ASSESSMENT OF THE TENSION CONDITION THAT OCCURS WHEN THE BODY IS IMMOBILIZED IN A BOSTON-TYPE PLASTIC CORSET
The paper aims to make a theoretical study - using the finite element method, on the local elastic stresses and deformations that appear in the sagital plane under the use of a Boston corset made of polypropylene.The studies prove the fact that the elastic stress state appearing in the xOy plane in...
Main Authors: | , , , , , |
---|---|
Format: | Article |
Language: | English |
Published: |
Politehnium Publishing House
2017-12-01
|
Series: | European Journal of Materials Science and Engineering |
Subjects: | |
Online Access: | http://ejmse.tuiasi.ro/articles/EJMSE_02_04_06_Nanu.pdf |
id |
doaj-6dbd1b8c8d024f1492f6e6bebeaaf1f2 |
---|---|
record_format |
Article |
spelling |
doaj-6dbd1b8c8d024f1492f6e6bebeaaf1f22020-11-25T01:20:12ZengPolitehnium Publishing HouseEuropean Journal of Materials Science and Engineering2537-43382017-12-0124139143ASSESSMENT OF THE TENSION CONDITION THAT OCCURS WHEN THE BODY IS IMMOBILIZED IN A BOSTON-TYPE PLASTIC CORSETConstantin NANU0Ion POEATA1Lucian EVA2Bogdan Florin TOMA3Raluca MUNTEANU4 Alin Marian CAZAC5Clinical Emergency Hospitals, P. Neamt, 610003, Romania University of Medicine & Pharm. “Grigore T Popa”, Fac. of Medicine, Iasi, 700115, Romania Clinical Emergency Hospitals “Prof. Dr. Nicolae. Oblu”: Iasi, 700050, Romania Clinical Emergency Hospitals “Prof. Dr. Nicolae. Oblu”: Iasi, 700050, Romania Clinical Emergency Hospitals “Prof. Dr. Nicolae. Oblu”: Iasi, 700050, Romania Technical University “Gheorghe Asachi” of Iasi-Romania, Department of Materials Engineering and Industrial Safety, Blvd. Mangeron, No. 69, 700050, Iasi, RomaniaThe paper aims to make a theoretical study - using the finite element method, on the local elastic stresses and deformations that appear in the sagital plane under the use of a Boston corset made of polypropylene.The studies prove the fact that the elastic stress state appearing in the xOy plane in the case of the flexion motion is higher whit 18% than the elastic stress state appearing in the same plane in the case of the extension movement and the higher elastic displacements are recorded in the flexion motion. The occurrence of local deformities at the level of the fractured vertebrae may lead to the wrong splicing of the vertebrae, which causes changes in the geometry of the spine.http://ejmse.tuiasi.ro/articles/EJMSE_02_04_06_Nanu.pdfspineboston orthosisvertebraefinite element |
collection |
DOAJ |
language |
English |
format |
Article |
sources |
DOAJ |
author |
Constantin NANU Ion POEATA Lucian EVA Bogdan Florin TOMA Raluca MUNTEANU Alin Marian CAZAC |
spellingShingle |
Constantin NANU Ion POEATA Lucian EVA Bogdan Florin TOMA Raluca MUNTEANU Alin Marian CAZAC ASSESSMENT OF THE TENSION CONDITION THAT OCCURS WHEN THE BODY IS IMMOBILIZED IN A BOSTON-TYPE PLASTIC CORSET European Journal of Materials Science and Engineering spine boston orthosis vertebrae finite element |
author_facet |
Constantin NANU Ion POEATA Lucian EVA Bogdan Florin TOMA Raluca MUNTEANU Alin Marian CAZAC |
author_sort |
Constantin NANU |
title |
ASSESSMENT OF THE TENSION CONDITION THAT OCCURS WHEN THE BODY IS IMMOBILIZED IN A BOSTON-TYPE PLASTIC CORSET |
title_short |
ASSESSMENT OF THE TENSION CONDITION THAT OCCURS WHEN THE BODY IS IMMOBILIZED IN A BOSTON-TYPE PLASTIC CORSET |
title_full |
ASSESSMENT OF THE TENSION CONDITION THAT OCCURS WHEN THE BODY IS IMMOBILIZED IN A BOSTON-TYPE PLASTIC CORSET |
title_fullStr |
ASSESSMENT OF THE TENSION CONDITION THAT OCCURS WHEN THE BODY IS IMMOBILIZED IN A BOSTON-TYPE PLASTIC CORSET |
title_full_unstemmed |
ASSESSMENT OF THE TENSION CONDITION THAT OCCURS WHEN THE BODY IS IMMOBILIZED IN A BOSTON-TYPE PLASTIC CORSET |
title_sort |
assessment of the tension condition that occurs when the body is immobilized in a boston-type plastic corset |
publisher |
Politehnium Publishing House |
series |
European Journal of Materials Science and Engineering |
issn |
2537-4338 |
publishDate |
2017-12-01 |
description |
The paper aims to make a theoretical study - using the finite element method, on the local elastic stresses and deformations that appear in the sagital plane under the use of a Boston corset made of polypropylene.The studies prove the fact that the elastic stress state appearing in the xOy plane in the case of the flexion motion is higher whit 18% than the elastic stress state appearing in the same plane in the case of the extension movement and the higher elastic displacements are recorded in the flexion motion. The occurrence of local deformities at the level of the fractured vertebrae may lead to the wrong splicing of the vertebrae, which causes changes in the geometry of the spine. |
topic |
spine boston orthosis vertebrae finite element |
url |
http://ejmse.tuiasi.ro/articles/EJMSE_02_04_06_Nanu.pdf |
work_keys_str_mv |
AT constantinnanu assessmentofthetensionconditionthatoccurswhenthebodyisimmobilizedinabostontypeplasticcorset AT ionpoeata assessmentofthetensionconditionthatoccurswhenthebodyisimmobilizedinabostontypeplasticcorset AT lucianeva assessmentofthetensionconditionthatoccurswhenthebodyisimmobilizedinabostontypeplasticcorset AT bogdanflorintoma assessmentofthetensionconditionthatoccurswhenthebodyisimmobilizedinabostontypeplasticcorset AT ralucamunteanu assessmentofthetensionconditionthatoccurswhenthebodyisimmobilizedinabostontypeplasticcorset AT alinmariancazac assessmentofthetensionconditionthatoccurswhenthebodyisimmobilizedinabostontypeplasticcorset |
_version_ |
1725134904633589760 |