Potentially serious adverse effects from application of a circumferential compression device for pelvic fracture: A report of three cases

Pelvic circumferential compression devices (PCCDs) have gained wide acceptance in the management of patients with pelvic fracture. These devices are considered safe due to their noninvasive nature and significant hazards associated with the use of PCCDs have not been reported previously. However, we...