description

Bibliographic Details
Main Authors: Hao Xiong, Yiping Luo, Dongsheng Ji, Hongjuan Ren, Dan Wei, Wentao Liu
Format: Article
Language:English
Published: SAGE Publishing 2021-04-01
Series:Advances in Mechanical Engineering
Online Access:https://doi.org/10.1177/16878140211012538
Description
Summary:When a large amount of heat is produced during the operation of a MRF (magnetorheological fluid) testing device, the temperature of the device will increase, which will in turn affect the characteristics of the MRF. Exploring the temperature field characteristics of the MRF yield stress testing device is necessary to improve the accuracy of the device. In this study, first, the yield stress testing device is designed, and then its temperature field model, including enameled wire and assembly gap, is established. Second, simulation software is used to simulate the temperature field change. Finally, a test platform is developed to test the simulation results, especially for two factors, namely, thermal conductivity of the coil winding and assembly gap, which demonstrate considerable influence. Comprehensive thermal conductivity and assembly clearance are determined, and the optimum temperature field of the device for measuring yield stress is resulted.
ISSN:1687-8140