A Fusion-Spliced Near-Field Optical Fiber Probe Using Photonic Crystal Fiber for Nanoscale Thermometry Based on Fluorescence-Lifetime Measurement of Quantum Dots

We have developed a novel nanoscale temperature-measurement method using fluorescence in the near-field called Fluorescence Near-field Optics Thermal Nanoscopy (Fluor-NOTN). Fluor-NOTN enables the temperature distributions of nanoscale materials to be measured in vivo/in situ. The proposed method me...

Full description

Bibliographic Details
Main Authors: Toshiharu Saiki, Yuji Nagasaka, Yoshihiro Taguchi, Takuro Fujii
Format: Article
Language:English
Published: MDPI AG 2011-08-01
Series:Sensors
Subjects:
Online Access:http://www.mdpi.com/1424-8220/11/9/8358/
Description
Summary:We have developed a novel nanoscale temperature-measurement method using fluorescence in the near-field called Fluorescence Near-field Optics Thermal Nanoscopy (Fluor-NOTN). Fluor-NOTN enables the temperature distributions of nanoscale materials to be measured in vivo/in situ. The proposed method measures temperature by detecting the temperature dependent fluorescence lifetimes of Cd/Se Quantum Dots (QDs). For a high-sensitivity temperature measurement, the auto-fluorescence generated from a fiber probe should be reduced. In order to decrease the noise, we have fabricated a novel near-field optical-fiber probe by fusion-splicing a photonic crystal fiber (PCF) and a conventional single-mode fiber (SMF). The validity of the novel fiber probe was assessed experimentally by evaluating the auto-fluorescence spectra of the PCF. Due to the decrease of auto-fluorescence, a six- to ten-fold increase of S/N in the near-field fluorescence lifetime detection was achieved with the newly fabricated fusion-spliced near-field optical fiber probe. Additionally, the near-field fluorescence lifetime of the quantum dots was successfully measured by the fabricated fusion-spliced near-field optical fiber probe at room temperature, and was estimated to be 10.0 ns.
ISSN:1424-8220