A new dynamic torque calibration device at PTB

Research in the field of dynamic torque calibration has gained some momentum in the last few years. At PTB, a calibration device for small dynamic torque levels up to about 0.5 N m proved the working principle of dynamic torque calibrations with sinusoidal excitation. In a next step, a new calibrati...

Full description

Bibliographic Details
Main Authors: Leonard Klaus, Michael Kobusch
Format: Article
Language:English
Published: Elsevier 2021-12-01
Series:Measurement: Sensors
Subjects:
Online Access:http://www.sciencedirect.com/science/article/pii/S2665917421001124
id doaj-4f161342ac834842983510f58c200fee
record_format Article
spelling doaj-4f161342ac834842983510f58c200fee2021-10-07T04:26:49ZengElsevierMeasurement: Sensors2665-91742021-12-0118100149A new dynamic torque calibration device at PTBLeonard Klaus0Michael Kobusch1Corresponding author.; Physikalisch-Technische Bundesanstalt (PTB), Bundesallee 100, 38116, Braunschweig, GermanyPhysikalisch-Technische Bundesanstalt (PTB), Bundesallee 100, 38116, Braunschweig, GermanyResearch in the field of dynamic torque calibration has gained some momentum in the last few years. At PTB, a calibration device for small dynamic torque levels up to about 0.5 N m proved the working principle of dynamic torque calibrations with sinusoidal excitation. In a next step, a new calibration device was developed with the intention of generating higher dynamic torque levels in the range of up to 20 N m. Although this is still less than many industrial applications require, it could pave the way to higher torque levels. The new calibration device features a horizontal arrangement of all components of the drive shaft. The device consists mostly of industrial components that have been modified, rather than custom-made components. After commissioning, optimising the device enabled a frequency range from 10 Hz to up to 400 Hz.http://www.sciencedirect.com/science/article/pii/S2665917421001124Dynamic calibrationDynamic measurementOptimisation of motor dynamicsQuality of vibration
collection DOAJ
language English
format Article
sources DOAJ
author Leonard Klaus
Michael Kobusch
spellingShingle Leonard Klaus
Michael Kobusch
A new dynamic torque calibration device at PTB
Measurement: Sensors
Dynamic calibration
Dynamic measurement
Optimisation of motor dynamics
Quality of vibration
author_facet Leonard Klaus
Michael Kobusch
author_sort Leonard Klaus
title A new dynamic torque calibration device at PTB
title_short A new dynamic torque calibration device at PTB
title_full A new dynamic torque calibration device at PTB
title_fullStr A new dynamic torque calibration device at PTB
title_full_unstemmed A new dynamic torque calibration device at PTB
title_sort new dynamic torque calibration device at ptb
publisher Elsevier
series Measurement: Sensors
issn 2665-9174
publishDate 2021-12-01
description Research in the field of dynamic torque calibration has gained some momentum in the last few years. At PTB, a calibration device for small dynamic torque levels up to about 0.5 N m proved the working principle of dynamic torque calibrations with sinusoidal excitation. In a next step, a new calibration device was developed with the intention of generating higher dynamic torque levels in the range of up to 20 N m. Although this is still less than many industrial applications require, it could pave the way to higher torque levels. The new calibration device features a horizontal arrangement of all components of the drive shaft. The device consists mostly of industrial components that have been modified, rather than custom-made components. After commissioning, optimising the device enabled a frequency range from 10 Hz to up to 400 Hz.
topic Dynamic calibration
Dynamic measurement
Optimisation of motor dynamics
Quality of vibration
url http://www.sciencedirect.com/science/article/pii/S2665917421001124
work_keys_str_mv AT leonardklaus anewdynamictorquecalibrationdeviceatptb
AT michaelkobusch anewdynamictorquecalibrationdeviceatptb
AT leonardklaus newdynamictorquecalibrationdeviceatptb
AT michaelkobusch newdynamictorquecalibrationdeviceatptb
_version_ 1716839893461630976