Tri-axial square Helmholtz coil system at the Alibag Magnetic Observatory: upgraded to a magnetic sensor calibration facility

A tri-axial square Helmholtz coil system for the study of palaeomagnetic studies, manufactured by GEOFYZIKA (former Czechoslovakia), was successfully commissioned at the Alibag Magnetic Observatory (IAGA code: ABG) in the year 1985. This system was used for a few years, after which the system en...

Full description

Bibliographic Details
Main Authors: P. Mahavarkar, J. John, V. Dhapre, V. Dongre, S. Labde
Format: Article
Language:English
Published: Copernicus Publications 2018-04-01
Series:Geoscientific Instrumentation, Methods and Data Systems
Online Access:https://www.geosci-instrum-method-data-syst.net/7/143/2018/gi-7-143-2018.pdf
id doaj-6c1973fc3e334cf896a208969822e227
record_format Article
spelling doaj-6c1973fc3e334cf896a208969822e2272020-11-24T21:29:48ZengCopernicus PublicationsGeoscientific Instrumentation, Methods and Data Systems2193-08562193-08642018-04-01714314910.5194/gi-7-143-2018Tri-axial square Helmholtz coil system at the Alibag Magnetic Observatory: upgraded to a magnetic sensor calibration facilityP. Mahavarkar0J. John1V. Dhapre2V. Dongre3S. Labde4Indian Institute of Geomagnetism, Dept. of Science & Technology, Govt. of India, Navi Mumbai 410218, IndiaIndian Institute of Geomagnetism, Dept. of Science & Technology, Govt. of India, Navi Mumbai 410218, IndiaIndian Institute of Geomagnetism, Dept. of Science & Technology, Govt. of India, Navi Mumbai 410218, IndiaIndian Institute of Geomagnetism, Dept. of Science & Technology, Govt. of India, Navi Mumbai 410218, IndiaIndian Institute of Geomagnetism, Dept. of Science & Technology, Govt. of India, Navi Mumbai 410218, IndiaA tri-axial square Helmholtz coil system for the study of palaeomagnetic studies, manufactured by GEOFYZIKA (former Czechoslovakia), was successfully commissioned at the Alibag Magnetic Observatory (IAGA code: ABG) in the year 1985. This system was used for a few years, after which the system encountered technical problems with the control unit. Rectification of the unit could not be undertaken, as the information document related to this system was not available, and as a result the system had been lying in an unused state for a long time, until 2015, when the system was recommissioned and upgraded to a test facility for calibrating the magnetometer sensors. We have upgraded the system with a constant current source and a data-logging unit. Both of these units have been designed and developed in the institute laboratory. Also, re-measurements of the existing system have been made thoroughly. The upgraded system is semi-automatic, enabling non-specialists to operate it after a brief period of instruction. This facility is now widely used at the parent institute and external institutions to calibrate magnetometers and it also serves as a national facility. Here the design of this system with the calibration results for the space-borne fluxgate magnetometers is presented.https://www.geosci-instrum-method-data-syst.net/7/143/2018/gi-7-143-2018.pdf
collection DOAJ
language English
format Article
sources DOAJ
author P. Mahavarkar
J. John
V. Dhapre
V. Dongre
S. Labde
spellingShingle P. Mahavarkar
J. John
V. Dhapre
V. Dongre
S. Labde
Tri-axial square Helmholtz coil system at the Alibag Magnetic Observatory: upgraded to a magnetic sensor calibration facility
Geoscientific Instrumentation, Methods and Data Systems
author_facet P. Mahavarkar
J. John
V. Dhapre
V. Dongre
S. Labde
author_sort P. Mahavarkar
title Tri-axial square Helmholtz coil system at the Alibag Magnetic Observatory: upgraded to a magnetic sensor calibration facility
title_short Tri-axial square Helmholtz coil system at the Alibag Magnetic Observatory: upgraded to a magnetic sensor calibration facility
title_full Tri-axial square Helmholtz coil system at the Alibag Magnetic Observatory: upgraded to a magnetic sensor calibration facility
title_fullStr Tri-axial square Helmholtz coil system at the Alibag Magnetic Observatory: upgraded to a magnetic sensor calibration facility
title_full_unstemmed Tri-axial square Helmholtz coil system at the Alibag Magnetic Observatory: upgraded to a magnetic sensor calibration facility
title_sort tri-axial square helmholtz coil system at the alibag magnetic observatory: upgraded to a magnetic sensor calibration facility
publisher Copernicus Publications
series Geoscientific Instrumentation, Methods and Data Systems
issn 2193-0856
2193-0864
publishDate 2018-04-01
description A tri-axial square Helmholtz coil system for the study of palaeomagnetic studies, manufactured by GEOFYZIKA (former Czechoslovakia), was successfully commissioned at the Alibag Magnetic Observatory (IAGA code: ABG) in the year 1985. This system was used for a few years, after which the system encountered technical problems with the control unit. Rectification of the unit could not be undertaken, as the information document related to this system was not available, and as a result the system had been lying in an unused state for a long time, until 2015, when the system was recommissioned and upgraded to a test facility for calibrating the magnetometer sensors. We have upgraded the system with a constant current source and a data-logging unit. Both of these units have been designed and developed in the institute laboratory. Also, re-measurements of the existing system have been made thoroughly. The upgraded system is semi-automatic, enabling non-specialists to operate it after a brief period of instruction. This facility is now widely used at the parent institute and external institutions to calibrate magnetometers and it also serves as a national facility. Here the design of this system with the calibration results for the space-borne fluxgate magnetometers is presented.
url https://www.geosci-instrum-method-data-syst.net/7/143/2018/gi-7-143-2018.pdf
work_keys_str_mv AT pmahavarkar triaxialsquarehelmholtzcoilsystematthealibagmagneticobservatoryupgradedtoamagneticsensorcalibrationfacility
AT jjohn triaxialsquarehelmholtzcoilsystematthealibagmagneticobservatoryupgradedtoamagneticsensorcalibrationfacility
AT vdhapre triaxialsquarehelmholtzcoilsystematthealibagmagneticobservatoryupgradedtoamagneticsensorcalibrationfacility
AT vdongre triaxialsquarehelmholtzcoilsystematthealibagmagneticobservatoryupgradedtoamagneticsensorcalibrationfacility
AT slabde triaxialsquarehelmholtzcoilsystematthealibagmagneticobservatoryupgradedtoamagneticsensorcalibrationfacility
_version_ 1725965616645406720