Flexible Hybrid Electrodes for Continuous Measurement of the Local Temperature in Long-Term Wounds

Long-term wounds need a continuous assessment of different biophysical parameters for their treatment, and there is a lack of affordable biocompatible devices capable of obtaining that uninterrupted flow of data. A portable prototype that allows caregivers to know the local temperature behavior of a...

Full description

Bibliographic Details
Main Authors: Ana María Rodes-Carbonell, Juan Torregrosa-Valls, Antonio Guill Ibáñez, Alvaro Tormos Ferrando, María Aránzazu Juan Blanco, Antonio Cebrián Ferriols
Format: Article
Language:English
Published: MDPI AG 2021-04-01
Series:Sensors
Subjects:
Online Access:https://www.mdpi.com/1424-8220/21/8/2741
Description
Summary:Long-term wounds need a continuous assessment of different biophysical parameters for their treatment, and there is a lack of affordable biocompatible devices capable of obtaining that uninterrupted flow of data. A portable prototype that allows caregivers to know the local temperature behavior of a long-term wound over time and compare it with different reference zones has been developed. Alternative flexible substrates, screen-printing techniques, polymeric inks, and an embedded system have been tested to achieve potential indicators of the status and evolution of chronic wounds. The final system is formed by temperature sensors attached to a flexible and stretchable medical-grade substrate, where silver conductive tracks have been printed as interconnections with the data-acquisition unit. In addition, a specific datalogger has been developed for this system. The whole set will enable health personnel to acquire the temperature of the wound and its surroundings in order to make decisions regarding the state and evolution of the wound.
ISSN:1424-8220