Calculating AMBU Bag Dimensions for Use in Portable Ventilators

Based on the design of non-portable ventilators, this project examines and analyzes the primary individual component used within a motorized AMBU bag ventilator - the AMBU bag. Calculations are conducted to assess the minimum volume for an AMBU bag to provide sufficient oxygen for average total lung...

Full description

Bibliographic Details
Main Author: Camden Smith
Format: Article
Language:English
Published: University of South Florida 2020-10-01
Series:Undergraduate Journal of Mathematical Modeling: One + Two
Subjects:
Online Access:https://scholarcommons.usf.edu/ujmm/vol11/iss1/4/
Description
Summary:Based on the design of non-portable ventilators, this project examines and analyzes the primary individual component used within a motorized AMBU bag ventilator - the AMBU bag. Calculations are conducted to assess the minimum volume for an AMBU bag to provide sufficient oxygen for average total lung capacity. It allows to determine the feasibility and effectiveness of an electrical, portable AMBU bag ventilator under the reduced size requirements. A calculus-based formula, known as the Disk Method, is utilized for these calculations. It is shown that the Disk method can be reversed to find a shape of an AMBU bag given a predetermined volume and desired dimensions. Thus an affordable, motorized, portable AMBU bag ventilator with full functionality and effectiveness can be developed. It will contribute to fulfilling the need for more ventilators due the current COVID-19 pandemic.
ISSN:2326-3652