Calibration of Gas-Metering Constrictions
To extend the range of usefulness of a water calibrated metering constriction in the field of gas flow measurement, a method is described for determining the expansion coefficient (∈) which does not involve measurement of gas flow rate, while the precision of the method is such as to achieve accurac...
Main Author: | |
---|---|
Format: | Article |
Language: | English |
Published: |
SAGE Publishing
1974-01-01
|
Series: | Measurement + Control |
Online Access: | https://doi.org/10.1177/002029407400700105 |
Summary: | To extend the range of usefulness of a water calibrated metering constriction in the field of gas flow measurement, a method is described for determining the expansion coefficient (∈) which does not involve measurement of gas flow rate, while the precision of the method is such as to achieve accuracy tolerances of the order ±0· 30%. When used in conjunction with flow coefficients appropriate to incompressible flow, the flow coefficient appropriate to compressible flow is determined with an overall tolerance well within the limits ±0· 5% over a range of ∈ of 1 > ∈ > 0· 90. |
---|---|