Temperature field investigation of the industrial autoclave ASCAMAT-230

Heterogeneity of the temperature field in the sterilization chambers of industrial autoclaves during heating and cooling phase is one of the problems in the development of new regime for product heat treatment on the preliminary selection and laboratory testing stages. This is the reason of studying...

Full description

Bibliographic Details
Main Authors: Stolyanov A. V., Kaychenov A. V., Kuranova L. K., Maslov A. A., Grokhovsky V. A.
Format: Article
Language:Russian
Published: Murmansk State Technical University 2017-09-01
Series:Vestnik MGTU
Subjects:
Online Access:http://vestnik.mstu.edu.ru/show.shtml?art=1901&pdf=1
id doaj-5e3ff5ecc3304c4a8d4f1a180260689d
record_format Article
spelling doaj-5e3ff5ecc3304c4a8d4f1a180260689d2020-11-24T22:34:50ZrusMurmansk State Technical UniversityVestnik MGTU1560-92781997-47362017-09-0120356357110.21443/1560-9278-2017-20-3-563-571Temperature field investigation of the industrial autoclave ASCAMAT-230Stolyanov A. V.0Kaychenov A. V. 1Kuranova L. K. 2Maslov A. A. 3Grokhovsky V. A. 4Murmansk State Technical UniversityMurmansk State Technical UniversityMurmansk State Technical UniversityMurmansk State Technical UniversityMurmansk State Technical UniversityHeterogeneity of the temperature field in the sterilization chambers of industrial autoclaves during heating and cooling phase is one of the problems in the development of new regime for product heat treatment on the preliminary selection and laboratory testing stages. This is the reason of studying the temperature field of the industrial autoclave ASCAMAT-230. The autoclave ASCAMAT-230 is a vertical autoclave of 230 liters capacity heated by three tubular electric heaters. Determination of temperature field parameters has been carried out according to the method for studying the temperature field of periodic devices' heating medium for sterilizing canned food. Six Thermochron iButton temperature loggers have been used to measure the temperatures in the sterilization chamber of the ASCAMAT-230 autoclave. Logger 1 has been put in the place of the standard thermometer installation, loggers 2 and 3, 5 and 6 – in the center and on periphery of the lower and upper parts of the autoclave respectively, logger 4 – in the center of the middle part of the autoclave's sterilization chamber. Temperature measurement in cans with the product has been carried out by temperature loggers from the Ellab TrackSense PRO complex. The loggers have been installed in two cans with homogeneous product located in the upper and lower parts of the autoclave. The time temperature dependences for the entire sterilization process, as well as for the heating and cooling stages have been constructed according to the acquired information. Based on these data conclusions about the uniformity of the temperature field inside the sterilization chamber of the autoclave ASCAMAT-230 during the heating, sterilization and cooling stages have been developed. Some recommendations for using the autoclave ASCAMAT-230 in the preliminary selection and verification of sterilization regimes for canned foods from hydrobionts, and also for the creation of an industrial autoclave control system have been given in the paper.http://vestnik.mstu.edu.ru/show.shtml?art=1901&pdf=1loggers 2 and 35 and 6 – in the center and on periphery of the lower and upper parts of the autoclave respectivelylogger 4 – in the center of the middle part of the autoclave's sterilization chamber. Temperature measurement in cans with the product has been carried out by temperature loggers from the Ellab TrackSense PRO complex. The loggers have been installed in two cans with homogeneous product located in the upper and lower parts of the autoclave. The time temperature dependences for the entire sterilization processas well as for the heating and cooling stages have been constructed according to the acquired information. Based on these data conclusions about the uniformity of the temperature field inside the sterilization chamber of the autoclave ASCAMAT-230 during the heatingsterilization and cooling stages have been developed. Some recommendations for using the autoclave ASCAMAT-230 in the preliminary selection and verification of sterilization regimes for canned foods from hydrobiontstemperature fieldASCAMAT-230temperature logger
collection DOAJ
language Russian
format Article
sources DOAJ
author Stolyanov A. V.
Kaychenov A. V.
Kuranova L. K.
Maslov A. A.
Grokhovsky V. A.
spellingShingle Stolyanov A. V.
Kaychenov A. V.
Kuranova L. K.
Maslov A. A.
Grokhovsky V. A.
Temperature field investigation of the industrial autoclave ASCAMAT-230
Vestnik MGTU
loggers 2 and 3
5 and 6 – in the center and on periphery of the lower and upper parts of the autoclave respectively
logger 4 – in the center of the middle part of the autoclave's sterilization chamber. Temperature measurement in cans with the product has been carried out by temperature loggers from the Ellab TrackSense PRO complex. The loggers have been installed in two cans with homogeneous product located in the upper and lower parts of the autoclave. The time temperature dependences for the entire sterilization process
as well as for the heating and cooling stages have been constructed according to the acquired information. Based on these data conclusions about the uniformity of the temperature field inside the sterilization chamber of the autoclave ASCAMAT-230 during the heating
sterilization and cooling stages have been developed. Some recommendations for using the autoclave ASCAMAT-230 in the preliminary selection and verification of sterilization regimes for canned foods from hydrobionts
temperature field
ASCAMAT-230
temperature logger
author_facet Stolyanov A. V.
Kaychenov A. V.
Kuranova L. K.
Maslov A. A.
Grokhovsky V. A.
author_sort Stolyanov A. V.
title Temperature field investigation of the industrial autoclave ASCAMAT-230
title_short Temperature field investigation of the industrial autoclave ASCAMAT-230
title_full Temperature field investigation of the industrial autoclave ASCAMAT-230
title_fullStr Temperature field investigation of the industrial autoclave ASCAMAT-230
title_full_unstemmed Temperature field investigation of the industrial autoclave ASCAMAT-230
title_sort temperature field investigation of the industrial autoclave ascamat-230
publisher Murmansk State Technical University
series Vestnik MGTU
issn 1560-9278
1997-4736
publishDate 2017-09-01
description Heterogeneity of the temperature field in the sterilization chambers of industrial autoclaves during heating and cooling phase is one of the problems in the development of new regime for product heat treatment on the preliminary selection and laboratory testing stages. This is the reason of studying the temperature field of the industrial autoclave ASCAMAT-230. The autoclave ASCAMAT-230 is a vertical autoclave of 230 liters capacity heated by three tubular electric heaters. Determination of temperature field parameters has been carried out according to the method for studying the temperature field of periodic devices' heating medium for sterilizing canned food. Six Thermochron iButton temperature loggers have been used to measure the temperatures in the sterilization chamber of the ASCAMAT-230 autoclave. Logger 1 has been put in the place of the standard thermometer installation, loggers 2 and 3, 5 and 6 – in the center and on periphery of the lower and upper parts of the autoclave respectively, logger 4 – in the center of the middle part of the autoclave's sterilization chamber. Temperature measurement in cans with the product has been carried out by temperature loggers from the Ellab TrackSense PRO complex. The loggers have been installed in two cans with homogeneous product located in the upper and lower parts of the autoclave. The time temperature dependences for the entire sterilization process, as well as for the heating and cooling stages have been constructed according to the acquired information. Based on these data conclusions about the uniformity of the temperature field inside the sterilization chamber of the autoclave ASCAMAT-230 during the heating, sterilization and cooling stages have been developed. Some recommendations for using the autoclave ASCAMAT-230 in the preliminary selection and verification of sterilization regimes for canned foods from hydrobionts, and also for the creation of an industrial autoclave control system have been given in the paper.
topic loggers 2 and 3
5 and 6 – in the center and on periphery of the lower and upper parts of the autoclave respectively
logger 4 – in the center of the middle part of the autoclave's sterilization chamber. Temperature measurement in cans with the product has been carried out by temperature loggers from the Ellab TrackSense PRO complex. The loggers have been installed in two cans with homogeneous product located in the upper and lower parts of the autoclave. The time temperature dependences for the entire sterilization process
as well as for the heating and cooling stages have been constructed according to the acquired information. Based on these data conclusions about the uniformity of the temperature field inside the sterilization chamber of the autoclave ASCAMAT-230 during the heating
sterilization and cooling stages have been developed. Some recommendations for using the autoclave ASCAMAT-230 in the preliminary selection and verification of sterilization regimes for canned foods from hydrobionts
temperature field
ASCAMAT-230
temperature logger
url http://vestnik.mstu.edu.ru/show.shtml?art=1901&pdf=1
work_keys_str_mv AT stolyanovav temperaturefieldinvestigationoftheindustrialautoclaveascamat230
AT kaychenovav temperaturefieldinvestigationoftheindustrialautoclaveascamat230
AT kuranovalk temperaturefieldinvestigationoftheindustrialautoclaveascamat230
AT maslovaa temperaturefieldinvestigationoftheindustrialautoclaveascamat230
AT grokhovskyva temperaturefieldinvestigationoftheindustrialautoclaveascamat230
_version_ 1725725778256068608