Do we deliver the pressures we intend to when using a T-piece resuscitator?

BACKGROUND: A T-piece resuscitator (TPR) uses a built-in manometer to set the inflation pressures, but we are not informed what pressures are actually delivered distally. Aim of this study was to measure the proximal and distal pressures under different gas conditions when using a TPR. METHODOLOGY/F...

Full description

Bibliographic Details
Main Authors: Evelien Roegholt, Jeroen J van Vonderen, Frans J Walther, Charles C Roehr, Arjan B te Pas
Format: Article
Language:English
Published: Public Library of Science (PLoS) 2013-01-01
Series:PLoS ONE
Online Access:http://europepmc.org/articles/PMC3661533?pdf=render
id doaj-4a691083a7c446538c5ba727337be43c
record_format Article
spelling doaj-4a691083a7c446538c5ba727337be43c2020-11-25T01:29:12ZengPublic Library of Science (PLoS)PLoS ONE1932-62032013-01-0185e6470610.1371/journal.pone.0064706Do we deliver the pressures we intend to when using a T-piece resuscitator?Evelien RoegholtJeroen J van VonderenFrans J WaltherCharles C RoehrArjan B te PasBACKGROUND: A T-piece resuscitator (TPR) uses a built-in manometer to set the inflation pressures, but we are not informed what pressures are actually delivered distally. Aim of this study was to measure the proximal and distal pressures under different gas conditions when using a TPR. METHODOLOGY/FINDINGS: A test lung was ventilated using a TPR (PIP 25 cmH₂O, PEEP 5 cmH₂O) with a gas flow rate of 8 L/min. A) Pressure delivered by six different TPRs was tested. To test variability 20 participants were asked to set PEEP and PIP pressures to 25/5 cmH₂O. B) PIP and PEEP were measured proximal and distal of the TPR when using standard tubing or heated tubing with or without a humidifier. In experiment A mean (SD) proximal PIP and PEEP of the TPRs were respectively 20.3 (0.3) cmH₂O (19.9-20.6 cmH₂O) and 4.9 (0.1) cmH₂O. When 20 participants set pressures; PIP 26.7 (0.5) cm H₂O and PEEP 5.9 (0.44) cmH2O were measured. Experiment B showed that the decrease of PIP between proximal and distal pressures was not clinically significant. However there was a significant decrease of PEEP using the standard tubing (5.1 (0.1) cmH₂O proximally versus 4.8 (0.2) cmH₂O distally; p<0.001) compared to, when using a humidifier with associated tubing and the humidifier turned on, 5.1 (0.1) proximally versus 3.9 (0.2) cmH₂O distally; (p<0.001). CONCLUSION/SIGNIFICANCE: The accuracy of the built-in manometer of a TPR is acceptable. Most pressures set proximally are comparable to the actual pressures delivered distally. However, when using tubing associated with the humidifier PEEP decreases distally by 1.1-1.2 cmH₂O and users should anticipate on this.http://europepmc.org/articles/PMC3661533?pdf=render
collection DOAJ
language English
format Article
sources DOAJ
author Evelien Roegholt
Jeroen J van Vonderen
Frans J Walther
Charles C Roehr
Arjan B te Pas
spellingShingle Evelien Roegholt
Jeroen J van Vonderen
Frans J Walther
Charles C Roehr
Arjan B te Pas
Do we deliver the pressures we intend to when using a T-piece resuscitator?
PLoS ONE
author_facet Evelien Roegholt
Jeroen J van Vonderen
Frans J Walther
Charles C Roehr
Arjan B te Pas
author_sort Evelien Roegholt
title Do we deliver the pressures we intend to when using a T-piece resuscitator?
title_short Do we deliver the pressures we intend to when using a T-piece resuscitator?
title_full Do we deliver the pressures we intend to when using a T-piece resuscitator?
title_fullStr Do we deliver the pressures we intend to when using a T-piece resuscitator?
title_full_unstemmed Do we deliver the pressures we intend to when using a T-piece resuscitator?
title_sort do we deliver the pressures we intend to when using a t-piece resuscitator?
publisher Public Library of Science (PLoS)
series PLoS ONE
issn 1932-6203
publishDate 2013-01-01
description BACKGROUND: A T-piece resuscitator (TPR) uses a built-in manometer to set the inflation pressures, but we are not informed what pressures are actually delivered distally. Aim of this study was to measure the proximal and distal pressures under different gas conditions when using a TPR. METHODOLOGY/FINDINGS: A test lung was ventilated using a TPR (PIP 25 cmH₂O, PEEP 5 cmH₂O) with a gas flow rate of 8 L/min. A) Pressure delivered by six different TPRs was tested. To test variability 20 participants were asked to set PEEP and PIP pressures to 25/5 cmH₂O. B) PIP and PEEP were measured proximal and distal of the TPR when using standard tubing or heated tubing with or without a humidifier. In experiment A mean (SD) proximal PIP and PEEP of the TPRs were respectively 20.3 (0.3) cmH₂O (19.9-20.6 cmH₂O) and 4.9 (0.1) cmH₂O. When 20 participants set pressures; PIP 26.7 (0.5) cm H₂O and PEEP 5.9 (0.44) cmH2O were measured. Experiment B showed that the decrease of PIP between proximal and distal pressures was not clinically significant. However there was a significant decrease of PEEP using the standard tubing (5.1 (0.1) cmH₂O proximally versus 4.8 (0.2) cmH₂O distally; p<0.001) compared to, when using a humidifier with associated tubing and the humidifier turned on, 5.1 (0.1) proximally versus 3.9 (0.2) cmH₂O distally; (p<0.001). CONCLUSION/SIGNIFICANCE: The accuracy of the built-in manometer of a TPR is acceptable. Most pressures set proximally are comparable to the actual pressures delivered distally. However, when using tubing associated with the humidifier PEEP decreases distally by 1.1-1.2 cmH₂O and users should anticipate on this.
url http://europepmc.org/articles/PMC3661533?pdf=render
work_keys_str_mv AT evelienroegholt dowedeliverthepressuresweintendtowhenusingatpieceresuscitator
AT jeroenjvanvonderen dowedeliverthepressuresweintendtowhenusingatpieceresuscitator
AT fransjwalther dowedeliverthepressuresweintendtowhenusingatpieceresuscitator
AT charlescroehr dowedeliverthepressuresweintendtowhenusingatpieceresuscitator
AT arjanbtepas dowedeliverthepressuresweintendtowhenusingatpieceresuscitator
_version_ 1725097791192039424