Evaluation of the Portal Imaging System Performance for an Elekta Precise Linac in Radiotherapy

Introduction: Electronic portal imaging devices (EPIDs) provide two- and three-dimensional planar and volumetric cone beam images to improve the accuracy of radiation treatment delivery. Periodic quality assurance (QA) of EPIDs is essential for dosimetric verification in radiotherapy. In this study,...

Full description

Bibliographic Details
Main Authors: Fatemeh Shahedi, Mehdi Momennezhad, shahrokh Nasseri, Hamid Gholamhosseinian
Format: Article
Language:English
Published: Mashhad University of Medical Sciences 2018-10-01
Series:Iranian Journal of Medical Physics
Subjects:
Online Access:http://ijmp.mums.ac.ir/article_10360_2d9a5cce0c6e89bad98663130cc5d16e.pdf
id doaj-540636daa6374d9882345e8d93e1f0fa
record_format Article
spelling doaj-540636daa6374d9882345e8d93e1f0fa2020-11-24T22:47:58ZengMashhad University of Medical SciencesIranian Journal of Medical Physics2345-36722345-36722018-10-0115429530310.22038/ijmp.2018.27401.128710360Evaluation of the Portal Imaging System Performance for an Elekta Precise Linac in RadiotherapyFatemeh Shahedi0Mehdi Momennezhad1shahrokh Nasseri2Hamid Gholamhosseinian3Medical Physics Research Center, Mashhad University of Medical Sciences, Mashhad, Iran.Medical Physics Research Center, Mashhad University of Medical Sciences, Mashhad, Iran.Medical Physics Research Center, Mashhad University of Medical Sciences, Mashhad, Iran.Medical Physics Research center, Mashhad University of Medical Sciences, Mashhad, IranIntroduction: Electronic portal imaging devices (EPIDs) provide two- and three-dimensional planar and volumetric cone beam images to improve the accuracy of radiation treatment delivery. Periodic quality assurance (QA) of EPIDs is essential for dosimetric verification in radiotherapy. In this study, a QA program was implemented to evaluate the function of the EPID to be confident in applying corrections for the uncertainty of patient set-up. Material and Methods: Firstly, the safety features were verified, and the uniformity of EPID response was evaluated using flat panel detector. Additionally, the contrast and spatial resolutions of the EPID were assessed using detail counting of the Los Vegas phantom images by visualization method and measuring the modulation transfer function using edge technique, respectively. Moreover, a combination of smoothing methods was used for optimal use of edge detection algorithm for the noisy portal images. Finally, the location of the central ray on the EPID surface at different gantry angles was determined to evaluate the mechanical stability of the supporting arm. Results: The safety interlocks were found to be functional.The EPID response variation was less than 3% according to the results obtained from the detector. The contrast resolution met the recommended tolerance; however, the visualization method was widely observer-dependent. The value of f50 for spatial resolution was 0.401±0.005 lp/mm for the photon energy of 6 MV. The supporting arm deviation was within ±1 mm. Conclusion: The periodic QA of image guidance system gave confidence to apply the corrections for set-up in clinic.http://ijmp.mums.ac.ir/article_10360_2d9a5cce0c6e89bad98663130cc5d16e.pdfQuality AssuranceRadiotherapyImaging
collection DOAJ
language English
format Article
sources DOAJ
author Fatemeh Shahedi
Mehdi Momennezhad
shahrokh Nasseri
Hamid Gholamhosseinian
spellingShingle Fatemeh Shahedi
Mehdi Momennezhad
shahrokh Nasseri
Hamid Gholamhosseinian
Evaluation of the Portal Imaging System Performance for an Elekta Precise Linac in Radiotherapy
Iranian Journal of Medical Physics
Quality Assurance
Radiotherapy
Imaging
author_facet Fatemeh Shahedi
Mehdi Momennezhad
shahrokh Nasseri
Hamid Gholamhosseinian
author_sort Fatemeh Shahedi
title Evaluation of the Portal Imaging System Performance for an Elekta Precise Linac in Radiotherapy
title_short Evaluation of the Portal Imaging System Performance for an Elekta Precise Linac in Radiotherapy
title_full Evaluation of the Portal Imaging System Performance for an Elekta Precise Linac in Radiotherapy
title_fullStr Evaluation of the Portal Imaging System Performance for an Elekta Precise Linac in Radiotherapy
title_full_unstemmed Evaluation of the Portal Imaging System Performance for an Elekta Precise Linac in Radiotherapy
title_sort evaluation of the portal imaging system performance for an elekta precise linac in radiotherapy
publisher Mashhad University of Medical Sciences
series Iranian Journal of Medical Physics
issn 2345-3672
2345-3672
publishDate 2018-10-01
description Introduction: Electronic portal imaging devices (EPIDs) provide two- and three-dimensional planar and volumetric cone beam images to improve the accuracy of radiation treatment delivery. Periodic quality assurance (QA) of EPIDs is essential for dosimetric verification in radiotherapy. In this study, a QA program was implemented to evaluate the function of the EPID to be confident in applying corrections for the uncertainty of patient set-up. Material and Methods: Firstly, the safety features were verified, and the uniformity of EPID response was evaluated using flat panel detector. Additionally, the contrast and spatial resolutions of the EPID were assessed using detail counting of the Los Vegas phantom images by visualization method and measuring the modulation transfer function using edge technique, respectively. Moreover, a combination of smoothing methods was used for optimal use of edge detection algorithm for the noisy portal images. Finally, the location of the central ray on the EPID surface at different gantry angles was determined to evaluate the mechanical stability of the supporting arm. Results: The safety interlocks were found to be functional.The EPID response variation was less than 3% according to the results obtained from the detector. The contrast resolution met the recommended tolerance; however, the visualization method was widely observer-dependent. The value of f50 for spatial resolution was 0.401±0.005 lp/mm for the photon energy of 6 MV. The supporting arm deviation was within ±1 mm. Conclusion: The periodic QA of image guidance system gave confidence to apply the corrections for set-up in clinic.
topic Quality Assurance
Radiotherapy
Imaging
url http://ijmp.mums.ac.ir/article_10360_2d9a5cce0c6e89bad98663130cc5d16e.pdf
work_keys_str_mv AT fatemehshahedi evaluationoftheportalimagingsystemperformanceforanelektapreciselinacinradiotherapy
AT mehdimomennezhad evaluationoftheportalimagingsystemperformanceforanelektapreciselinacinradiotherapy
AT shahrokhnasseri evaluationoftheportalimagingsystemperformanceforanelektapreciselinacinradiotherapy
AT hamidgholamhosseinian evaluationoftheportalimagingsystemperformanceforanelektapreciselinacinradiotherapy
_version_ 1725680326595837952