Long-Term Measurement of Impedance in Chronically Implanted Depth and Subdural Electrodes During Responsive Neurostimulation in Humans

Long-term stability of the electrode–tissue interface may be required to maintain optimal neural recording with subdural and deep brain implants and to permit appropriate delivery of neuromodulation therapy. Although short-term changes in impedance at the electrode–tissue interface are known to occu...

Full description

Bibliographic Details
Main Authors: Karl A. Sillay, Paul Rutecki, Kathy Cicora, Greg Worrell, Joseph Drazkowski, Jerry J. Shih, Ashwini D. Sharan, Martha J. Morrell, Justin Williams, Brett Wingeier
Format: Article
Language:English
Published: Elsevier 2013-09-01
Series:Brain Stimulation
Subjects:
Online Access:http://www.sciencedirect.com/science/article/pii/S1935861X13000752
Description
Summary:Long-term stability of the electrode–tissue interface may be required to maintain optimal neural recording with subdural and deep brain implants and to permit appropriate delivery of neuromodulation therapy. Although short-term changes in impedance at the electrode–tissue interface are known to occur, long-term changes in impedance have not previously been examined in detail in humans. To provide further information about short- and long-term impedance changes in chronically implanted electrodes, a dataset from 191 persons with medically intractable epilepsy participating in a trial of an investigational responsive neurostimulation device (the RNS® System, NeuroPace, Inc.) was reviewed. Monopolar impedance measurements were available for 391 depth and subdural leads containing a total of 1564 electrodes; measurements were available for median 802 days post-implant (range 28–1634). Although there were statistically significant short-term impedance changes, long-term impedance was stable after one year. Impedances for depth electrodes transiently increased during the third week after lead implantation and impedances for subdural electrodes increased over 12 weeks post-implant, then were stable over the subsequent long-term follow-up. Both depth and subdural electrode impedances demonstrated long-term stability, suggesting that the quality of long-term electrographic recordings (the data used to control responsive brain stimulation) can be maintained over time.
ISSN:1935-861X