Effect of blood volume in standard anaerobic blood culture bottles of the BacT/ALERT 3D system used for the detection of pathogens and time to detection.
Blood volume may profoundly affect the isolation of microorganisms in blood cultures. The effect of blood volume in standard anaerobic bottles of the BacT/ALERT 3D system was investigated.Adult patients who visited the emergency department and referred for blood culture (n = 824) were enrolled from...
Main Authors: | Seong Chun Kim, Sunjoo Kim, Dong-Hyun Lee, Sae-Rom Choi, Jeong-Sook Kim |
---|---|
Format: | Article |
Language: | English |
Published: |
Public Library of Science (PLoS)
2015-01-01
|
Series: | PLoS ONE |
Online Access: | http://europepmc.org/articles/PMC4315493?pdf=render |
Similar Items
-
Comparison of ‘time to detection’ values between BacT/ALERT VIRTUO and BacT/ALERT 3D instruments for clinical blood culture samples
by: Francesco Congestrì, et al.
Published: (2017-09-01) -
Evaluation of BacT/Alert FAN Plus Bottles for
the Culture of Peritoneal Dialysate
by: Min-Seung Park, et al.
Published: (2019-12-01) -
The Performance of the Four Anaerobic Blood Culture Bottles BacT/ALERT-FN, -FN Plus, BACTEC-Plus and -Lytic in Detection of Anaerobic Bacteria and Identification by Direct MALDI-TOF MS.
by: Mohammed Almuhayawi, et al.
Published: (2015-01-01) -
Efficiency of BacT/Alert in detecting <em>Salmonella</em>
by: Ida Parwati, et al.
Published: (1998-10-01) -
Evaluation of the novel FAplus, FNplus, and PFplus Blood Culture Bottles (BacT/Alert) – Performance with conventional and MALDI-TOF protocols
by: H. Seifert, et al.
Published: (2014-04-01)