Summary: | Abstract Objective Commercial kits of column tests for pre-transfusion testing have progressively replaced conventional tube tests in most laboratories. Aim of this study was to compare three commercial test cell panels for the identification of irregular red blood cell (RBC) alloantibodies. Overall, 44 samples with a positive indirect antiglobulin test (IAT) by routine testing were used for comparison of following panels: Ortho RESOLVE® panelC (Ortho Clinical Diagnostics (OCD), Milan, Italy), ID-DiaPanel(-P) (Bio-Rad Laboratories, CA, USA) and Identisera Diana(P) (Grifols, Barcelona, Spain). Column agglutination techniques were used, with microtubes containing either microgel (Bio-Rad/Grifols) or glass bead microparticles (Ortho). Results Alloantibody identification was possible in 38 samples, of which identical identification was shown in 33 samples by all methods. The remaining samples showed differences between certain methods, with the gel card system being superior to the glass card system for analyzing stored samples Considering that not all samples were evaluated in all three methods, the concordance rate reached 100% between Bio-Rad and Grifols, 90.5% between Bio-Rad and OCD, 86.5% between OCD and Grifols and 90.5% between all methods. Although differences in sensitivities were seen for specific antibodies, the three methods showed comparable performance for the identification of RBC alloantibodies.
|