Serum bile acid analysis: a rapid, direct enzymatic method using dual-beam spectrophotofluorimetry

The direct quantitative measurement of total bile acids in serum has been achieved using an enzymatic fluorescent method with a dual-beam spectrophotofluorimeter. By use of a 3α-hydroxysteroid dehydrogenase, oxidation of bile acids with NAD is completed in 200 seconds with the observed NADH fluoresc...

Full description

Bibliographic Details
Main Authors: P A Siskos, P T Cahill, N B Javitt, M.D., Ph.D.
Format: Article
Language:English
Published: Elsevier 1977-09-01
Series:Journal of Lipid Research
Subjects:
Online Access:http://www.sciencedirect.com/science/article/pii/S0022227520416104
Description
Summary:The direct quantitative measurement of total bile acids in serum has been achieved using an enzymatic fluorescent method with a dual-beam spectrophotofluorimeter. By use of a 3α-hydroxysteroid dehydrogenase, oxidation of bile acids with NAD is completed in 200 seconds with the observed NADH fluorescence being proportional to the concentration of serum bile acids. This method is rapid (8 minutes per individual sample), has an intrinsic sensitivity of ± µM of total bile acids, requires no sample preparation and less than 0.8 ml of serum. Paired data analysis using enzymatic fluorescence and gas–liquid chromatographic methods gives a correlation coefficient (r) of 0.99 for 34 samples ranging from 2 to 530 µM.