Potential errors in conventional DOT measurement techniques in shake flasks and verification using a rotating flexitube optical sensor

<p>Abstract</p> <p>Background</p> <p>Dissolved oxygen tension (DOT) is an important parameter for evaluating a bioprocess. Conventional means to measure DOT in shake flasks using fixed Clark-type electrodes immersed in the bulk liquid are problematic, because they inher...

Full description

Bibliographic Details
Main Authors: Käser Andreas, Kensy Frank, Hansen Sven, Büchs Jochen
Format: Article
Language:English
Published: BMC 2011-05-01
Series:BMC Biotechnology
Online Access:http://www.biomedcentral.com/1472-6750/11/49
id doaj-600af980a83a417fa562febabd4a4907
record_format Article
spelling doaj-600af980a83a417fa562febabd4a49072020-11-25T02:50:03ZengBMCBMC Biotechnology1472-67502011-05-011114910.1186/1472-6750-11-49Potential errors in conventional DOT measurement techniques in shake flasks and verification using a rotating flexitube optical sensorKäser AndreasKensy FrankHansen SvenBüchs Jochen<p>Abstract</p> <p>Background</p> <p>Dissolved oxygen tension (DOT) is an important parameter for evaluating a bioprocess. Conventional means to measure DOT in shake flasks using fixed Clark-type electrodes immersed in the bulk liquid are problematic, because they inherently alter the hydrodynamics of the systems. Other approaches to measure DOT that apply fluorescing sensor spots fixed at the inside wall of a shake flask are also suboptimal. At low filling volumes for cultivating microorganisms with a high oxygen demand, the measured DOT signal may be erroneous. Here, the sensor spot is sometimes exposed to gas in the head space of the flask. Merely repositioning the sensor spot elsewhere in the flask does not address this problem, since there is no location in the shake flask that is always covered by the rotating bulk liquid. Thus, the aim of this prospective study is first, to verify the systemic error of Clark-type electrodes for measuring DOT in shake flasks. The second principle aim is to use the newly built "flexitube optical sensor" to verify potential errors in conventional optical DOT measurements based on fixed sensor spots.</p> <p>Results</p> <p>With the Clark-type electrode, the maximum oxygen transfer capacity in shake flasks rose compared to that of an analogous system without an electrode. This proves changed hydrodynamics in the system with the Clark-type electrode. Furthermore, regarding the sensor spot experiments under oxygen-limited conditions where the DOT value ought to approach zero, the acquired signals were clearly above zero. This implies that the sensor spot is influenced by oxygen present in the headspace and not only by oxygen in the bulk liquid.</p> <p>Conclusions</p> <p>The Clark-type electrode is unsuitable for measuring DOT. Moreover, the newly built rotating flexitube optical sensor is useful to verify potential errors of conventional optical DOT measurement techniques applying fixed sensor spots.</p> http://www.biomedcentral.com/1472-6750/11/49
collection DOAJ
language English
format Article
sources DOAJ
author Käser Andreas
Kensy Frank
Hansen Sven
Büchs Jochen
spellingShingle Käser Andreas
Kensy Frank
Hansen Sven
Büchs Jochen
Potential errors in conventional DOT measurement techniques in shake flasks and verification using a rotating flexitube optical sensor
BMC Biotechnology
author_facet Käser Andreas
Kensy Frank
Hansen Sven
Büchs Jochen
author_sort Käser Andreas
title Potential errors in conventional DOT measurement techniques in shake flasks and verification using a rotating flexitube optical sensor
title_short Potential errors in conventional DOT measurement techniques in shake flasks and verification using a rotating flexitube optical sensor
title_full Potential errors in conventional DOT measurement techniques in shake flasks and verification using a rotating flexitube optical sensor
title_fullStr Potential errors in conventional DOT measurement techniques in shake flasks and verification using a rotating flexitube optical sensor
title_full_unstemmed Potential errors in conventional DOT measurement techniques in shake flasks and verification using a rotating flexitube optical sensor
title_sort potential errors in conventional dot measurement techniques in shake flasks and verification using a rotating flexitube optical sensor
publisher BMC
series BMC Biotechnology
issn 1472-6750
publishDate 2011-05-01
description <p>Abstract</p> <p>Background</p> <p>Dissolved oxygen tension (DOT) is an important parameter for evaluating a bioprocess. Conventional means to measure DOT in shake flasks using fixed Clark-type electrodes immersed in the bulk liquid are problematic, because they inherently alter the hydrodynamics of the systems. Other approaches to measure DOT that apply fluorescing sensor spots fixed at the inside wall of a shake flask are also suboptimal. At low filling volumes for cultivating microorganisms with a high oxygen demand, the measured DOT signal may be erroneous. Here, the sensor spot is sometimes exposed to gas in the head space of the flask. Merely repositioning the sensor spot elsewhere in the flask does not address this problem, since there is no location in the shake flask that is always covered by the rotating bulk liquid. Thus, the aim of this prospective study is first, to verify the systemic error of Clark-type electrodes for measuring DOT in shake flasks. The second principle aim is to use the newly built "flexitube optical sensor" to verify potential errors in conventional optical DOT measurements based on fixed sensor spots.</p> <p>Results</p> <p>With the Clark-type electrode, the maximum oxygen transfer capacity in shake flasks rose compared to that of an analogous system without an electrode. This proves changed hydrodynamics in the system with the Clark-type electrode. Furthermore, regarding the sensor spot experiments under oxygen-limited conditions where the DOT value ought to approach zero, the acquired signals were clearly above zero. This implies that the sensor spot is influenced by oxygen present in the headspace and not only by oxygen in the bulk liquid.</p> <p>Conclusions</p> <p>The Clark-type electrode is unsuitable for measuring DOT. Moreover, the newly built rotating flexitube optical sensor is useful to verify potential errors of conventional optical DOT measurement techniques applying fixed sensor spots.</p>
url http://www.biomedcentral.com/1472-6750/11/49
work_keys_str_mv AT kaserandreas potentialerrorsinconventionaldotmeasurementtechniquesinshakeflasksandverificationusingarotatingflexitubeopticalsensor
AT kensyfrank potentialerrorsinconventionaldotmeasurementtechniquesinshakeflasksandverificationusingarotatingflexitubeopticalsensor
AT hansensven potentialerrorsinconventionaldotmeasurementtechniquesinshakeflasksandverificationusingarotatingflexitubeopticalsensor
AT buchsjochen potentialerrorsinconventionaldotmeasurementtechniquesinshakeflasksandverificationusingarotatingflexitubeopticalsensor
_version_ 1724740425634283520