The Accuracy and Precision of the Continuously Stored Data from Flash Glucose Monitoring System in Type 2 Diabetes Patients during Standard Meal Tolerance Test

Background. The purpose of this study was to investigate the accuracy of the continuously stored data from the Abbott FreeStyle Libre flash glucose monitoring (FGM) system in Chinese diabetes patients during standard meal tests when glucose concentrations were rapidly changing. Subjects and Methods....

Full description

Bibliographic Details
Main Authors: Rengna Yan, Huiqin Li, Xiaocen Kong, Xiaofang Zhai, Maoyuan Chen, Yixuan Sun, Lei Ye, Xiaofei Su, Jianhua Ma
Format: Article
Language:English
Published: Hindawi Limited 2020-01-01
Series:International Journal of Endocrinology
Online Access:http://dx.doi.org/10.1155/2020/5947680
id doaj-52678792066444c68f29ab2182562ba3
record_format Article
spelling doaj-52678792066444c68f29ab2182562ba32020-11-25T02:26:54ZengHindawi LimitedInternational Journal of Endocrinology1687-83371687-83452020-01-01202010.1155/2020/59476805947680The Accuracy and Precision of the Continuously Stored Data from Flash Glucose Monitoring System in Type 2 Diabetes Patients during Standard Meal Tolerance TestRengna Yan0Huiqin Li1Xiaocen Kong2Xiaofang Zhai3Maoyuan Chen4Yixuan Sun5Lei Ye6Xiaofei Su7Jianhua Ma8Department of Endocrinology, Nanjing First Hospital, Nanjing Medical University, Nanjing 210012, ChinaDepartment of Endocrinology, Nanjing First Hospital, Nanjing Medical University, Nanjing 210012, ChinaDepartment of Endocrinology, Nanjing First Hospital, Nanjing Medical University, Nanjing 210012, ChinaDepartment of Endocrinology, Nanjing First Hospital, Nanjing Medical University, Nanjing 210012, ChinaDepartment of Endocrinology, Nanjing First Hospital, Nanjing Medical University, Nanjing 210012, ChinaDepartment of Endocrinology, Nanjing First Hospital, Nanjing Medical University, Nanjing 210012, ChinaNational Heart Research Institute Singapore, National Heart Centre Singapore, SingaporeDepartment of Endocrinology, Nanjing First Hospital, Nanjing Medical University, Nanjing 210012, ChinaDepartment of Endocrinology, Nanjing First Hospital, Nanjing Medical University, Nanjing 210012, ChinaBackground. The purpose of this study was to investigate the accuracy of the continuously stored data from the Abbott FreeStyle Libre flash glucose monitoring (FGM) system in Chinese diabetes patients during standard meal tests when glucose concentrations were rapidly changing. Subjects and Methods. Interstitial glucose levels were monitored for 14 days in 26 insulin-treated patients with type 2 diabetes using the FGM system. Standard meal tests were conducted to induce large glucose swings. Venous blood glucose (VBG) was tested at 0, 30, 60, and 120 min after standard meal tests in one middle day of the first and second weeks, respectively. The corresponding sensor glucose values were obtained from interpolating continuously stored data points. Assessment of accuracy was according to recent consensus recommendations with median absolute relative difference (MARD) and Clarke and Parkes error grid analysis (CEG and PEG). Results. Among 208 paired sensor-reference values, 100% were falling within zones A and B of the Clarke and Parkes error grid analysis. The overall MARD was 10.7% (SD, 7.8%). Weighted least squares regression analysis resulted in high agreement between the FGM sensor glucose and VBG readings. The overall MTT results showed that FGM was lower than actual VBG, with MAD of 22.1 mg/dL (1.2 mmol/L). At VBG rates of change of -1 to 0, 0 to 1, 1 to 2, and 2 to 3 mg/dl/min, MARD results were 11.4% (SD, 8.7%), 9.4% (SD, 6.5%), 9.9% (SD, 7.5%), and 9.5% (SD, 7.7%). At rapidly changing VBG concentrations (>3 mg/dl/min), MARD increased to 19.0%, which was significantly higher than slow changing BG groups. Conclusions. Continuously stored interstitial glucose measurements with the FGM system were found to be acceptable to evaluate VBG in terms of clinical decision during standard meal tests. The continuously stored data from the FGM system appeared to underestimate venous glucose and performed less well during rapid glucose changes.http://dx.doi.org/10.1155/2020/5947680
collection DOAJ
language English
format Article
sources DOAJ
author Rengna Yan
Huiqin Li
Xiaocen Kong
Xiaofang Zhai
Maoyuan Chen
Yixuan Sun
Lei Ye
Xiaofei Su
Jianhua Ma
spellingShingle Rengna Yan
Huiqin Li
Xiaocen Kong
Xiaofang Zhai
Maoyuan Chen
Yixuan Sun
Lei Ye
Xiaofei Su
Jianhua Ma
The Accuracy and Precision of the Continuously Stored Data from Flash Glucose Monitoring System in Type 2 Diabetes Patients during Standard Meal Tolerance Test
International Journal of Endocrinology
author_facet Rengna Yan
Huiqin Li
Xiaocen Kong
Xiaofang Zhai
Maoyuan Chen
Yixuan Sun
Lei Ye
Xiaofei Su
Jianhua Ma
author_sort Rengna Yan
title The Accuracy and Precision of the Continuously Stored Data from Flash Glucose Monitoring System in Type 2 Diabetes Patients during Standard Meal Tolerance Test
title_short The Accuracy and Precision of the Continuously Stored Data from Flash Glucose Monitoring System in Type 2 Diabetes Patients during Standard Meal Tolerance Test
title_full The Accuracy and Precision of the Continuously Stored Data from Flash Glucose Monitoring System in Type 2 Diabetes Patients during Standard Meal Tolerance Test
title_fullStr The Accuracy and Precision of the Continuously Stored Data from Flash Glucose Monitoring System in Type 2 Diabetes Patients during Standard Meal Tolerance Test
title_full_unstemmed The Accuracy and Precision of the Continuously Stored Data from Flash Glucose Monitoring System in Type 2 Diabetes Patients during Standard Meal Tolerance Test
title_sort accuracy and precision of the continuously stored data from flash glucose monitoring system in type 2 diabetes patients during standard meal tolerance test
publisher Hindawi Limited
series International Journal of Endocrinology
issn 1687-8337
1687-8345
publishDate 2020-01-01
description Background. The purpose of this study was to investigate the accuracy of the continuously stored data from the Abbott FreeStyle Libre flash glucose monitoring (FGM) system in Chinese diabetes patients during standard meal tests when glucose concentrations were rapidly changing. Subjects and Methods. Interstitial glucose levels were monitored for 14 days in 26 insulin-treated patients with type 2 diabetes using the FGM system. Standard meal tests were conducted to induce large glucose swings. Venous blood glucose (VBG) was tested at 0, 30, 60, and 120 min after standard meal tests in one middle day of the first and second weeks, respectively. The corresponding sensor glucose values were obtained from interpolating continuously stored data points. Assessment of accuracy was according to recent consensus recommendations with median absolute relative difference (MARD) and Clarke and Parkes error grid analysis (CEG and PEG). Results. Among 208 paired sensor-reference values, 100% were falling within zones A and B of the Clarke and Parkes error grid analysis. The overall MARD was 10.7% (SD, 7.8%). Weighted least squares regression analysis resulted in high agreement between the FGM sensor glucose and VBG readings. The overall MTT results showed that FGM was lower than actual VBG, with MAD of 22.1 mg/dL (1.2 mmol/L). At VBG rates of change of -1 to 0, 0 to 1, 1 to 2, and 2 to 3 mg/dl/min, MARD results were 11.4% (SD, 8.7%), 9.4% (SD, 6.5%), 9.9% (SD, 7.5%), and 9.5% (SD, 7.7%). At rapidly changing VBG concentrations (>3 mg/dl/min), MARD increased to 19.0%, which was significantly higher than slow changing BG groups. Conclusions. Continuously stored interstitial glucose measurements with the FGM system were found to be acceptable to evaluate VBG in terms of clinical decision during standard meal tests. The continuously stored data from the FGM system appeared to underestimate venous glucose and performed less well during rapid glucose changes.
url http://dx.doi.org/10.1155/2020/5947680
work_keys_str_mv AT rengnayan theaccuracyandprecisionofthecontinuouslystoreddatafromflashglucosemonitoringsystemintype2diabetespatientsduringstandardmealtolerancetest
AT huiqinli theaccuracyandprecisionofthecontinuouslystoreddatafromflashglucosemonitoringsystemintype2diabetespatientsduringstandardmealtolerancetest
AT xiaocenkong theaccuracyandprecisionofthecontinuouslystoreddatafromflashglucosemonitoringsystemintype2diabetespatientsduringstandardmealtolerancetest
AT xiaofangzhai theaccuracyandprecisionofthecontinuouslystoreddatafromflashglucosemonitoringsystemintype2diabetespatientsduringstandardmealtolerancetest
AT maoyuanchen theaccuracyandprecisionofthecontinuouslystoreddatafromflashglucosemonitoringsystemintype2diabetespatientsduringstandardmealtolerancetest
AT yixuansun theaccuracyandprecisionofthecontinuouslystoreddatafromflashglucosemonitoringsystemintype2diabetespatientsduringstandardmealtolerancetest
AT leiye theaccuracyandprecisionofthecontinuouslystoreddatafromflashglucosemonitoringsystemintype2diabetespatientsduringstandardmealtolerancetest
AT xiaofeisu theaccuracyandprecisionofthecontinuouslystoreddatafromflashglucosemonitoringsystemintype2diabetespatientsduringstandardmealtolerancetest
AT jianhuama theaccuracyandprecisionofthecontinuouslystoreddatafromflashglucosemonitoringsystemintype2diabetespatientsduringstandardmealtolerancetest
AT rengnayan accuracyandprecisionofthecontinuouslystoreddatafromflashglucosemonitoringsystemintype2diabetespatientsduringstandardmealtolerancetest
AT huiqinli accuracyandprecisionofthecontinuouslystoreddatafromflashglucosemonitoringsystemintype2diabetespatientsduringstandardmealtolerancetest
AT xiaocenkong accuracyandprecisionofthecontinuouslystoreddatafromflashglucosemonitoringsystemintype2diabetespatientsduringstandardmealtolerancetest
AT xiaofangzhai accuracyandprecisionofthecontinuouslystoreddatafromflashglucosemonitoringsystemintype2diabetespatientsduringstandardmealtolerancetest
AT maoyuanchen accuracyandprecisionofthecontinuouslystoreddatafromflashglucosemonitoringsystemintype2diabetespatientsduringstandardmealtolerancetest
AT yixuansun accuracyandprecisionofthecontinuouslystoreddatafromflashglucosemonitoringsystemintype2diabetespatientsduringstandardmealtolerancetest
AT leiye accuracyandprecisionofthecontinuouslystoreddatafromflashglucosemonitoringsystemintype2diabetespatientsduringstandardmealtolerancetest
AT xiaofeisu accuracyandprecisionofthecontinuouslystoreddatafromflashglucosemonitoringsystemintype2diabetespatientsduringstandardmealtolerancetest
AT jianhuama accuracyandprecisionofthecontinuouslystoreddatafromflashglucosemonitoringsystemintype2diabetespatientsduringstandardmealtolerancetest
_version_ 1715484149083537408