Determination of Optimum Planar Imaging Parameters for Small Structures with Diameters Less Than the Resolution of the Gamma Camera

Introduction: The limited spatial resolution of the gamma camera hinders the absolute quantification of planar images of small structures. The imaged structures are affected by partial volume effects (PVEs), which can spread activity and lead to underestimation of the regional distribution.  The use...

Full description

Bibliographic Details
Main Author: Mpumelelo Nyathi
Format: Article
Language:English
Published: Mashhad University of Medical Sciences 2017-12-01
Series:Iranian Journal of Medical Physics
Subjects:
Online Access:http://ijmp.mums.ac.ir/article_9267_2020713899b0a0e8b080c581dfdbd514.pdf
id doaj-51d66815f1ca4d2ea3b948b234f7d334
record_format Article
spelling doaj-51d66815f1ca4d2ea3b948b234f7d3342020-11-24T22:13:28ZengMashhad University of Medical SciencesIranian Journal of Medical Physics2345-36722345-36722017-12-0114421922810.22038/ijmp.2017.24559.12469267Determination of Optimum Planar Imaging Parameters for Small Structures with Diameters Less Than the Resolution of the Gamma CameraMpumelelo Nyathi0Department of Medical Physics Sefako Makgatho Health Sciences University SouthAfricaIntroduction: The limited spatial resolution of the gamma camera hinders the absolute quantification of planar images of small structures. The imaged structures are affected by partial volume effects (PVEs), which can spread activity and lead to underestimation of the regional distribution.  The use of optimum planar parameters reduces the impact of the limited spatial resolution of the gamma camera and the statistical noise inherent to low photon count, thus improving quantification. In this study, we aimed to determine the optimum planar imaging parameters for small structures. Materials and Methods: A thyroid protocol was used to acquire planar images of the spheres A, B, and C (16 mm, 12 mm, and 11 mm in diameter, respectively) whilst filled with a targeted activity concentration of technetium-99m. One sphere was mounted at the centre of the Jaszczak Phantom and the other two adjacent to its walls using capillary stems fitted on the spheres. The phantom was filled with distilled water. The targeted activity concentrations used were 74 kBq/mL, 100 kBq/mL, 150 kBq/mL, and 300 kBq/mL. Images of the same count per pixel were acquired on 64  64, 128  128, 256  256, 512  512, and 1024  1024 pixels using a vertical detector mounted 5 cm above the phantom. All the images were quantified using ImageJ software, version 1.48a, Java 1.70_51 [64-bit]. Results: The optimum planar imaging parameters established were a matrix size of 128 128 pixels and technetium-99m solution of activity concentration of 300 kBq/ml. Conclusion: The use of optimal imaging parameters reduces the impact of PVEs, leading to improved quantitative accuracy.http://ijmp.mums.ac.ir/article_9267_2020713899b0a0e8b080c581dfdbd514.pdfMedical ImagingRadioisotope ImagingPartial Volume Effects
collection DOAJ
language English
format Article
sources DOAJ
author Mpumelelo Nyathi
spellingShingle Mpumelelo Nyathi
Determination of Optimum Planar Imaging Parameters for Small Structures with Diameters Less Than the Resolution of the Gamma Camera
Iranian Journal of Medical Physics
Medical Imaging
Radioisotope Imaging
Partial Volume Effects
author_facet Mpumelelo Nyathi
author_sort Mpumelelo Nyathi
title Determination of Optimum Planar Imaging Parameters for Small Structures with Diameters Less Than the Resolution of the Gamma Camera
title_short Determination of Optimum Planar Imaging Parameters for Small Structures with Diameters Less Than the Resolution of the Gamma Camera
title_full Determination of Optimum Planar Imaging Parameters for Small Structures with Diameters Less Than the Resolution of the Gamma Camera
title_fullStr Determination of Optimum Planar Imaging Parameters for Small Structures with Diameters Less Than the Resolution of the Gamma Camera
title_full_unstemmed Determination of Optimum Planar Imaging Parameters for Small Structures with Diameters Less Than the Resolution of the Gamma Camera
title_sort determination of optimum planar imaging parameters for small structures with diameters less than the resolution of the gamma camera
publisher Mashhad University of Medical Sciences
series Iranian Journal of Medical Physics
issn 2345-3672
2345-3672
publishDate 2017-12-01
description Introduction: The limited spatial resolution of the gamma camera hinders the absolute quantification of planar images of small structures. The imaged structures are affected by partial volume effects (PVEs), which can spread activity and lead to underestimation of the regional distribution.  The use of optimum planar parameters reduces the impact of the limited spatial resolution of the gamma camera and the statistical noise inherent to low photon count, thus improving quantification. In this study, we aimed to determine the optimum planar imaging parameters for small structures. Materials and Methods: A thyroid protocol was used to acquire planar images of the spheres A, B, and C (16 mm, 12 mm, and 11 mm in diameter, respectively) whilst filled with a targeted activity concentration of technetium-99m. One sphere was mounted at the centre of the Jaszczak Phantom and the other two adjacent to its walls using capillary stems fitted on the spheres. The phantom was filled with distilled water. The targeted activity concentrations used were 74 kBq/mL, 100 kBq/mL, 150 kBq/mL, and 300 kBq/mL. Images of the same count per pixel were acquired on 64  64, 128  128, 256  256, 512  512, and 1024  1024 pixels using a vertical detector mounted 5 cm above the phantom. All the images were quantified using ImageJ software, version 1.48a, Java 1.70_51 [64-bit]. Results: The optimum planar imaging parameters established were a matrix size of 128 128 pixels and technetium-99m solution of activity concentration of 300 kBq/ml. Conclusion: The use of optimal imaging parameters reduces the impact of PVEs, leading to improved quantitative accuracy.
topic Medical Imaging
Radioisotope Imaging
Partial Volume Effects
url http://ijmp.mums.ac.ir/article_9267_2020713899b0a0e8b080c581dfdbd514.pdf
work_keys_str_mv AT mpumelelonyathi determinationofoptimumplanarimagingparametersforsmallstructureswithdiameterslessthantheresolutionofthegammacamera
_version_ 1725800946696454144